RESTORIS 180616Recalls & reported problems
Mako Surgical Corp. · MCK Onlay Tibia Baseplate · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00848486000790), received between Aug 29, 2019 and Nov 19, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K082172, K090763, K150307, K220930, K242831); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information3
- Osseointegration Problem1
Patient problems
- Pain3
- Inadequate Osseointegration1
- Injury1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 19, 2024Osseointegration Problem; Osseointegration Problem
REVISION OF RIGHT PARTIAL MEDIAL UNI TO RIGHT TOTAL KNEE. ORIGINAL SURGERY WAS RIGHT PARTIAL MEDIAL UNICOMPARTMENTAL KNEE DONE IN (B)(6) 2014 (EXACT DATE UNKNOWN). REASON FOR REVISION WAS TIBIAL COMPONENT WAS LOOSE. ALL THREE COMPONENTS (MCK FEMUR, MCK TIBIA, AND POLY) WERE REVISED AND REPLACED WITH TOTAL KNEE IMPLANTS.
Patient problems: Inadequate Osseointegration
MAUDE report 20727578 ↗InjuryMar 5, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
THIS PI IS FOR THE IMPLANTS USED IN PATIENT 3 OF 3. (B)(6) MALE, SURGERY (B)(6) 2019. RIGHT KNEE. AS REPORTED BY SURGEON: ATTACHED ARE XRAYS OF THREE PATIENTS I PREVIOUSLY DISCUSSED. ALL HAVE CONCERNING RADIOLUCENCIES UNDERNEATH THE TIBIAL BASEPLATE THAT APPEARED SEVERAL MONTHS POST OP. THUS FAR, NONE HAVE REQUIRED REVISION BUT ALL HAVE SOME DEGREE OF PAIN (MOSTLY MEDIAL). AGAIN I NEVER SAW THIS PRIOR THE MICS TRANSITION. LET ME KNOW YOUR THOUGHTS AND IF ANY OTHER SURGEONS HAVE SEEN THIS ISSUE. MORE RECENTLY, I HAVE BEEN USING A 2MM DRILL TO PERFORATE THE SCLEROTIC TIBIAL SURFACE TO PROMOTE BETTER CEMENT INTERDIGITATION AND MY RESULTS SEEM TO BE MUCH BETTER. CASE TYPE: PKA 3.0.
Patient problems: Pain
MAUDE report 9794039 ↗InjuryMar 5, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
THIS PI IS FOR THE IMPLANTS USED IN PATIENT 2 OF 3. (B)(6) YO FEMALE, SURGERY (B)(6) 2019. AS REPORTED BY SURGEON: ATTACHED ARE XRAYS OF THREE PATIENTS I PREVIOUSLY DISCUSSED. ALL HAVE CONCERNING RADIOLUCENCIES UNDERNEATH THE TIBIAL BASEPLATE THAT APPEARED SEVERAL MONTHS POST OP. THUS FAR, NONE HAVE REQUIRED REVISION BUT ALL HAVE SOME DEGREE OF PAIN (MOSTLY MEDIAL). AGAIN I NEVER SAW THIS PRIOR THE MICS TRANSITION. LET ME KNOW YOUR THOUGHTS AND IF ANY OTHER SURGEONS HAVE SEEN THIS ISSUE. MORE RECENTLY, I HAVE BEEN USING A 2MM DRILL TO PERFORATE THE SCLEROTIC TIBIAL SURFACE TO PROMOTE BETTER CEMENT INTERDIGITATION AND MY RESULTS SEEM TO BE MUCH BETTER. CASE TYPE: PKA 3.0.
Patient problems: Pain
MAUDE report 9794038 ↗InjuryAug 29, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
REVISION OF A RESTORIS MEDIAL UNICOMPARTMENT KNEE TO A TOTAL KNEE REPLACEMENT. PATIENT HAD A REVISION UNDERTAKEN DUE TO PAIN UPON WALKING. FEMORAL COMPONENT APPEARED WELL FIXED, TIBIAL COMPONENT WAS EASILY MOVED DURING THE REVISION. ONCE IMPLANTS WERE REMOVED THE TIBIAL IMPLANT SHOWED NO VISIBLE SIGNS OF BONDING TO THE CEMENT NOR THE BONE. FEMUR WAS WELL FIXED. A TOTAL KNEE REPLACEMENT WAS COMPLETED SATISFACTORILY USING A TRIATHLON PS IMPLANT, WITHOUT DELAY.
Patient problems: Injury; Pain
MAUDE report 8947010 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082172, K090763, K150307, K220930, K242831. One submission often covers many models, so check the product and lot details in each record.
- Z-1386-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-1385-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-2051-2018 · Jan 19, 2018 · TerminatedRestoris MCK Onlay Insert Extractor.
- Z-1031-2016 · Jan 15, 2016 · TerminatedRestoris Multicompartmental Knee (MCK) System RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO.
- Z-1132-2014 · Jan 9, 2014 · TerminatedRestoris Multicompartmental Knee (MCK) System is an implant system designed to be used with MAKO's Robotic Arm Interactive Orthopedic System (RIO). It is composed of a unicompartme
Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Scorpio 73-3910Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls4 reports
- Scorpio 73-3508Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls3 reports
- Arthrex® AR-503-TTTCArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall15 reports
- Arthrex® AR-503-TTTFArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- Scorpio 73-3308Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- Arthrex® AR-504-PSC9Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER11 reports
- Arthrex® AR-504-PSD0Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
- RESTORIS 160405Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.