RESTORIS 180601Recalls & reported problems
Mako Surgical Corp. · MCK Onlay Tibia Baseplate · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00848486001995), received between Aug 26, 2020 and Aug 26, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Loss of Osseointegration” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K082172, K090763, K150307, K220930, K242831); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Loss of Osseointegration1
Patient problems
- Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2020Loss of Osseointegration; Loss of Osseointegration
THIS PI IS FOR THE LEFT KNEE. AS REPORTED IN PI (B)(4): IT WAS REPORTED BY SURGEON THAT PRE OP X-RAYS SHOWED MEDIAL COMPARTMENT WEAR. ALL THE STANDARD CRITERIA FOR MEDIAL COMPARTMENT ARTHROPLASTY WERE FULFILLED AND SURGEON PERFORMED A BILATERAL PARTIAL KNEE MEDIAL MAKOPLASTY. PRIMARY SURGERY PERFORMED ON (B)(6) 2017, BY DR. THE SURGEON REPORTED THAT THE CEMENTING TECHNIQUE IS AS WHAT HE WOULD DO FOR ALL MEDIAL MAKO KNEES AND HAVE HAD NO ISSUES. PATIENT POST OP X-RAYS WERE GOOD POSITION. EVEN HER POST OP MRI DONE ON HER KNEES 9 MONTHS AFTER POST SURGERY SHOWED NO LOOSENING AND STABLE IMPLANTS, NOR EDEMA. POST OP X-RAYS SHOWED THE IMPLANTS TO BE PERFECTLY PLACED. HOWEVER IN 2019 IT WAS NOTED THAT SHE DEVELOPED ACCELERATED LOOSENING OF BOTH MEDIAL IMPLANTS AND THIS RESULTED IN A BILATERAL REVISION TOTAL KNEE ARTHROPLASTY. THE REVISION WAS PERFORMED BY DR (B)(6) 2019. THE OPERATIVE NOTES SHOWED BILATERAL EXTENSIVE POLYETHYLENE WEAR WITH RESULTANT EXTENSIVE SYNOVITIS AND LOOSENING. REVISION PERFORMED USING NON STRYKER PRODUCT. NO DETAILS PROVIDED AS PATIENT WENT TO CONSULT ANOTHER SURGEON. UPDATED: 03 AUG 2020 - REFER IMPLANT PICTURE IN INTAKE TAB. RECENTLY, THE PATIENT CAME BACK TO VIS
Patient problems: Injury; No Clinical Signs, Symptoms or Conditions
MAUDE report 10454103 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082172, K090763, K150307, K220930, K242831. One submission often covers many models, so check the product and lot details in each record.
- Z-1386-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-1385-2023 · Mar 10, 2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
- Z-2051-2018 · Jan 19, 2018 · TerminatedRestoris MCK Onlay Insert Extractor.
- Z-1031-2016 · Jan 15, 2016 · TerminatedRestoris Multicompartmental Knee (MCK) System RESTORIS MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO.
- Z-1132-2014 · Jan 9, 2014 · TerminatedRestoris Multicompartmental Knee (MCK) System is an implant system designed to be used with MAKO's Robotic Arm Interactive Orthopedic System (RIO). It is composed of a unicompartme
Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Scorpio 73-3910Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls4 reports
- Scorpio 73-3508Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls3 reports
- Arthrex® AR-503-TTTCArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall15 reports
- Arthrex® AR-503-TTTFArthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- Scorpio 73-3308Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- Arthrex® AR-504-PSC9Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER11 reports
- Arthrex® AR-504-PSD0Arthrex, Inc. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
- RESTORIS 160405Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.