Skip to content

Axium Detachable Coil System QC-5-15-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-5-15-HELIX V05 · Class 2 · Implant

All models of Axium Detachable Coil System (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037410), received between Jun 3, 2022 and Nov 7, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Break1
  • Physical Resistance/Sticking1
  • Stretched1
  • Therapy Delivered to Incorrect Body Area1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Paresis1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2023Therapy Delivered to Incorrect Body Area; Stretched; Detachment of Device or Device Component; Physical Resistance/Sticking; Therapy Delivered to Incorrect Body Area; Stretched; Detachment of Device or Device Component; Physical Resistance/Sticking

    ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO ADDITIONAL MEDICAL/SURGICAL INTERVENTION NEEDED TO TREAT THE PREMATURELY DETACHED COIL. IT WAS NOTED THAT THE PATIENT'S RIGHT LIMB IS SLIGHTLY LIMITED, AND THE DEGREE DECREASES WITH TIME. IT'S NOT CLEAR WHETHER IT IS LONG-TERM OR SHORT-TERM.

    Patient problems: Foreign Body In Patient; Paresis

    MAUDE report 18083140 ↗
  • MalfunctionJun 3, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    MEDTRONIC RECEIVED REPORT THAT AN AXIUM COIL BROKE OR DETACHED PREMATURELY DURING DELIVERY. THE COIL WAS REMOVED AND REPLACED TO CONTINUE THE PROCEDURE. THERE WAS NO FRICTION OR OTHER DIFFICULTY DURING DELIVERY. THE PHYSICIAN DID NOT ROTATE THE DELIVERY PUSHER OR ATTEMPT TO REPOSITION THE COIL. NO ATTEMPT HAD BEEN MADE TO DETACH THE COIL.  THE PUSHWIRE WAS NOT BENT OR BROKEN AND THE PUSHWIRE WAS DISCARDED. IT WAS NOTED THE COIL AND ALL ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO HARM OR INJURY TO THE PATIENT WHO WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT AN UNRUPTURED SACCULAR ANTERIOR COMMUNICATING (ACOM) ARTERY ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 5.4MM AND THE NECK DIAMETER WAS 2MM. BLOOD FLOW WAS NORMAL. VESSEL TORTUOSITY WAS MINIMAL. ANCILLARY DEVICE: ECHELON 10 MICROCATHETER (MODEL: 105-5091-150, LOT: B323576).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14594478 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page