Axium Detachable Coil System QC-25-50-3DRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-25-50-3D V05 · Class 2 · Implant

All models of Axium Detachable Coil System (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536038035), received between Nov 4, 2022 and Nov 14, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure2
  • Stretched2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Stretched; Stretched

    MEDTRONIC RECEIVED A REPORT THAT DURING TRANSHEPATIC INTRAJUGULAR PORTOSYSTEMIC SHUNT (TIPS) SURGERY FOR VASCULAR MALFORMATION EMBOLIZATION, THE AXIUM PRIME BARE COIL WAS SELECTED FOR DELIVERY, AND THE SPRING COIL WAS STRETCHED. IT WAS REPLACED WITH A NEW MEDTRONIC COIL TO COMPLETE THE PROCEDURE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED. THE LESION WAS CHARACTERIZED AS PORTAL VEIN VARICOSE VEINS. THE TIMING OF WHEN THE STRETCH WAS OBSERVED IS UNKNOWN. NO ISSUES WERE REPORTED THAT RESULTED IN THE DAMAGE FOUND. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF VARICOSE VEINS THAT WERE NOT LOCATED IN THE ELOQUENT REGION. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS HIGH AND VESSEL TORTUOSITY WAS MINIMAL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23560459 ↗
  • MalfunctionNov 14, 2025Stretched; Stretched

    MEDTRONIC RECEIVED A REPORT THAT DURING TRANSHEPATIC INTRAJUGULAR PORTOSYSTEMIC SHUNT (TIPS) SURGERY FOR VASCULAR MALFORMATION EMBOLIZATION, THE AXIUM PRIME BARE COIL WAS SELECTED FOR DELIVERY, AND THE SPRING COIL WAS STRETCHED. IT WAS REPLACED WITH A NEW MEDTRONIC COIL TO COMPLETE THE PROCEDURE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED. THE LESION WAS CHARACTERIZED AS PORTAL VEIN VARICOSE VEINS. THE TIMING OF WHEN THE STRETCH WAS OBSERVED IS UNKNOWN. NO ISSUES WERE REPORTED THAT RESULTED IN THE DAMAGE FOUND. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF VARICOSE VEINS THAT WERE NOT LOCATED IN THE ELOQUENT REGION. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS HIGH AND VESSEL TORTUOSITY WAS MINIMAL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23560442 ↗
  • MalfunctionApr 17, 2025Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT BARE AXIUM 3D COIL WAS NOT DETACHED. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM WITH A MAX DIAMETER OF 23 MM AND A 12 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT AFTER THE SPRING COIL WAS TAKEN OUT OF THE PACKAGE, IT WAS INSERTED INTO THE ECHELON MICROCATHETER, AND THE SPRING COIL WAS DETACHED AFTER THE ENTIRE SPRING COIL WAS PUSHED. AFTER BREAKING THE DETACHMENT POINT AT THE TAIL OF THE PUSH ROD AND COMPLETELY PULLING OUT THE NYLON THREAD, THE PUSH ROD WAS PULLED BACK AND IT WAS FOUND THAT THE SPRING COIL HAS NOT BEEN DETACHED. THE UNDETACHED SPRING COIL WAS THEN RETRIEVED AND REPLACED WITH A SPRING COIL OF THE SAME MODEL TO COMPLETE FORMING A HOOP AND SUCCESSFULLY DETACHED IT. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE AN ECHELON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21860072 ↗
  • MalfunctionNov 4, 2022Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THE SPRING COIL COULD NOT BE DETACHED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM. IT WAS REPORTED THAT THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE IFU. THE PHYSICIAN TRIED TO DETACH WITH ANOTHER INSTANT DETACHER. THERE WAS A MANUAL ATTEMPT AT DETACHMENT VIA HYPOTUBE. THERE WAS NO ISSUE WITH THE INSTANT DETACHER.  ANCILLARY DEVICES INCLUDE A ECHELON MICROCATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15725446 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page