Concerto Detachable Coil System NV-2-6-HELIXRecalls & reported problems
Micro Therapeutics, Inc. · COIL NV-2-6-HELIX V07 · Class 2 · Implant
All models of Concerto Detachable Coil System (34)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Concerto detachable coil from Micro Therapeutics is an implanted device used to block blood flow in a vessel, such as to stop bleeding. The 70 reports mostly describe coils that would not release from the catheter or would not detach and be placed, along with coils that resisted passage through the microcatheter. Reports include 4 deaths and 4 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 70
- Deaths reported
- 4
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00847536044876), received between Mar 18, 2022 and Jul 20, 2026 — classified as 4 deaths, 4 injuries, 62 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (40 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure40
- Physical Resistance/Sticking27
- Detachment of Device or Device Component5
- Adverse Event Without Identified Device or Use Problem4
- Inadequacy of Device Shape and/or Size4
- Break2
- Material Deformation2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions66
- Convulsion/Seizure3
- Hematoma3
- Pseudoaneurysm3
- Aneurysm1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2026Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT EACH OF THE 10 CONCERTO HELIX COILS BECAME STUCK IN THE DISTAL SECTION OF A NON-MEDTRONIC MICROCATHETER. NO CATHETER DAMAGE WAS OBSERVED, HOWEVER, THE COIL PUSHWIRE DISTAL SEGMENT WAS OBSERVED TO HAVE KINK/DAMAGE. IT WAS UNKNOWN IF THERE WAS ANY FRICTION OR DIFFICULTY DURING TESTING OR DELIVERY. CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE PHYSICIAN DID NOT REPOSITION THE COIL DURING THE PROCEDURE. IT WAS UNKNOWN WHAT CONDITION WAS BEING TREATED, WHAT THE PATIENT'S BLOOD FLOW OR VESSEL TORTUOSITY WERE, OR IF THERE WERE ANY PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE REPORTED DEVICES AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). ANCILLARY DEVICES INCLUDED TERUMO PROGREAT LAMBDA 1.9FR 110CM MICROCATHETER. ADDITIONAL INFORMATION WAS RECEIVED THAT RESISTANCE AND PUSHWIRE KINK WERE OBSERVED INTRA-OPERATIVELY DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED ACCORDING TO THE STANDARD PROCEDURE; HOWEVER, THE PATIENT¿S RECOVERY STATUS IS CURRENTLY UNKNOWN. NO SPECIFIC CAUSES OR CONTRIBUTING FACTORS FOR THE REPORTED RESISTANCE OR PUSHWIRE KINK DAMAGE WERE IDENTIFIED. THE COILS WER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25882860 ↗MalfunctionMay 29, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL WAS STUCK IN THE CATHETER AND WAS UNABLE TO ADVANCE. THE DEVICE WAS STUCK IN THE MIDDLE OF THE CATHETER. THE CATHETER WAS NOT DAMAGED. THE COIL WAS NOT DAMAGED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT FOR BLEEDS. ANCILLARY DEVICES INCLUDE A BOSTON SCIENTIFIC TRUSELECT MICROCATHETER. (B)(6) 2026 E1 (REP, HCP, FOR): ADDITIONAL INFORMATION RECEIVED. NO CAUSES OR CONTRIBUTING FACTORS FOR THE EVENT WERE IDENTIFIED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER CONCERTO COIL WITH NO ISSUES. THE COIL CAME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THE CATHETER WAS CONTINUOUSLY FLUSHED THROUGHOUT THE PROCEDURE. A TOTAL OF FOUR COILS ENCOUNTERED RESISTANCE, ONE NV-3-8-HELIX COIL AND THREE NV-2-6-HELIX COILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25338114 ↗MalfunctionMay 29, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL WAS STUCK IN THE CATHETER AND WAS UNABLE TO ADVANCE. THE DEVICE WAS STUCK IN THE MIDDLE OF THE CATHETER. THE CATHETER WAS NOT DAMAGED. THE COIL WAS NOT DAMAGED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT FOR BLEEDS. ANCILLARY DEVICES INCLUDE A BOSTON SCIENTIFIC TRUSELECT MICROCATHETER. ADDITIONAL INFORMATION RECEIVED. NO CAUSES OR CONTRIBUTING FACTORS FOR THE EVENT WERE IDENTIFIED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER CONCERTO COIL WITH NO ISSUES. THE COIL CAME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THE CATHETER WAS CONTINUOUSLY FLUSHED THROUGHOUT THE PROCEDURE. A TOTAL OF FOUR COILS ENCOUNTERED RESISTANCE, ONE NV-3-8-HELIX COIL AND THREE NV-2-6-HELIX COILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25338092 ↗MalfunctionApr 30, 2026Inadequacy of Device Shape and/or Size; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT CONCERTO COILS WOULD NOT DETACH. THE PATIENT WAS UNDERGOING SURGERY OF A RENAL BLEED. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MINIMAL. IT WAS REPORTED THAT UPON ACCESS OF THE RENAL VESSEL THE PHYSICIAN ATTEMPTED TO DEPLOY THE CONCERTO COIL AND WAS UNSUCCESSFUL X 3. THE STAFF REPORTED THAT THE COIL WOULD SUCCESSFULLY EXIT THE CATHETER BUT WOULD NOT TAKE SHAPE. THREE SIZES WERE ATTEMPTED AND FAILED. THREE SIZES WERE SUCCESSFULLY DEPLOYED. ALL DEVICES WE PREPARED AS INDICATED IN THE IFU. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE A AVANTI 6F SHEATH, OMNI SOS GUIDE CATHETER, RENEGADE HIGH FLOW MICRO CATHETER, FATHOM GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25047380 ↗MalfunctionMar 16, 2026Separation Failure; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE COIL WAS USED TO TREAT A BLEEDING AFTER A HIP TOTAL ENDOPROSTHESIS (TEP). THE COIL COULD NOT BE PUSHED THROUGH THE PROGREAT 2.4. THERE WAS A RESISTANCE IN THE DISTAL PART OF THE MICROCATHETER. THEREFORE NO IMPLANT OF THE COIL WAS POSSIBLE. THE PHYSICIAN USED AN NV-2-4-HELIX TO COMPLETE THE PROCEDURE (WITH THE SAME MICROCATHETERS) WITHOUT ANY PROBLEMS. THE REPORT WAS CLARIFIED THAT THE COIL COULD NOT BE DELIVERED AND DEPLOYED BECAUSE OF RESISTANCE. THE INITIAL INFORMATION REPORTED WAS INCORRECT. NO CONTINUOUS FLUSH WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24608072 ↗MalfunctionMar 13, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT DURING AN ATTEMPT TO ADVANCE A CONCERTO NV-2-6-HELIX COIL THROUGH A PHENOM 21 MICROCATHETER, THE COIL COULD NOT BE ADVANCED INTO THE MICROCATHETER. EVEN AFTER THE PROTECTIVE SHEATH WAS FULLY ADVANCED TO THE BASE OF THE MICROCATHETER HUB, THE COIL COULD NOT BE DELIVERED SMOOTHLY. THE COIL WAS REPLACED WITH A CONCERTO NV-2-8-HELIX COIL, HOWEVER, THE SAME RESISTANCE WAS ENCOUNTERED. THE COIL WAS REPLACED WITH A NON-MEDTRONIC PRODUCT TO COMPLETE THE PROCEDURE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE PATIENT WAS UNDERGOING COIL EMBOLIZATION SURGERY FOR TREATMENT FOR HYPOPHARYNGEAL CANCER TUMOR. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MINIMAL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). ANCILLARY DEVICES INCLUDED RSR18 GUIDECATHETER, ASAHI WAMS-180-16ST GUIDEWIRE. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. RESISTANCE WAS ENCOUNTERED AT THE HUB PORTION OF THE CATHETER. THE COIL EXITED THE PROTECTIVE SHEATH SMOOTHLY, AND A CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24592708 ↗MalfunctionFeb 13, 2026Separation Failure
MEDTRONIC RECEIVED A REPORT THAT A CONCERTO COIL WOULD NOT DETACH. THE PHYSICIAN ATTEMPTED TO DETACH THE COIL WITH THE INSTANT DETACHER AND THE MANUAL METHOD WITH THE HYPOTUBE 2 TIMES. THE PHYSICIAN DID NOT TRY TO DETACH WITH A SECOND INSTANT DETACHER DEVICE. IT WAS INDICATED THAT THE REPORTED DEVICE WAS NOT PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), NOTED AS NO CONTINUOUS FLUSH PREFORMED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24333001 ↗MalfunctionFeb 5, 2026Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL WOULD NOT DETACH. THE INSTANT DETACHER WAS NOT USED TO ATTEMPT TO DETACH THE COIL. THE PHYSICIAN DID NOT TRY TO DETACH WITH ANOTHER INSTANT DETACHER. THERE WAS A MANUAL ATTEMPT AT DETACHMENT VIA HYPOTUBE. THE MANUAL METHOD WAS ATTEMPTED NUMEROUS TIMES. THERE WAS NO ISSUE WITH THE INSTANT DETACHER. THERE WAS NON-DETACHMENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE A PROGREAT MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268481 ↗MalfunctionJan 21, 2026Detachment of Device or Device Component; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT A CONCERTO HELIX COIL WAS DETACHED BEFORE USE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. IT WAS UNKNOWN WHAT CONDITION WAS BEING TREATED OR WHAT THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24136726 ↗MalfunctionJan 20, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING INTERVENTIONAL RADIOLOGY AND VASCULAR SURGERY SUPPORT FOR AN INFERIOR EPIGASTRIC BLEED. IT WAS NOTED THAT THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE UNKNOWN. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, THAT REQUIRED THE USE OF EIGHT EV3 CONCERTO COILS TO EMBOLIZE THE VESSEL. THERE WERE ISSUES INTRODUCING THREE 2MMX6CM HELIX COILS INTO A TRUSELECT MICROCATHETER, TWO OF THE COILS WOULD NOT GO PASSED THE HUB OF THE MICROCATHETER AND ONE WAS SEIZED WITHIN THE CATHETER AT THE DISTAL END. IT WAS UNKNOWN WHETHER THE IMPLANT COIL WAS PUSHED SMOOTHLY OUT FROM THE INTRODUCER SHEATH. THIS RESULTED IN THE REMOVAL OF THE TRUSELECT MICROCATHETER WHERE THE COIL AND THE CATHETER WERE FOUND TO BE INTACT. WHEN THE SECOND TRUSELECT WAS INSERTED, AND THE CORRECT VESSEL WAS RESELECTED, THE THIRD CONCERTO COIL INTRODUCTION WAS ATTEMPTED BUT FAILED TO GET PASSED THE HUB OF THE NEW MICROCATHETER. THE EMBOLIZATION OF THE VESSEL WAS OBTAINED BY USING A DIFFERENT BRAND COIL. THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. A CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. THE REPORTED DE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24126624 ↗MalfunctionJan 16, 2026Separation Failure
MEDTRONIC RECEIVED A REPORT THAT A CONCERTO COIL DID NOT DETACH. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU). IT WAS UNKNOWN IF THERE WERE ANY PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24101906 ↗MalfunctionJan 16, 2026Separation Failure
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE EVENT WAS NOT DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24094364 ↗MalfunctionNov 26, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THREE CONCERTO COILS EXPERIENCED RESISTANCE IN THE MICROCATHETER. THE PATIENT WAS UNDERGOING TREATMENT FOR A LEFT RENAL ANGIOGRAM WITH A POSSIBLE PSEUDO ANEURYSM. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT THE CONCERTO COILS WOULD NOT TRACK THROUGH THE MICRO CATHETER. THE PHYSICIAN FELT RESISTANCE AND COULD NOT ADVANCE THE COILS. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY DEPLOY OTHER COILS AFTER THIS EVENT THROUGH THE SAME MICROCATHETER. NEITHER THE COIL NOR THE CATHETER WERE DAMAGED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE EVENT WAS NOT DETERMINED. CATHETER RESISTANCE WAS AT THE HUB. THE COIL DID NOT COME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THE CATHETER WAS FLUSHED IN BETWEEN THE USE OF COILS. THE INFORMATION IS CONFIRMED BY THE PHYSICIAN. ANCILLARY DEVICES INCLUDE A SHEATH: 5FR PINNACLE, AND A MICROCATHETER: 2FR TRUE SELECT. ADDITIONAL INFORMATION RECEIVED REPORTED THE COILS PASSED THROU
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23659751 ↗MalfunctionNov 21, 2025Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE PATIENT'S BASELINE WAS NORMAL/THE FISTULA WAS IN THE PATIENT'S DIALYSIS GRAFT IN THE UPPER EXTREMITY. THERE WAS NOT A POST THROMBECTOMY CONDITION. THE CAUSE OF THE RESISTANCE WAS BELIEVED TO BE DUE TO NOT HYDRATING THE COIL BEFORE THE INTRODUCTION INTO THE MICROCATHETER. THE CAUSE OF THE PREMATURE DEPLOYMENT WAS DUE TO RESISTANCE IN THE MICROCATHETER AND FORWARD FORCE. THE COIL SEEMED TO MOVE FREELY OUT OF THE COIL INTRODUCER SHEATH. THE INFORMATION WAS PROVIDED BY THE SCRUB ASSISTANT IN THE PROCEDURE. THREE OTHER COILS WERE PLACED AFTER THE RESISTANCE/PREMATURE DEPLOYMENT WITH A NEW MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23623883 ↗MalfunctionNov 14, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PRODUCT WAS BEING USED AS PER THE INSTRUCTIONS FOR USE (IFU), HOWEVER, THE CONCERTO HELIX COIL GOT STUCK WITHIN THE MICROCATHETER. A NEW MEDTRONIC DEVICE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT A CONTINUOUS CATHETER FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE CATHETER WAS NOT A MEDTRONIC PRODUCT. THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. THE INFORMATION IS CONFIRMED BY THE PHYSICIAN. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23560672 ↗MalfunctionNov 10, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A CONCERTO HELIX COIL COULD NOT BE ADVANCED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED. THE PROCEDURE WAS COMPLETED BY USING A NEW PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23513586 ↗MalfunctionNov 3, 2025Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT NV-2-4-HELIX (BATCH NO: 231316177) WAS DISCOVERED KINKED UPON OPENING THE PACKAGE. THERE WAS PUSHER KINK/BROKEN. THE DAMAGED LOCATION WAS THE PUSHWIRE MIDDLE SEGMENT. ANOTHER COIL NV-2-6-HELIX (BATCH NO: 231221824) WAS UNABLE TO DEPLOY. THERE WAS NON-DETACHMENT. THE PHYSICIAN ATTEMPTED TO DETACH THE COIL WITH THE INSTANT DETACHER TWO TIMES. THE PHYSICIAN TRIED TO DETACH WITH ANOTHER INSTANT DETACHER. THERE WAS ONE MANUAL ATTEMPT AT DETACHMENT VIA HYPOTUBE. THERE WAS NO ISSUE WITH THE INSTANT DETACHER. ANOTHER COIL NV-2-6-HELIX (BATCH NO: 231221824) WAS DISCOVERED KINKED UPON OPENING. THERE WAS PUSHER KINK/BROKEN. THERE WAS NO FRICTION OR DIFFICULTY DURING TESTING OR DELIVERY. THE CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE DAMAGED SECTION WAS THE PUSHWIRE MIDDLE SEGMENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT IS ALIVE WITH NO INJURY. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ANEURYSM. ANCILLARY DEVICES INCLUDE A TERUMO SHEATH, BOSTON TRUSLECT MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23458266 ↗MalfunctionJul 31, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT THE COIL CONCERTO COULD NOT BE ADVANCED. IT WAS REPORTED THAT THE REPORTED DEVICE AND ALL ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT A NEW PRODUCT WAS USED TO COMPLETE THE PROCEDURE. CAUSE OF THE RESISTANCE WAS UNKNOWN. A CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. THE INFORMATION IS CONFIRMED BY THE PHYSICIAN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22683462 ↗MalfunctionJul 22, 2025Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PRODUCT WAS BEING USED AS PER THE INSTRUCTIONS FOR USE (IFU) BUT THE COIL WAS KINKED AND GOT STUCK WITHIN THE MICROCATHETER. A NEW MEDTRONIC DEVICE WAS USED TO FINISH THE PROCEDURE. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THE COIL CAME OUT OF THE INTRODUCER SHEATH SMOOTHLY. CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. THE CATHETER WAS NOT A MEDTRONIC DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22572967 ↗MalfunctionJul 15, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT THE CONCERTO COIL COULD NOT PASS THROUGH THE MICROCATHETER DURING THE PROCEDURE. A NEW MEDTRONIC DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICE USE WAS A MICROCATHETER. ADDITIONAL INFORMATION RECEIVED REPORTED THE RESISTANCE WAS IN THE CATHETER HUB. CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE OBSERVED TO THE COIL OR CATHETER. THE CATHETER WAS NOT A MEDTRONIC PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22512802 ↗MalfunctionMay 31, 2025Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED REPORTED CAN'T CONFIRM IF THE COIL WOULD BE ABLE TO DETACH. THERE WAS RESISTANCE WHEN THE PHYSICIAN ATTEMPTED TO ADVANCE THE COIL TO DEPLOY IT, HOWEVER SINCE THERE'S NO SUCCESSFUL DEPLOYMENT, THERE WAS NO ATTEMPT TO DETACH IT. A COIL HAS TO BE SUCCESSFULLY DEPLOYED INTO THE VESSEL BEFORE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22120467 ↗MalfunctionMay 27, 2025Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THERE WAS A DETACH DEFECT WITH A CONCERTO COIL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). IT WAS UNKNOWN IF ANY PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087994 ↗MalfunctionFeb 26, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE COIL CONCERTO HELIX 2MM X 6CM, THE COIL COULD NOT BE ADVANCED. THE DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE PATIENT OUTCOME WAS REPORTED AS ALIVE WITH NO INJURY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21470992 ↗MalfunctionJan 21, 2025Separation Failure; Separation Failure
IT WAS REPORTED THAT DURING THE USE OF THE COIL CONCERTO HELIX 2 MM X 6 CM, THE DEVICE WOULD NOT DEPLOY, AND THERE WAS A NON-DETACHMENT OF THE COIL. THE DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE. THE INSTANT DETACHER WAS DISCARDED, AND IT IS UNKNOWN IF THERE WAS ANY ISSUE WITH IT. THE PRODUCT WAS REPLACED BY ANOTHER MEDTRONIC PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21200779 ↗InjuryDec 5, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A THE COIL COULD NOT BE DEPLOYED THROUGH THE CATHETER. THE COIL WAS REMOVED AND REPLACED WITH NO HARM TO THE PATIENT.
Patient problems: Aneurysm; No Clinical Signs, Symptoms or Conditions
MAUDE report 20855027 ↗
FDA clearance and classification
Cleared through 510(k) K081465 on Aug 19, 2008 (decision: Substantially Equivalent). Submitted by Micro Therapeutics, Inc. as “Axium Detachable Coil System”. The FDA review took 84 days. FDA record ↗
The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Common questions about the Concerto Detachable Coil System NV-2-6-HELIX
Has the Concerto Detachable Coil System NV-2-6-HELIX been recalled?
No FDA recall record in our data names this exact device identifier (DI 00847536044876), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Concerto Detachable Coil System NV-2-6-HELIX?
The FDA's MAUDE database lists 70 adverse event reports for this device, received between Mar 18, 2022 and Jul 20, 2026. In those reports, the most common specific device problems are “Separation Failure” (40), “Physical Resistance/Sticking” (27) and “Detachment of Device or Device Component” (5); the most common patient problems are “Convulsion/Seizure” (3), “Hematoma” (3) and “Pseudoaneurysm” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Concerto Detachable Coil System NV-2-6-HELIX?
4 reports are classified as deaths, 4 as injuries and 62 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Concerto Detachable Coil System NV-2-6-HELIX FDA approved?
It was cleared through 510(k) K081465 on Aug 19, 2008. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Concerto Detachable Coil System NV-2-6-HELIX MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Concerto Detachable Coil System NV-2-6-HELIX, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand Concerto Detachable Coil System, model NV-2-6-HELIX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization117 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page