Axium Detachable Coil System QC-6-20-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-6-20-3D V05 · Class 2 · Implant
All models of Axium Detachable Coil System (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037847), received between Nov 18, 2024 and Jan 30, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Physical Resistance/Sticking1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2026Detachment of Device or Device Component; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL ENCOUNTERED RESISTANCE IN THE MIDDLE OF THE MICROCATHETER. THE PATIENT WAS UNDERGOING RENAL ARTERY EMBOLISM SURGERY. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MINIMAL. IT WAS REPORTED THAT DURING A RENAL ARTERY EMBOLIZATION PROCEDURE, THE OPERATOR USED A TERUMO MICROCATHETER TO DELIVER THE AXIUM COIL. WHEN THE COIL WAS DELIVERED TO THE MIDSECTION OF THE MICROCATHETER, DELIVERY BECAME DIFFICULT AND THE COIL COULD NOT BE ADVANCED FURTHER. AFTER DETACHING THE COIL WITHIN THE MICROCATHETER, THE OPERATOR ATTEMPTED TO FLUSH FROM THE PROXIMAL END WITH A SYRINGE TO PUSH THE COIL FURTHER DISTALLY, BUT FAILED. A NEW COIL WAS THEN USED INSTEAD TO CONTINUE THE PROCEDURE. THE COIL WAS STUCK IN THE MIDDLE OF THE CATHETER. THE PERIPHERAL INTERVENTIONAL COIL COULD NOT BE ADVANCED INTO THE MICROCATHETER. THE CATHETER WAS NOT DAMAGED. THE COIL WAS NOT DAMAGED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE. ANCILLARY DEVICES INCLUDE THE TERUMO MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24226568 ↗MalfunctionNov 18, 2024Separation Failure; Separation Failure
ADDITIONAL INFORMATION WAS RECEIVED THAT THE COIL WAS ATTEMPTED TO BE DETACHED WITH THE DETACHER 8 TIMES. THERE WAS NO ATTEMPT WITH THE MANUAL METHOD. PRIOR TO COIL NON-DETACHMENT, THERE WERE NO ISSUES ENCOUNTERED. THERE WAS NO PUSHWIRE DAMAGE OBSERVED AFTER REMOVAL FROM THE PATIENT. THE CAUSE OF THE EVENT WAS NOT DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20715889 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page