Mako Surgical Corp. · Sagittal Saw Attachment · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Mako sagittal saw attachment is a bone-cutting attachment used in robot-assisted knee surgery. Its 154 reports mostly describe cleaning residue left on the device, attachments loosening or detaching, and excess vibration when cutting. The reports include no deaths and 1 injury, and 153 are malfunctions. One recall, started in April 2019, is closed and concerns saw attachments whose bearings were not lubricated.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
153
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 153 adverse event reports that list this exact device (DI 00848486032111), received between Sep 5, 2018 and Jan 16, 2026 — classified as 1 injury, 152 malfunctions.

The most frequently reported specific device problem is “Residue After Decontamination” (94 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 25, 2019Terminated

    The bearings of the saw attachments were ungreased.

    Z-1824-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Residue After Decontamination94
  • Mechanical Problem23
  • Detachment of Device or Device Component14
  • Contamination of Device Ingredient or Reagent12
  • Non Reproducible Results10
  • Premature Separation8
  • Device Slipped6
  • Noise, Audible6

Patient problems

  • No Clinical Signs, Symptoms or Conditions97
  • No Known Impact Or Consequence To Patient57
  • No Patient Involvement6
  • Muscle/Tendon Damage1

Recent adverse event reports

The 25 most recent of 153 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Fracture; Mechanical Problem

    THE FOLLOWING INFORMATION WAS PROVIDED: STRAIGHT ATTACHMENT BROKE DURING BONE PREPARATION. SURGEON WAS CUTTING THE POSTERIOR CONDYLE WHEN THE ATTACHMENT CLEARLY BECAME LOOSE. SECOND MAKO KNEE TRAY OPENED AND STRAIGHT ATTACHMENT REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24095629 ↗
  • MalfunctionJan 5, 2026Mechanical Problem; Positioning Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23968223 ↗
  • MalfunctionDec 17, 2025Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component

    THE FOLLOWING INFORMATION WAS PROVIDED: AFTER THE PROCEDURE, THIS WAS DISCOVERED DURING INSTRUMENT INSPECTION. SINCE THE PROCEDURE CONCLUDED WITHOUT INCIDENT, NO ADVERSE EVENTS OCCURRED. WHEN SECURELY ATTACHING THE SAW BLADE TO THE MAKO MICS SAGITTAL SAW STRAIGHT ATTACHMENT, LATERAL MOVEMENT WAS OBSERVED. ADDITIONALLY, THE PROTRUDING SECTION OF THE MICS CONNECTION POINT ALSO ROTATES. CONTINUED USE MAY CAUSE THE SAW BLADE TO LOOSEN DURING MICS OPERATION, SO REPLACEMENT IS REQUESTED. UPDATE ON 12/05/2025: 1. COULD YOU PLEASE ELABORATE ON WHAT YOU MEAN BY LATERAL MOVEMENT? ARE YOU REFERRING TO A WIDER OSCILLATION COMPARED TO THE REGULAR ONE? YES. DURING POSTOPERATIVE INSPECTION, IT WAS CONFIRMED THAT THE SAW BLADE WAS SECURELY FASTENED TO THE STRAIGHT ATTACHMENT, HOWEVER LATERAL MOVEMENT WAS OBSERVED BY MANUAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23825985 ↗
  • InjuryJun 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Noise, Audible; Incorrect, Inadequate or Imprecise Result or Readings; Noise, Audible

    STRAIGHT SAW ATTACHMENT MAKES EXTREMELY LOUD NOISE WHILE EXECUTING THE CUTS WITH A NARROW BLADE. IT IS SUSPECTED THAT THE OSCILLATION MOTION IS WIDER THAN NORMAL. HALF WAY OF EXECUTING THE TIBIA CUT, IT WAS NOTICED THE PATELLAR TENDON WAS DAMAGED BY THE BLADE. IT HAD TO BE STITCHED.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 19562149 ↗
  • MalfunctionNov 6, 2023Residue After Decontamination; Residue After Decontamination

    LEAKING OIL CASE TYPE / APPLICATION: TKA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18071736 ↗
  • MalfunctionOct 23, 2023Difficult to Insert; Premature Separation; Difficult to Insert; Premature Separation

    THE SALES REP RECEIVED THE FOLLOWING MESSAGE: "THAT ATTACHMENT NOT WORK( IT'S NOT HOLDING THE SAW BLADE). WE CANCEL THE SECOND CASE TODAY"C UPDATE: LOCKING MECHANISM FAILURE OCCURRED INTRA-OPERATIVELY. DUE TO THE FAILURE, A FOLLOWING CASE WAS CANCELLED PRIOR TO ANY PATIENT (ANESTHESIA) INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17985595 ↗
  • MalfunctionOct 10, 2023Contamination; Residue After Decontamination; Contamination; Residue After Decontamination

    SAWS STAINED STERILIZING TUBING WITH SOME OILY RESIDUE. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17906908 ↗
  • MalfunctionOct 10, 2023Contamination; Residue After Decontamination; Contamination; Residue After Decontamination

    SAWS STAINED STERILIZING TUBING WITH SOME OILY RESIDUE. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17906907 ↗
  • MalfunctionOct 9, 2023Residue After Decontamination; Residue After Decontamination

    ATTACHMENT LEAKING OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17894146 ↗
  • MalfunctionOct 5, 2023Contamination; Residue After Decontamination; Contamination; Residue After Decontamination

    SAW LEAKING FLUID/ OIL CASE TYPE / APPLICATION: TKA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17874825 ↗
  • MalfunctionSep 18, 2023Residue After Decontamination; Residue After Decontamination

    SAW LEAKING GREASE AND OIL CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17762920 ↗
  • MalfunctionSep 18, 2023Residue After Decontamination; Residue After Decontamination

    ALL SAWS LEAKING GREASE AND OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17762911 ↗
  • MalfunctionSep 18, 2023Residue After Decontamination; Residue After Decontamination

    ALL SAWS LEAKING GREASE AND OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17762906 ↗
  • MalfunctionSep 15, 2023Residue After Decontamination; Residue After Decontamination

    ALL SAWS LEAKING GREASE AND OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17760643 ↗
  • MalfunctionSep 15, 2023Residue After Decontamination; Residue After Decontamination

    ALL SAWS LEAKING GREASE AND OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17760642 ↗
  • MalfunctionAug 25, 2023Residue After Decontamination; Residue After Decontamination

    OIL LEAKING OUT OF STRAIGHT SAW ATTACHMENT IN TRAY. CONTAMINATED TRAY. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17622411 ↗
  • MalfunctionMay 31, 2023Residue After Decontamination; Residue After Decontamination

    WORN SAW ATTACHMENT- REPLACEMENT NEEDED UNABLE TO USE - RUST. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17029661 ↗
  • MalfunctionMay 26, 2023Residue After Decontamination; Residue After Decontamination

    INNER MECHANISM LOOSE, OIL LEAKED. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17012242 ↗
  • MalfunctionMay 26, 2023Residue After Decontamination; Residue After Decontamination

    SPD SAYS THAT THE ATTACHMENTS ARE LEAKING LUBRICANT. THAT THEY STERILIZED THEM AND THAT WHEN OPENED IN THE ROOM THAT THERE WAS GREASE ALL OVER ALL THE INSTRUMENTS. THEY AREN'T SURE WHICH PART IS LEAKING THE LUBRICANT AND WANT TO GET NEW ONE. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17012209 ↗
  • MalfunctionApr 28, 2023Residue After Decontamination; Noise, Audible; Residue After Decontamination; Noise, Audible

    SAW AND ATTACHMENTS VERY LOUD. CHECKPOINTS AT THE EDGE OF RANGE AT 1.2 AND 1.4 FOR STRAIGHT AND SAGITTAL ATTACHMENTS. ALL PIECES SHOWING SIGNS OF WEAR WITH DARK GREASE LIKE MATERIAL AROUND MOUTH OF MICS AND ATTACHMENTS. CUTS WERE 1-2 MM DEEP BUT STILL ACCEPTABLE FOR PRESSFIT FOR SURGEON. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16830989 ↗
  • MalfunctionApr 11, 2023Residue After Decontamination; Residue After Decontamination

    OIL LEAKING FROM ATTACHMENT. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16712587 ↗
  • MalfunctionApr 7, 2023Residue After Decontamination; Residue After Decontamination

    STRAIGHT SAW ATTACHMENT IS LEAKING OIL. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16702347 ↗
  • MalfunctionMar 20, 2023Residue After Decontamination; Residue After Decontamination

    BOTH SAWS LEAKING FLUID. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16574185 ↗
  • MalfunctionMar 20, 2023Residue After Decontamination; Residue After Decontamination

    THE SAW ATTACHMENT LEAKS OIL ON VARIOUS AREAS. ADDITIONAL INFORMATION: "THE LEAKAGE WAS NOTICED PRE-OP WHEN OPENING THE INSTRUMENTS. WE HAD THEM WASH IT AGAIN BUT IT WAS STILL THE SAME PROBLEM.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16574182 ↗
  • MalfunctionMar 7, 2023Residue After Decontamination; Residue After Decontamination

    STRAIGHT SAW ATTACHMENT HAS BECOME GREASY BLACK AFTER STERILIZATION PROCESS AND WASHING PROCESS, AS IF A SEAL HAS BEEN BROKEN. OIL AND WASH WATER LEAKED FROM THE INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16499227 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →