Skip to content

Axium Prime Detachable Coil System APB-6-20-HX-SSRecalls & reported problems

Micro Therapeutics, Inc. · COIL APB-6-20-HX-SS V05 · Class 2 · Implant

All models of Axium Prime Detachable Coil System (70)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536038608), received between Jun 13, 2023 and Mar 23, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Physical Resistance/Sticking1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Detachment of Device or Device Component; Physical Resistance/Sticking

    MEDTRONIC RECEIVED A REPORT OF AXIUM COIL PREMATURE DETACHMENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED RIGHT INTERNAL CAROTID ARTERY POSTERIOR COMMUNICATING SEGMENT ANEURYSM WITH A MAX DIAMETER OF 8 MM AND A 5.6 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT DURING TREATMENT OF THE ANEURYSM, AFTER THE MICROCATHETER WAS POSITIONED IN PLACE, A COIL WAS SUCCESSFULLY FILLED TO FORM A HOOP. WHEN DELIVERING THIS COIL, AFTER THE COIL BODY ENTERED THE ANEURYSM SAC, THE DISTRIBUTION WAS UNSATISFACTORY. WHILE PREPARING TO ADJUST THE COIL BODY, DURING WITHDRAWAL IT WAS FOUND THAT THE COIL HAD SPONTANEOUSLY DETACHED WITHIN THE MICROCATHETER. THEREFORE, THE PUSHER ROD WAS WITHDRAWN, ANOTHER COIL WAS FILLED, THE COIL WITHIN THE MICROCATHETER WAS ADVANCED INTO THE ANEURYSM SAC, AND SUBSEQUENT FINISHING COIL WAS FILLED, AFTER WHICH THE PROCEDURE WAS COMPLETED. THE COIL REMAINED IN THE PATIENT. THE COIL WAS IMPLANTED AT THE INTENDED LOCATION. THERE WERE NO SURGICAL OR MEDICINAL INTERVENTION REQUIRED. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY. THE PHYSICIAN REPOSITIONED THE COIL O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24664321 ↗
  • MalfunctionJun 13, 2023Stretched; Stretched

    MEDTRONIC RECEIVED A REPORT THAT THE COIL WAS STRETCHED. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AMORPHOUS, RUPTURED LEFT MIDDLE CEREBRAL ARTERY M1 BIFURCATION ANEURYSM WITH A MAX DIAMETER OF 14MM AND A 7MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. IT WAS REPORTED THAT WHEN TREATING THE RUPTURED AORTIC ANEURYSM AT THE M1 BIFURCATION OF THE LEFT MIDDLE CEREBRAL ARTERY, AFTER THE ECHELON WAS IN PLACE, THE FC COIL FIRST FORMED A HOOP AND WAS NOT DETACHED. AFTER THE STENT MICROCATHETER WAS IN PLACE, THE EP STENT WAS DEPLOYED SMOOTHLY, AND THEN ANOTHER ECHELON WAS PIERCED THROUGH THE MESH, AND THE PRIME SS 6-20 WAS READY TO BE FILLED IN FOR DOUBLE MICROCATHETER EMBOLIZATION. DURING THE PROCESS OF FILLING IN THE COIL, THE RESISTANCE WAS NORMAL, AND THERE WAS NO OBVIOUS RESISTANCE DURING ADJUSTMENT. DURING THE OPERATION, THE COIL WAS FOUND TO BE STRETCHED, SO THE COIL WAS WITHDRAWN. ALTHOUGH IT INTERACTED WITH THE FC THAT FORMED THE HOOP BEFORE, BUT BECAUSE OF THE STENT, THE STRETCHED SS WAS SUCCESSFULLY WITHDRAWN FROM THE BODY COMPLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17118063 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page