Skip to content

NA 112280Recalls & reported problems

Mako Surgical Corp. · KNEE TIBIAL ARRAY · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00848486015923), received between Aug 2, 2019 and Mar 13, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Device Slipped1
  • Failure to Read Input Signal1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Non Reproducible Results1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2023Device Slipped; Detachment of Device or Device Component; Device Slipped; Detachment of Device or Device Component

    UPON INSPECTION OF BUMPED ARRAY, THE SURGEON NOTED THE ARRAY ITSELF HAD COME LOOSE FROM THE POST. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16530439 ↗
  • MalfunctionFeb 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Failure to Read Input Signal; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Read Input Signal

    TIBIA REGISTRATION COULD NOT BE COMPLETED. THEY FOUND THE BASE OF THE ARRAY WAS MOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16319166 ↗
  • MalfunctionAug 2, 2019Output Problem; Non Reproducible Results; Output Problem; Non Reproducible Results

    DURING A MAKO TKA THE SURGEON HAD DIFFICULT MAINTAINING A REGISTRATION ON THE TIBIA AND FEMUR. ON INSPECTION THE ARRAY ITSELF WAS WOBBLY AT THE CENTRE ATTACHMENT. SURGICAL DELAY OF APPROXIMATELY 1 HOUR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8858171 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources and updates