RESTORIS 180604Recalls & reported problems
Mako Surgical Corp. · MCK Onlay Tibia Baseplate · Class 2 · Implant
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00848486000707), received between Dec 11, 2020 and Sep 12, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K082172, K090763, K150307, K220930, K242831); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem3
- Fracture2
- Break1
- Loss of Osseointegration1
Patient problems
- Ambulation or Postural Difficulties3
- Inadequate Osseointegration2
- Insufficient Information2
- Pain2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 12, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT'S LEFT MEDIAL UNI KNEE WAS REVISED DUE TO THE TIBIAL BASEPLATE BREAKING IN HALF (NO CAUSE OR CONTRIBUTOR TO THE BREAKAGE WAS REPORTED TO THE REP). THE INSERT BECAME DAMAGED/ WARPED AS A RESULT. THE PATIENT WAS CONVERTED TO A TKA WITH STEM AND AUGMENTS. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 23036275 ↗InjuryMar 4, 2024Fracture; Fracture
MAKO UNI TIBIA BASEPLATE WAS BROKE IN HALF. POLY WAS STILL INTACT. THE UNI WAS REVISED INTO A TKA. NOTICED VIA X-RAY. ORIGINAL SURGERY WAS AROUND 7-8 YEARS AGO. UPDATE 08/FEBRUARY/2024 WG: LEFT MEDIAL KNEE. SALES MANAGER CONFIRMED THERE WERE NO ALLEGATIONS AGAINST THE REVISED INSERT.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 18824584 ↗InjuryMay 25, 2023Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem
REVISION FROM MEDIAL UNI TO TOTAL KNEE REPLACEMENT. THE TIBIAL IMPLANT WAS LOOSE WITH MINIMAL BONDING TO THE CEMENT MANTLE. FEMUR WAS WELL FIXATED. PATIENT PRESENTED WITH MEDIAL KNEE PAIN. LEFT KNEE.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 17000851 ↗InjuryJun 30, 2022Osseointegration Problem; Osseointegration Problem
RESTORIS PKA WAS REVISED AFTER A PATIENT FALL AND SUBSEQUENT LOOSENING OF THE TIBIAL COMPONENT. LEFT SIDE.
Patient problems: Inadequate Osseointegration
MAUDE report 14878874 ↗InjuryMar 9, 2021Osseointegration Problem; Osseointegration Problem
PATIENT HAD A REVISION. PATIENT HAD A LEFT MAKO PARTIAL KNEE (RESTORIS) INSITU APPROXIMATELY 12 MONTHS OLD. THEY PRESENTED WITH MEDIAL PAIN, MRI SHOWED MEDIAL TIBIAL LOOSENING. INTRA OP TIBIAL BASEPLATE LIFTED OFF WITH NO SIGN OF CEMENT INTERDIGITATION. PRIMARY TRIATHLON KNEE IMPLANTED.
Patient problems: Pain
MAUDE report 11447284 ↗InjuryDec 11, 2020Break; Break
MAKO BASEPLATE WAS REVISED AND WAS BROKEN IN 2 PIECES.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 10992246 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082172, K090763, K150307, K220930, K242831. One submission often covers many models, so check the product and lot details in each record.
- Z-1385-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-1386-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-2051-2018 · Jan 19, 2018 · TerminatedRestoris MCK Onlay Insert Extractor.
- Z-1031-2016 · Jan 15, 2016 · TerminatedRestoris Multicompartmental Knee (MCK) System RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO.
- Z-1132-2014 · Jan 9, 2014 · TerminatedRestoris Multicompartmental Knee (MCK) System is an implant system designed to be used with MAKO's Robotic Arm Interactive Orthopedic System (RIO). It is composed of a unicompartme
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.