Axium Detachable Coil System QC-14-40-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-14-40-3D V05 · Class 2 · Implant
All models of Axium Detachable Coil System (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037984), received between Oct 24, 2022 and Apr 1, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Packaging Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Packaging Problem1
- Physical Resistance/Sticking1
- Separation Failure1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2025Stretched; Packaging Problem; Stretched; Packaging Problem
MEDTRONIC RECEIVED INFORMATION REGARDING AN AXIUM COIL THAT WAS STRETCHED PRIOR TO USE IN THE PATIENT'S BODY. THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT C4 SEGMENT GIANT LATERAL UNRUPTURED SACCULAR ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 13MM AND THE NECK DIAMETER WAS 9MM. PATIENT'S VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT DURING THE INITIAL UNPACKING OF A THE AXIUM COIL, IT WAS FOUND STRETCHED AT THE COIL BODY AND CONNECTING ROD OUTSIDE THE PATIENT'S BODY. IT WAS NOTED THAT THE COIL AND ALL ACCESSORY DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE STRETCHED COIL WAS REPLACED WITH ANOTHER MODEL, AND THE SURGERY PROCEEDED SUCCESSFULLY. AS THE ISSUE WAS OBSERVED DURING UNPACKING AND OUTSIDE THE PATIENT'S BODY, IT IS CONSIDERED THAT THERE IS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED REPORTED THERE WAS NO DAMAGE OR EVIDENCE OF TAMPERING TO DEVICE PACKAGING. THE PROXIMAL END OF THE PUSHWIRE WAS SECURED IN THE GUIDEWIRE CLIP (BLACK RUBBER STOPPER) WHEN THE PACKAGE WAS OPENED. THE CAUSE OF THE DAMAGE WAS UNKNOWN. NO PREP WAS PERFORMED PRIOR TO THE DAMAGE BEING SEEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21749128 ↗MalfunctionOct 24, 2022Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED THAT RESISTANCE OCCURRED IN THE DISTAL SECTION OF THE CATHETER. THE COIL CAME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THERE WAS NO KINK/DAMAGE OBSERVED TO THE COIL OR CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15659165 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page