M SERIES BIPHASIC 40010511101123010Recalls & reported problems

Zoll Medical Corporation · Defibrillator ,MAIN, MNL-1, SPO2, NIBP, BIPHASIC, PACE, RCD, SMR, CDMK, AC, DMST, NELLCOR · Class 2

All models of M SERIES BIPHASIC (59)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00847946004217), received between Sep 29, 2017 and Oct 2, 2019 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Failure of Device to Self-Test” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure of Device to Self-Test2
  • Failure to Power Up2
  • Failure to Deliver Shock/Stimulation1
  • Failure to Discharge1
  • Loss of Power1
  • Use of Device Problem1

Patient problems

  • No Patient Involvement7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2019Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE KNOB WILL NOT TURN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 9145830 ↗
  • MalfunctionJun 12, 2019Failure to Deliver Shock/Stimulation; Failure to Power Up; Failure to Deliver Shock/Stimulation; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. WHEN POWERED UP WITH AC, THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8693500 ↗
  • MalfunctionNov 5, 2018Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8039353 ↗
  • MalfunctionMay 4, 2018Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7487722 ↗
  • MalfunctionApr 10, 2018Loss of Power; Loss of Power

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7414225 ↗
  • MalfunctionFeb 20, 2018Failure to Discharge; Failure to Discharge

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WAS UNABLE TO DISCHARGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7282346 ↗
  • MalfunctionSep 29, 2017Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 6903587 ↗