Concerto Detachable Coil System PV-2-6-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL PV-2-6-3D V05 · Class 2 · Implant
All models of Concerto Detachable Coil System (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536045095), received between Nov 20, 2025 and Mar 6, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking2
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL WAS STUCK IN THE MIDDLE OF THE CATHETER. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A HEMORRHOIDS. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE TREATMENT OF CHOICE WAS EMBOLIZATION OF SUPERIOR RECTAL ARTERY BRANCHES USING A NON-MEDTRONIC MICROCATHETER WITH COILS. ALL DEVICES INVOLVED WERE PREPARED AS PER INSTRUCTIONS FOR USE (IFU). THE MICROCATHETER WAS UNDER CONTINUOUS FLUSHING THROUGH THE ENTIRE OPERATION. THE PV-2-6-3D CONCERTO COIL WAS PREPARED AS PER IFU. THE PHYSICIAN ADVANCED THE COIL THROUGH THE MICROCATHETER, AND AT ABOUT HALF OF THE MICROCATHETER'S LENGTH THE COIL GOT STUCK AND COULD NOT MOVE FORWARD. THE PHYSICIAN DECIDED NOT TO USE EXCESSIVE FORCE AND TO RETRIEVE BACK THE COIL. THE COIL WAS TAKEN OUT OF THE PATIENT. THE OPERATION WENT ON USING THE BALD SPI018 COILS ON THE SAME MICROCATHETER WITH NO ISSUE AT ALL. THE COIL WAS STUCK IN THE MIDDLE OF THE CATHETER. THERE WAS NO DAMAGE TO THE CATHETER AND COIL. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24537283 ↗MalfunctionNov 20, 2025Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR A SACCULAR, UNRUPTURED GASTROINTESTINAL ANEURYSM WITH A MAX DIAMETER OF 3 MM AND A 1 MM NECK DIAMETER. IT WAS NOTED THAT THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT CONCERTO COIL 2 MM X 6MM WAS UNABLE TO DEPLOY AND BECAME STUCK IN MIDDLE SECTION OF THE HEADWAY 17 MICRO CATHETER. THE COIL WAS REMOVED AND DIFFERENT COIL WAS USED. THE CATHETER AND THE COIL WERE NOT DAMAGED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23606647 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page