Skip to content

NA 144140Recalls & reported problems

Mako Surgical Corp. · BONE PINS (4MM X 140MM) · Class 2 · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00848486018177), received between Jun 11, 2018 and Dec 1, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Insufficient Device Problem Information2
  • Non Reproducible Results2
  • Break1
  • Fracture1
  • Material Integrity Problem1
  • Positioning Failure1

Patient problems

  • Bone Fracture(s)4
  • No Known Impact Or Consequence To Patient2
  • Abdominal Distention1
  • Abdominal Pain1
  • Anemia1
  • Hematoma1
  • Hemorrhage/Blood Loss/Bleeding1
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 1, 2025Positioning Failure; Insufficient Device Problem Information; Positioning Failure; Insufficient Device Problem Information

    NO NEW INFORMATION.

    Patient problems: Abdominal Distention; Abdominal Pain; Anemia; Hematoma; Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension; Perforation of Vessels

    MAUDE report 23680481 ↗
  • InjuryJan 6, 2025Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem

    A FRACTURE OCCURRED IN THE ILIAC CREST OF THE HEALTHY LEG WHERE THE ANCHORING PIN WAS PLACED.

    Patient problems: Bone Fracture(s)

    MAUDE report 21082755 ↗
  • InjuryFeb 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ONE OF MY SURGEONS TOLD ME THAT ONE OF HIS PATIENTS HAD A FEMORAL FRACTURE NEAR THE SITE OF A PIN PLACED FOR A PRIMARY MAKO KNEE. THE FRACTURE DID NOT OCCUR INTRAOPERATIVELY. THE SURGEON SAID THE POST OP X-RAY SHOWED NO FRACTURE. THE PATIENT REPORTED ROLLING OVER IN BED, FEELING A "POP", AND GOING INTO THE ER TO FIND SHE HAD A FRACTURED FEMUR. CASE TYPE/PROCEDURE: LEFT TKA. PER RESPONSE: "THE CASE WAS A TKA. NO REVISION WAS DONE. THE FEMUR WAS REDUCED AND PLATED BY ANOTHER SURGEON."

    Patient problems: Bone Fracture(s)

    MAUDE report 16342552 ↗
  • InjuryFeb 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ONE OF MY SURGEONS TOLD ME THAT ONE OF HIS PATIENTS HAD A FEMORAL FRACTURE NEAR THE SITE OF A PIN PLACED FOR A PRIMARY MAKO KNEE. THE FRACTURE DID NOT OCCUR INTRAOPERATIVELY. THE SURGEON SAID THE POST OP X-RAY SHOWED NO FRACTURE. THE PATIENT REPORTED ROLLING OVER IN BED, FEELING A "POP," AND GOING INTO THE ER TO FIND SHE HAD A FRACTURED FEMUR. CASE TYPE/PROCEDURE: LEFT TKA. PER RESPONSE: "THE CASE WAS A TKA. NO REVISION WAS DONE. THE FEMUR WAS REDUCED AND PLATED BY ANOTHER SURGEON."

    Patient problems: Bone Fracture(s)

    MAUDE report 16342509 ↗
  • InjuryMay 8, 2019Material Integrity Problem; Material Integrity Problem

    SUPRACONDYLAR FRACTURE OF FEMUR AT OR NEAR PIN LOCATION AT 6 WEEKS POST-OP. CASE TYPE: TKA.

    Patient problems: Bone Fracture(s)

    MAUDE report 8593687 ↗
  • MalfunctionApr 23, 2019Break; Non Reproducible Results; Break; Non Reproducible Results

    SURGEON WAS REMOVING FEMORAL PINS AFTER COMPLETING MAKO PORTION OF THE TOTAL KNEE REPLACEMENT AND THE FEMORAL DISTAL PIN SNAPPED WHILE BEING PULLED OUT. CASE TYPE: TKA. SURGICAL DELAY: =15 MINUTES. DID THE EVENT OCCUR DURING THE CASE? AS PER THE MPS: YES, THE EVENT OCCURRED DURING THE CASE. WE HAD JUST OPENED FINAL IMPLANTS BUT HAD NOT IMPACTED YET. WAS THE PATIENT UNDER ANESTHESIA? YES. WERE ANY DEBRIS/COMPONENTS LEFT INSIDE OF THE PATIENT? (YES/NO) AS PER THE MPS: YES, A PIECE OF THE PIN WAS LEFT INSIDE THE BONE. THE SURGERY WAS PAUSED AND AN X-RAY WAS TAKEN. THE SURGEON DECIDED TO LEAVE THE PIECE OF PIN INSIDE OF THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8540296 ↗
  • InjuryJun 11, 2018Insufficient Device Problem Information; Non Reproducible Results; Insufficient Device Problem Information; Non Reproducible Results

    AS REPORTED: CONVERSATION WITH DR. (B)(6) ABOUT HAVING ISSUES WITH THE 4X140 AND 4X170 PINS CRACKING THE FEMORAL CORTEX WHEN BEING DRILLED IN FOR ARRAY PLACEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7586291 ↗

Other Orthopedic stereotaxic instrument devices