Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037984), received between Oct 24, 2022 and Apr 1, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Packaging Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Packaging Problem1
  • Physical Resistance/Sticking1
  • Separation Failure1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2025Stretched; Packaging Problem; Stretched; Packaging Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING AN AXIUM COIL THAT WAS STRETCHED PRIOR TO USE IN THE PATIENT'S BODY. THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT C4 SEGMENT GIANT LATERAL UNRUPTURED SACCULAR ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 13MM AND THE NECK DIAMETER WAS 9MM. PATIENT'S VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT DURING THE INITIAL UNPACKING OF A THE AXIUM COIL, IT WAS FOUND STRETCHED AT THE COIL BODY AND CONNECTING ROD OUTSIDE THE PATIENT'S BODY. IT WAS NOTED THAT THE COIL AND ALL ACCESSORY DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE STRETCHED COIL WAS REPLACED WITH ANOTHER MODEL, AND THE SURGERY PROCEEDED SUCCESSFULLY. AS THE ISSUE WAS OBSERVED DURING UNPACKING AND OUTSIDE THE PATIENT'S BODY, IT IS CONSIDERED THAT THERE IS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED REPORTED THERE WAS NO DAMAGE OR EVIDENCE OF TAMPERING TO DEVICE PACKAGING. THE PROXIMAL END OF THE PUSHWIRE WAS SECURED IN THE GUIDEWIRE CLIP (BLACK RUBBER STOPPER) WHEN THE PACKAGE WAS OPENED. THE CAUSE OF THE DAMAGE WAS UNKNOWN. NO PREP WAS PERFORMED PRIOR TO THE DAMAGE BEING SEEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21749128 ↗
  • MalfunctionOct 24, 2022Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking

    ADDITIONAL INFORMATION WAS RECEIVED THAT RESISTANCE OCCURRED IN THE DISTAL SECTION OF THE CATHETER. THE COIL CAME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THERE WAS NO KINK/DAMAGE OBSERVED TO THE COIL OR CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15659165 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
  • Z-0873-2012 · Mar 1, 2010 · Terminated