Axium Detachable Coil System QC-8-30-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-8-30-3D V05 · Class 2 · Implant
All models of Axium Detachable Coil System (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00847536037908), received between Dec 11, 2022 and Sep 3, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component2
- Material Deformation2
- Physical Resistance/Sticking2
- No Apparent Adverse Event1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Break; Detachment of Device or Device Component; No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50587415 ↗MalfunctionMar 12, 2026Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A BARE AXIUM 3D COIL (QC-8-30-3D) ENCOUNTERED RESISTANCE AND BECAME STUCK IN THE MIDDLE OF AN ECHELON 10 MICROCATHETER DURING DELIVERY, AND THE MIDDLE SEGMENT OF THE PUSHWIRE WAS FOUND TO BE KINKED. SINCE THE SAME MODEL COIL WAS UNAVAILABLE, THE COIL WAS REPLACED WITH A MEDTRONIC BARE AXIUM 3D COIL (QC-7-30-3D), HOWEVER, DELIVERY WAS STILL DIFFICULT, AND THE TIP OF THE COIL (PUSHWIRE DISTAL SEGMENT) WAS FOUND TO BE DAMAGED WHEN REMOVED. THE COIL WAS REPLACED WITH A NON-MEDTRONIC PRODUCT TO COMPLETE THE PROCEDURE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE PATIENT WAS UNDERGOING COIL EMBOLIZATION SURGERY FOR TREATMENT OF AN UNRUPTURED, SACCULAR, HEPATIC ARTERY ANEURYSM WITH A MAX DIAMETER OF 8 MM AND A 3 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO FRICTION OR DIFFICULTY DURING TESTING OR DELIVERY AND CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. ADDITIONAL INFORMATION RECEIVED. THE CAUSE OF THE RESISTANCE AND THE PUSHWIRE KINK WAS NOT DETERMINED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582924 ↗MalfunctionJan 9, 2026Separation Failure
MEDTRONIC RECEIVED A REPORT THAT AXIUM COIL FAILED TO DETACH. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED C6 SEGMENT ANEURYSM WITH A MAX DIAMETER OF 9 MM AND A 3.5 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT DURING THE SURGERY, IT WAS FOUND THAT THE AXIUM SPRING COIL COULD NOT BE DETACHED. AFTER REPLACING IT WITH ANOTHER SPRING COIL, THE DETACHMENT WAS SUCCESSFUL. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE AXIUM COIL AND ANY ACCESSORY DEVICES PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24020356 ↗MalfunctionFeb 20, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE PREMATURE DETACHMENT WAS NOT DETERMINED. THE PREMATURE DETACHMENT OF THE AXIUM WAS OBSERVED IMMEDIATELY AFTER TAKING THE COIL OUT FROM THE PACKAGE. THERE WAS PREPARATION COMPLETED TO THE COIL PRIOR TO NOTICING THE PREMATURE DETACHMENT. THERE WERE NO DETACHMENT ATTEMPTS MADE WITH THE INSTANT DETACHER OR VIA THE MANUAL METHOD. THERE WAS A KINK OBSERVED IN THE PUSHWIRE. THE AXIUM COIL WAS NOT IMPLANTED IN THE HUMAN BODY. ONLY 1 COIL WAS IMPLANTED BEFORE AND AFTER THIS COIL MALFUNCTIONED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18739168 ↗MalfunctionDec 11, 2022Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT TWO COILS HAD RESISTANCE/STUCK IN SHEATH. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AR TERIOVENOUS FISTULA. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT AFTER THE PRE-OPERATIVE WORK PREPARATION WAS COMPLETED, THE SURGEON SELECTED AND USED THE ECHELON-10 MICROCATHETER TO FILL THE SPRING COIL. THE FIRST SEVERAL COILS WERE FILLED SMOOTHLY AND THE SURGEON SELECTED AND USED QC-9-30-3D. AFTER IT WAS HYDRATED, THE LEATHER SHEATH WAS ALIGNED WITH THE MICROCATHETER TO PUSH AND DELIVER THE SPRING COIL. AT THIS TIME, THE SURGEON INDICATED THAT THE PUSH RESISTANCE OF THE COIL WAS LARGE AND IT COULD NOT BE PUSHED OUT. AFTER IT WAS WITHDRAWN, THE PUSH ROD WAS PULLED OUT AND IT WAS FOUND THAT THE COIL WAS BROKEN IN THE LEATHER SHEATH. AFTER REPLACING WITH ANOTHER QC COIL OF THE SAME MODEL, IT COULD BE USED NORMALLY AND SMOOTHLY. THEN, QC-8-30-3D WAS SELECTED AND USED, THE SAME PROBLEM WAS ENCOUNTERED, IT COULD NOT BE PUSHED OUT. LATER, MULTIPLE QC COILS WERE SELECTED AND USED, AND THE FILLING WENT SMOOTHLY, THE SURGERY WAS COMPLETED. THE SURGEON BELIEVED THAT THESE QC-9-30-3D, QC-8-30-3D COIL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15956366 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.