Concerto Detachable Coil System PV-6-20-3DRecalls & reported problems

Micro Therapeutics, Inc. · COIL PV-6-20-3D V05 · Class 2 · Implant

All models of Concerto Detachable Coil System (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00847536045163), received between May 16, 2024 and Nov 6, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking4
  • Break1
  • Material Deformation1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2025Break; Material Deformation; Physical Resistance/Sticking; Break; Material Deformation; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED REPORTED THE COIL HAD SOME RESISTANCE DURING INSERTION IN THE MICROCATHETER. THE MICROCATHETER WAS NOT A MEDTRONIC DEVICE. WHEN THE COIL WAS WITHDRAWN, IT WAS NOTED THE THE WIRE WAS KINKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23490194 ↗
  • MalfunctionJun 9, 2025Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THERE WAS A FAILURE TO RELEASE THE SPRING AT THE TIME OF SURGERY. THE DEVICE WAS REMOVED FROM THE PATIENT AND A NEW SPRING WAS INSTALLED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. NO MEDICAL OR SURGICAL INTERVENTION WAS NEEDED TO PREVENT A PERMANENT IMPAIRMENT OF A FUNCTION. THE EVENT DID NOT LEAD TO OR EXTEND PATIENT HOSPITALIZATION. THERE WAS NO INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22170367 ↗
  • MalfunctionJan 31, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT TWO CONCERTO COILS BECAME STUCK IN THE CATHETER. THE SPECIFIC LOCATION WITHIN THE CATHETER WHERE THE RESISTANCE WAS ENCOUNTERED IS UNKNOWN, AND IT IS NOT SPECIFIED WHETHER THE CATHETER OR COILS WERE DAMAGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THAT IT WAS UNKNOWN WHAT CATHETER WAS USED, BUT IT WAS NOT A MEDTRONIC DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283084 ↗
  • MalfunctionMay 17, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    MEDTRONIC RECEIVED A REPORT THAT THE COIL COULD NOT BE ADVANCED IN ATTEMPTS TO USE IT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19343958 ↗
  • MalfunctionMay 16, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    MEDTRONIC RECEIVED A REPORT THAT THE COIL COULD NOT BE ADVANCED IN ATTEMPTS TO USE IT. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). IT WAS UNKNOWN IF THERE WERE ANY PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PROCEDURE WAS COMPLETED WITH A NEW PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336620 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →