Axium Detachable Coil System QC-3-8-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-3-8-HELIX V05 · Class 2 · Implant

All models of Axium Detachable Coil System (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536037373), received between Feb 7, 2023 and Nov 7, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Separation Failure2
  • Break1
  • Inadequacy of Device Shape and/or Size1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2024Separation Failure; Detachment of Device or Device Component; Separation Failure; Detachment of Device or Device Component

    MEDTRONIC RECEIVED A REPORT THAT AN AXIUM COIL COULDN'T BE DETACHED. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, RUPTURED ANEURYSM OF THE POSTERIOR COMMUNICATING ARTERY WITH A MAX DIAMETER OF 5MM AND A 3MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE SURGEON USED THE COIL TO EMBOLIZE A POSTERIOR COMMUNICATING ARTERY ANEURYSM. WHEN THE COIL WAS DETACHED BY THE BACKUP DETACHMENT METHOD, THE DETACHMENT WIRE HAD BEEN COMPLETELY BROKEN OUT, BUT THE COIL WAS NOT DETACHED. SO THE WHOLE WAS COMPLETELY WITHDRAWN FROM THE MICROCATHETER, AND IT AUTOMATICALLY FELL OFF OUTSIDE THE MICROCATHETER AFTER WITHDRAWN. THE PHYSICIAN DID NOT ATTEMPT TO DETACH THE COIL WITH THE INSTANT DETACHER. THERE WAS 1 MANUAL ATTEMPT AT DETACHMENT. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20630727 ↗
  • MalfunctionMay 22, 2024Separation Failure; Separation Failure

    ADDITIONAL INFORMATION RECEIVED THAT THE CAUSE OF THE NON-DETACHMENT WAS NOT DETERMINED. PRIOR TO COIL NON-DETACHMENT, THERE WERE NO ISSUES ENCOUNTERED. THERE WAS NO PUSHWIRE DAMAGE OBSERVED AFTER REMOVAL FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19369912 ↗
  • MalfunctionNov 10, 2023Break; Inadequacy of Device Shape and/or Size; Stretched; Break; Inadequacy of Device Shape and/or Size; Stretched

    ADDITIONAL INFORMATION WAS RECEIVED THAT ISSUES THAT RESULTED IN THE DAMAGE THAT WAS FOUND WAS DURING THE OPERATION, IT WAS TRIED TO RETRACT THE COIL BODY AND FOUND THAT THE DISTAL END WAS UNCONTROLLABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18113119 ↗
  • MalfunctionFeb 7, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    NEW INFORMATION WAS RECEIVED. THE FIELD CONTACT INITIALLY REPORTED THE EVENT TO MEDTRONIC. THE TREATING PHYSICIAN¿S NAME AND CONTACT INFORMATION (PHONE NUMBER) IS ZHOU CHUANKAI. CONTACT INFORMATION THE QC-14-40-3D DETACHABLE SPRING COIL WAS DETACHED WHEN IT WAS TAKEN OUT OF THE PROTECTIVE SHEATH. THE OPERATOR BELIEVED THAT THIS WAS A PRODUCT QUALITY PROBLEM AND REQUESTED TO RETURN IT. PLEASE IDENTIFY IT. THE COIL PREMATURELY SEPARATED FROM THE DELIVERY WIRE. PART OF THE COIL DID NOT FRACTURE AND SEPARATE FROM THE REST. THE CATHETER MODEL AND LOT NUMBER ARE UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310923 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →