Axium PrimeTM FC-7-30-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL FC-7-30-3D V03 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536034273), received between Jun 7, 2022 and Jun 28, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Packaging Problem1
- Stretched1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
- Stroke/CVA1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2023Break; Stretched; Break; Stretched
MEDTRONIC RECEIVED INFORMATION THAT A FRAMING COIL STRETCHED AND BROKE. THE DAMAGED SECTION WAS IN THE PUSHWIRE PROXIMAL SEGMENT. THE PATIENT HAD AN ASSOCIATED RIGHT SIDE STROKE. DURING THE RETRACTION OF THE 7X30 FRAME 3D, THE COIL STRETCHED. ATTEMPTED TO SNARE THE COIL; UPON SNARING, THE COIL BROKE OFF AND REMAINED IN THE PATIENT. ONCE IMAGING WAS OBTAINED, THE INTERNAL CAROTID ARTERY (ICA) HAD CLOTTED OFF. ATTEMPTED TO ASPIRATE BUT WAS UNSUCCESSFUL. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY OR TESTING. CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE PHYSICIAN REPOSITIONED THE COIL DURING THE PROCEDURE 3 TIMES. THE COIL WAS NOT IMPLANTED AT THE INTENDED LOCATION. ATTEMPTED TO SNARE THE COIL BUT FAILED, THEN THE ICA CLOTTED OFF. THE PUSHWIRE WAS NOT BENT OR BROKEN. THE PHYSICIAN DID NOT ROTATE THE DELIVERY PUSHER DURING THE PROCEDURE. CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE PATIENT WAS BEING TREATED FOR NEUROVASCUALR ABNORMALITIES. IT WAS NOT IN THE ELOQUENT REGION. PATIENT BLOOD FLOW WAS LOW AND VESSEL TORTUOSITY WAS SEVERE.
Patient problems: Device Embedded In Tissue or Plaque; Stroke/CVA
MAUDE report 17221235 ↗MalfunctionJun 7, 2022Detachment of Device or Device Component; Packaging Problem; Detachment of Device or Device Component; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT THE COIL WAS PRE-DETACHED AFTER UNPACKING. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU. THE COIL WAS NOT IMPLANTED IN THE PATIENT. THE PUSHWIRE WAS NOT BENT OR BROKEN. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE ICA PARACLINOID WITH A MAX DIAMETER OF 8MM AND A 3.8MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS SEVERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14622653 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.