Internal Handle 1011-0139-02Recalls & reported problems
Zoll Medical Corporation · Class 2
All models of Internal Handle (14)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00847946022334), received between Jun 11, 2018 and Mar 3, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver Shock/Stimulation4
- Failure to Discharge2
- Output Problem2
- Failure of Device to Self-Test1
- Grounding Malfunction1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information3
- No Patient Involvement2
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Grounding Malfunction
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24497185 ↗MalfunctionFeb 27, 2025Failure of Device to Self-Test; Failure of Device to Self-Test
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21482729 ↗MalfunctionSep 15, 2022Unable to Obtain Readings; Unable to Obtain Readings
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN) THE ASSOCIATED DEVICE WAS UNABLE TO DETECT THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER SET OF INTERNAL HANDLES TO CONTINUE TREATING THE PATIENT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 15428236 ↗MalfunctionAug 10, 2022Output Problem; Output Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE INTERNAL HANDLES DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 15200807 ↗MalfunctionAug 10, 2022Output Problem; Output Problem
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE INTERNAL HANDLES DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 15200303 ↗MalfunctionDec 28, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13102205 ↗MalfunctionMar 19, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE ASSOCIATED DEVICE FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11529529 ↗MalfunctionNov 17, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO CARDIOVERT A PATIENT (AGE & GENDER UNKNOWN), THESE INTERNAL HANDLES FAILED TO DISCHARGE WHILE ATTACHED TO AN ASSOCIATED DEFIBRILLATOR. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10853736 ↗MalfunctionJan 31, 2019Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE ASSOCIATED DEFIBRILLATOR FAILED TO DISCHARGE USING THESE ATTACHED INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 8293754 ↗MalfunctionJun 13, 2018Failure to Discharge; Failure to Discharge
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE ASSOCIATED DEFIBRILLATOR WAS UNABLE TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Information
MAUDE report 7596435 ↗MalfunctionJun 11, 2018Failure to Discharge; Failure to Discharge
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE ASSOCIATED DEVICE FAILED TO DISCHARGE USING THESE INTERNAL HANDLES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Patient problems: No Patient Involvement
MAUDE report 7586957 ↗
FDA clearance and classification
Cleared through Special 510(k) K042302 on Oct 26, 2004 (decision: Substantially Equivalent). Submitted by Zoll Medical Corp as “Zoll Autoclavable Internal Handles, Zoll Autoclavable Internal Handles With Integrated Electrodes, Zoll Autoclavable”. The FDA review took 62 days. FDA record ↗
The FDA regulates product code LDD (dc-defibrillator, low-energy, (including paddles)) under 21 CFR 870.5300 as a Class II device. 21 CFR 870.5300 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Internal Handle 1011-0139-02
Has the Internal Handle 1011-0139-02 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00847946022334), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Internal Handle 1011-0139-02?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Jun 11, 2018 and Mar 3, 2026. In those reports, the most common specific device problems are “Failure to Deliver Shock/Stimulation” (4), “Failure to Discharge” (2) and “Output Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Internal Handle 1011-0139-02?
None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Internal Handle 1011-0139-02 FDA approved?
It was cleared through Special 510(k) K042302 on Oct 26, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Internal Handle 1011-0139-02, and is it still sold?
The FDA's device database lists Zoll Medical Corporation as the company responsible for it (brand Internal Handle, model 1011-0139-02). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Dc-defibrillator, low-energy, (including paddles) devices
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall6 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall5 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall1 report
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall0 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall0 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall0 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall0 reports
- NA 11131Physio-Control, Inc. · Dc-defibrillator, low-energy, (including paddles)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page