Axium Detachable Coil System QC-12-30-HELIXRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-12-30-HELIX V05 · Class 2 · Implant
All models of Axium Detachable Coil System (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536037526), received between Oct 30, 2025 and Oct 30, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Retraction Problem1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2025Retraction Problem; Detachment of Device or Device Component; Retraction Problem; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT THAT DURING THE DELIVERY OF THE COIL, THE OPERATOR FOUND THAT THE COIL BODY HAD DETACHED FROM THE PUSHER ROD. AT THIS POINT, THE COIL COULD NOT BE RETRIEVED AND COULD ONLY BE PUSHED FURTHER; HOWEVER, THE COIL COULD NO LONGER BE PRECISELY DELIVERED INTO THE ANEURYSM THROUGH THE PUSHER ROD, RESULTING IN PART OF THE COIL REMAINING FREE WITHIN THE PATIENT¿S VESSEL. TO ENSURE PATIENT SAFETY, THE OPERATOR DEPLOYED AN ADDITIONAL STENT TO PRESS THE FREE COIL AGAINST THE VESSEL WALL. THE IMPLANT COIL REMAINS IN THE PATIENT. THE COIL WAS NOT IMPLANTED IN THE INTENDED LOCATION. NO SURGICAL OR MEDICINAL INTERVENTION WAS REQUIRED. THE PUSHWIRE WAS NOT BENT OR BROKEN. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY. THE PHYSICIAN REPOSITIONED THE COIL 1 TIME. THE PHYSICIAN DID NOT ATTEMPT TO DETACH THE COIL. THE PHYSICIAN DID NOT ROTATE THE DELIVERY PUSHER DURING THE PROCEDURE. CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SAC
Patient problems: Foreign Body In Patient
MAUDE report 23425629 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page