TRIDENT, RIO 209820Recalls & reported problems
Mako Surgical Corp. · STRAIGHT SHELL IMPACTOR · Class 2
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00848486030278), received between Dec 5, 2018 and Oct 31, 2019 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (7 reports).
4 other recalls cover devices cleared under the same FDA submission (K141989, K170593); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Non Reproducible Results7
- Break6
- Contamination /Decontamination Problem3
- Fracture2
- Difficult to Remove1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2019Break; Fracture; Break; Fracture
BACK OF STRAIGHT IMPACTION HANDLE SHEARED OFF DURING IMPACTION. CASE TYPE: THA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9260160 ↗MalfunctionOct 3, 2019Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
DURING IMPACTION, PHYSICIAN WAS UNABLE TO LOCK PLATFORM INTO PLACE. UPON CLOSER INSPECTION, FOUND STRAIGHT IMPACTOR TO BE BROKEN. CASE TYPE: THA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9151434 ↗MalfunctionSep 12, 2019Break; Fracture; Break; Fracture
THE END OF THE IMPACTOR SHAFT BROKE OFF. CASE TYPE: THA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9007959 ↗MalfunctionJun 21, 2019Contamination /Decontamination Problem; Non Reproducible Results; Contamination /Decontamination Problem; Non Reproducible Results
IMPACTOR CONTINUES TO CONTAIN BI BURDEN AFTER STERILIZATION. CASE TYPE: THA; SURGICAL DELAY: X - =15 MINUTES. UPDATE: "YES PATIENT WAS UNDER ANESTHESIA".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8724081 ↗MalfunctionJun 21, 2019Contamination /Decontamination Problem; Non Reproducible Results; Contamination /Decontamination Problem; Non Reproducible Results
IMPACTOR CONTINUES TO CONTAIN BIO BURDEN AFTER STERILIZATION. CASE TYPE: THA. SURGICAL DELAY: 15 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8724076 ↗MalfunctionJun 21, 2019Contamination /Decontamination Problem; Non Reproducible Results; Contamination /Decontamination Problem; Non Reproducible Results
IMPACTOR CONTAINING BIO BURDEN AFTER STERILIZATION. CASE TYPE: THA. SURGICAL DELAY = 15 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8723982 ↗MalfunctionMay 30, 2019Break; Break
THE END OF THE MAKO STRAIGHT CUP IMPACTOR BROKE OFF INSIDE OF THE IMPACTION PLATFORM. CASE TYPE: THA. UPDATE: "YES THE IMPACTOR BROKE DURING IMPACTION".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8653509 ↗MalfunctionFeb 14, 2019Mechanical Problem; Non Reproducible Results; Mechanical Problem; Non Reproducible Results
AFTER THE CUP WAS FULLY IMPACTED DR. ALI HAD A LOT OF DIFFICULTY UNSCREWING THE CUP FROM THE IMPACTION HANDLE. UPON REMOVAL I NOTICED THE THREADS WERE DAMAGED CAUSING THE CUP TO CROSS THREAD. CASE TYPE: (B)(4). SURGICAL DELAY: SEVERAL MINUTES ATTEMPTING TO UNSCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8340946 ↗MalfunctionJan 3, 2019Break; Non Reproducible Results; Break; Non Reproducible Results
END OF IMPACTION HANDLE SHEARED OFF DURING IMPACTION - IMPACTED THE CUP AND WHEN THE ARM WAS TAKEN AWAY FROM THE PATIENT AND THE IMPACTION PLATFORM TAKEN OFF THE IMPACTION HANDLE, PART OF THE IMPACTION HANDLE HAD BEEN SHEARED OFF AND FELL OUT OF THE IMPACTION PLATFORM ONTO THE STERILE BACK TABLE. CASE TYPE: THA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8211778 ↗MalfunctionDec 26, 2018Difficult to Remove; Non Reproducible Results; Difficult to Remove; Non Reproducible Results
THA EVENT DESCRIPTION: WHEN IMPACTING THE CUP IT WAS CROSS-THREADED AND IT WAS DIFFICULT TO UNSCREW THE CUP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8196828 ↗MalfunctionDec 5, 2018Break; Non Reproducible Results; Break; Non Reproducible Results
DURING IMPACTION, SHAFT AND PLATFORM BOTH BROKE. SURGICAL DELAY: 15 MINUTES. CASE TYPE: THA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8135094 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141989, K170593. One submission often covers many models, so check the product and lot details in each record.
- Z-2745-2020 · Jul 9, 2020 · TerminatedMako Hip End Effector, Variable Angle Catalog Number: 206967 GTIN:00848486022167
- Z-3182-2018 · Dec 27, 2017 · TerminatedMakoplasty RIO Standard System; Ethernet to Fiber Optic Converter Catalog # 200933 Product Usage: The Partial Knee Application (PKA), for use with the Mako System, is indicated fo
- Z-3131-2017 · Aug 7, 2017 · TerminatedTotal Hip Application (THA) Product Usage: The Robotic Arm Interactive Orthopedic System (RIO) is intended to assist the surgeon in providing software define spatial boundaries f
- Z-1411-2017 · Dec 22, 2016 · TerminatedMAKO RIO THA Application User Guides, for total hip arthroplasty, orthopedic surgery.
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.