Scorpio 73-3910Recalls & reported problems
Howmedica Osteonics Corp. · Universal Dome Patella · Class 2 · Implant
All models of Scorpio (107)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327033915), received between Nov 8, 2023 and Nov 8, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Discolored” (4 reports).
2 FDA recall records name this device's identifier, 2 still open.
8 other recalls cover devices cleared under the same FDA submission (K051948, K052917, K243817, K252898, K972967); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Open, Classified
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Z-3206-2026 · Root cause: Process change control · Recall details · FDA record ↗
- Dec 1, 2023Open, Classified
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Z-0898-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Discolored4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2023Material Discolored; Material Discolored
73-3910 LOT 4873 CAME OUT OF THE BOX YELLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18097225 ↗MalfunctionNov 8, 2023Material Discolored; Material Discolored
73-3910 LOT 4873 CAME OUT OF THE BOX YELLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18097171 ↗MalfunctionNov 8, 2023Material Discolored; Material Discolored
73-3910 LOT 4873 CAME OUT OF THE BOX YELLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18097167 ↗MalfunctionNov 8, 2023Material Discolored; Material Discolored
73-3910 LOT 4873 CAME OUT OF THE BOX YELLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18097166 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051948, K052917, K243817, K252898, K972967. One submission often covers many models, so check the product and lot details in each record.
- Z-3197-2026 · Jul 30, 2026 · Open, ClassifiedALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PAT
- Z-0897-2024 · Dec 1, 2023 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
- Z-0896-2024 · Dec 1, 2023 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708
- Z-0895-2024 · Dec 1, 2023 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
- Z-0894-2024 · Dec 1, 2023 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308
- Z-0900-2024 · Dec 1, 2023 · Open, ClassifiedScorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0910
- Z-0893-2024 · Dec 1, 2023 · Open, ClassifiedSCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
- Z-0606-2012 · Aug 17, 2011 · TerminatedEius Unicompartmental Knee System Stryker Howmedica Osteonics Corp. Howmedica International S. de R.L., Raheen Business Park, Limerick, Ireland. Intended for use in Unicompartment
Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Scorpio 73-3508Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls3 reports
- Arthrex® AR-503-TTTCArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall15 reports
- Arthrex® AR-503-TTTFArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- Scorpio 73-3308Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- Arthrex® AR-504-PSC9Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER11 reports
- RESTORIS 160405Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
- Arthrex® AR-504-PSD0Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
- RESTORIS 180613Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page