Skip to content

RESTORIS 180613Recalls & reported problems

Mako Surgical Corp. · MCK Onlay Tibia Baseplate · Class 2 · Implant

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00848486000769), received between Nov 20, 2020 and Apr 1, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (K082172, K090763, K150307, K220930, K242831); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Loss of Osseointegration2
  • Fracture1
  • Migration1
  • Osseointegration Problem1
  • Unstable1

Patient problems

  • Inadequate Osseointegration3
  • Bone Fracture(s)2
  • Joint Laxity1
  • Pain1
  • Periprosthetic fracture1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2026Fracture; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING INFORMATION WAS PROVIDED: BOOKED REVISION SURGERY. PATIENT WAS EXPERIENCING PAIN AND SHOWED SOME SIGNS OF TIBIAL LOOSENING ON SPEC CT SCAN, SURGEON MENTIONED HE BELIEVES THE PATIENT EXPERIENCED A STRESS FRACTURE AFTER PRIMARY SURGERY FOR THE POTENTIAL CAUSE. IMPLANT APPEARED WELL BONDED TO THE CEMENT MANTLE DURING EXPLANTATION.

    Patient problems: Inadequate Osseointegration; Pain; Periprosthetic fracture

    MAUDE report 24758374 ↗
  • InjuryFeb 28, 2024Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    REVISED A RIGHT MEDIAL MCK UNI. REVISION WAS DUE TO TIBIAL LOOSENING/SUBSIDENCE. PATIENT WAS CONVERTED TO A TOTAL KNEE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18796494 ↗
  • InjuryNov 27, 2023Migration; Migration

    REVISED A RIGHT MEDIAL UNI DUE TO PER-ARTICULAR FRACTURE AND SUBSEQUENT TIBIAL BASEPLATE COLLAPSE. HE REVISED TO A TOTAL KNEE WITH MEDIAL TIBIAL AUGMENT, STEM.

    Patient problems: Bone Fracture(s)

    MAUDE report 18212119 ↗
  • InjuryOct 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING INSERTION OF MCK TIBIAL BASEPLATE, THE PROXIMAL MEDIAL SIDE OF THE TIBIA WAS FRACTURED. MANUAL TKA WAS PERFORMED.

    Patient problems: Bone Fracture(s)

    MAUDE report 17997102 ↗
  • InjuryMay 20, 2021Unstable; Unstable

    PKA WAS REVISED TO TKA DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 11856017 ↗
  • InjuryNov 20, 2020Loss of Osseointegration; Loss of Osseointegration

    A RESTORIS BASEPLATE WAS REVISED DUE TO LOOSENING. RIGHT MEDIAL UNICOMP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 10874768 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082172, K090763, K150307, K220930, K242831. One submission often covers many models, so check the product and lot details in each record.

  • Z-1386-2023 · Mar 10, 2023 · Open, Classified
    stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure
  • Z-1385-2023 · Mar 10, 2023 · Open, Classified
    stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
  • Z-2051-2018 · Jan 19, 2018 · Terminated
    Restoris MCK Onlay Insert Extractor.
  • Z-1031-2016 · Jan 15, 2016 · Terminated
    Restoris Multicompartmental Knee (MCK) System RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO.
  • Z-1132-2014 · Jan 9, 2014 · Terminated
    Restoris Multicompartmental Knee (MCK) System is an implant system designed to be used with MAKO's Robotic Arm Interactive Orthopedic System (RIO). It is composed of a unicompartme

Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.