The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00847946022334), received between Jun 11, 2018 and Mar 3, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deliver Shock/Stimulation4
  • Failure to Discharge2
  • Output Problem2
  • Failure of Device to Self-Test1
  • Grounding Malfunction1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information3
  • No Patient Involvement2
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2026Grounding Malfunction

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497185 ↗
  • MalfunctionFeb 27, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21482729 ↗
  • MalfunctionSep 15, 2022Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) THE ASSOCIATED DEVICE WAS UNABLE TO DETECT THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER SET OF INTERNAL HANDLES TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 15428236 ↗
  • MalfunctionAug 10, 2022Output Problem; Output Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE INTERNAL HANDLES DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 15200807 ↗
  • MalfunctionAug 10, 2022Output Problem; Output Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE INTERNAL HANDLES DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 15200303 ↗
  • MalfunctionDec 28, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13102205 ↗
  • MalfunctionMar 19, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE ASSOCIATED DEVICE FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11529529 ↗
  • MalfunctionNov 17, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO CARDIOVERT A PATIENT (AGE & GENDER UNKNOWN), THESE INTERNAL HANDLES FAILED TO DISCHARGE WHILE ATTACHED TO AN ASSOCIATED DEFIBRILLATOR. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10853736 ↗
  • MalfunctionJan 31, 2019Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED TO DISCHARGE USING THESE ATTACHED INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 8293754 ↗
  • MalfunctionJun 13, 2018Failure to Discharge; Failure to Discharge

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Information

    MAUDE report 7596435 ↗
  • MalfunctionJun 11, 2018Failure to Discharge; Failure to Discharge

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Patient Involvement

    MAUDE report 7586957 ↗

Other Dc-defibrillator, low-energy, (including paddles) devices

All Dc-defibrillator, low-energy, (including paddles) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.