Mako Surgical Corp. · Right Angle Saw Attachment · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Mako right angle saw attachment is an orthopedic stereotaxic instrument. The 81 reports mostly describe locking mechanisms failing so the saw blade is loose, along with grease or residue after cleaning and occasional inaccurate cuts. There are 81 reports, one injury and no deaths. One recall started April 2019, now closed, was about ungreased bearings.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
81
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 00848486032128), received between Jun 11, 2018 and Dec 17, 2025 — classified as 1 injury, 80 malfunctions.

The most frequently reported specific device problem is “Residue After Decontamination” (24 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 25, 2019Terminated

    The bearings of the saw attachments were ungreased.

    Z-1823-2019 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Residue After Decontamination24
  • Detachment of Device or Device Component20
  • Mechanical Problem20
  • Non Reproducible Results10
  • Contamination of Device Ingredient or Reagent9
  • Premature Separation7
  • Device Slipped6
  • Noise, Audible6

Patient problems

  • No Known Impact Or Consequence To Patient46
  • No Clinical Signs, Symptoms or Conditions36
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Mechanical Problem; Application Program Problem: Parameter Calculation Error; Mechanical Problem; Application Program Problem: Parameter Calculation Error

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23825988 ↗
  • MalfunctionApr 21, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21874829 ↗
  • InjuryMar 26, 2025Difficult to Insert; Difficult to Insert

    LOCKING MECHANISM BROKE ON ANGLED SAW ATTACHMENT (CASE WAS CANCELLED). CASE TYPE / APPLICATION: TKA. IT WAS PREOP AND THE CASE WAS CANCELLED BECAUSE THE SAW ATTACHMENT LOCKING MECHANISM BROKE ONE THE SECOND SIDE OF A BILATERAL KNEE CASE AND THE DOCTOR DID NOT WANT TO WAIT AN HOUR AND A HALF FOR THE NEXT SAW ATTACHMENT TO BE PROCESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21699144 ↗
  • MalfunctionOct 25, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error

    MAKO ANGLED SAW ATTACHMENT LOCKING MECHANISM IS BROKEN. IT WILL NOT HOLD THE SAW IN PLACE AND AS A RESULT HAS ABOUT 2MM OF TOGGLE IN THE BLADE. THIS INTRODUCES 2MM OF INACCURACY DURING CUTS AND WAS EVIDENT DURING THE CASE. INITIALLY THE SURGEON ALIGNED THE SAW FOR THE DISTAL FEMORAL CUT AND STARTED THE SAW BLADE OFF THE BONE. BEFORE CUTTING THE BONE, THE SURGEON MADE A COMMENT THAT THE SAW SEEMED TO HAVE MORE VIBRATION THAN USUAL SO WE RE CHECKED THE SAW CHECKPOINT WITH THE GREEN PROBE AND IT PASSED. THE SURGEON PERFORMED BOTH ANGLED SAW CUTS AND PROCEEDED TO THE STRAIGHT ATTACHMENT CUTS. WE THEN TRAILED THE FEMUR AND IT WASN'T SITTING FLUSH. I GOT THE SURGEON TO CHECK HIS CUTS WITH THE PLANAR PROBE AND BOTH THE DISTAL AND POSTERIOR CHAMFER WERE OFF BY ABOUT 2MM. WE THEN SWITCHED THE SAW ATTACHMENT AND RECUT BOTH THAT WERE OFF. THE SURGEON NOTICED THE DIFFERENCE IN AMOUNT OF BONE THAT WAS CUT AND WAS SATISFIED WHEN HE RE CHECKED THE CUTS AND TRAILED THE FEMUR AGAIN. UPON EXAMINING THE OLD ATTACHMENT WITH THE SAW ATTACHED, THERE WAS VISIBLE ¿PLAY/TOGGLE¿ WHEN THE SAW WAS ¿LOCKED IN¿. MAKO TKA 2.0.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20540078 ↗
  • MalfunctionApr 12, 2024

    STRYKER MAKO ROBOT SAGGITAL SAW RIGHT ANGLE ATTACHMENT AND STRAIGHT SAGITTAL SAW ATTACHMENT 2.7 BREAKING AND UNABLE TO HOLD SAW BLADE SECURELY. REPAIRS AND REPLACEMENT REQUIRED FREQUENTLY. RISK TO PATIENT HARM DURING PROCEDURE. MFG IFU FOLLOWED FOR PROCESSING. REF REPORT MW5153825.

    Patient problems: Insufficient Information

    MAUDE report 19104558 ↗
  • MalfunctionMay 26, 2023Residue After Decontamination; Residue After Decontamination

    SPD SAYS THAT THE ATTACHMENTS ARE LEAKING LUBRICANT. THAT THEY STERILIZED THEM AND THAT WHEN OPENED IN THE ROOM THAT THERE WAS GREASE ALL OVER ALL THE INSTRUMENTS. THEY AREN'T SURE WHICH PART IS LEAKING THE LUBRICANT AND WANT TO GET NEW ONE. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17012442 ↗
  • MalfunctionApr 27, 2023Residue After Decontamination; Noise, Audible; Residue After Decontamination; Noise, Audible

    SAW AND ATTACHMENTS VERY LOUD. CHECKPOINTS AT THE EDGE OF RANGE AT 1.2 AND 1.4 FOR STRAIGHT AND SAGITTAL ATTACHMENTS. ALL PIECES SHOWING SIGNS OF WEAR WITH DARK GREASE LIKE MATERIAL AROUND MOUTH OF MICS AND ATTACHMENTS. CUTS WERE 1-2 MM DEEP BUT STILL ACCEPTABLE FOR PRESSFIT FOR SURGEON. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16827230 ↗
  • MalfunctionApr 11, 2023Positioning Failure; Incorrect, Inadequate or Imprecise Result or Readings; Positioning Failure; Incorrect, Inadequate or Imprecise Result or Readings

    DURING A RIGHT MAKO TKA, AFTER THE FIRST 2 CUTS WITH THE ANGLED SAW ATTACHMENT (DISTAL AND POSTERIOR CHAMFER), BOTH CUTS WERE CHECKED AND WERE BOTH 2-2.5MM DEEP. WE CONFIRMED THIS WITH THE X-ARRAY AS WELL AS THE GREEN PROBE AND BOTH SHOWED THE SAME DEEP RESULT. WHEN CUTTING THERE WAS NO YELLOW WARNING TO SUGGEST THE CUT WAS DEEP. ALSO, THE ANGLED SAW ATTACHMENT HAD BEEN CHECKED WITH THE SAWBLADE CHECKPOINT PRIOR TO AND AFTER THE CUTS AND WAS 0.4. THE FEMORAL ARRAY WAS ALSO CHECKED AND WAS 0.3. WHEN TRIALLING THE FEMORAL IMPLANT HAD A ~2MM GAP DISTALLY SO WE HAD TO USE CEMENTED IMPLANTS WHICH IS NOT THE SURGEON¿S PREFERENCE. ALL OF THE CUTS MADE WITH THE STRAIGHT SAW ATTACHMENT WERE WITHIN 0.5 OF THE PLANNED CUT. NOT AS PRECISE TKA WHICH IS WHY THE SURGEON USES MAKO. WE HAD TO USE A CEMENTED IMPLANT TO FILL THE 2MM GAP DUE TO THE DEEP DISTAL AND POSTERIOR CHAMFER CUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16712808 ↗
  • MalfunctionMar 20, 2023Residue After Decontamination; Residue After Decontamination

    BOTH SAWS LEAKING FLUID. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16574188 ↗
  • MalfunctionFeb 9, 2023Residue After Decontamination; Residue After Decontamination

    AFTER MULTIPLE WASHES, STERILIZATION AND DETAILED CLEANING, THERE IS PRESENT OIL OR FLUIDS COMING OUT OF SAW ATTACHMENT. ALSO BLACK RESIDUE LEFT ON LIP OF SAW ATTACHMENT THAT IS COMING OFF ON GLOVES BUT IS NOT COMING OFF WITH STERILIZATION. DISCOVERED DURING PRE-OP, NO PATIENT UNDER ANESTHESIA, NO DELAY. CASE TYPE / APPLICATION: PKA 3.0.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16342210 ↗
  • MalfunctionOct 18, 2022Residue After Decontamination; Residue After Decontamination

    OIL LEAKING FROM ANGLED SAW ATTACHMENT. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624188 ↗
  • MalfunctionOct 5, 2022Residue After Decontamination; Residue After Decontamination

    ANGLED ATTACHMENT IS LEAKING OIL. NO DELAY. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15543549 ↗
  • MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination

    ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14632666 ↗
  • MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination

    ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14632531 ↗
  • MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination

    ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14632528 ↗
  • MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination

    ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14632456 ↗
  • MalfunctionMay 6, 2022Residue After Decontamination; Noise, Audible; Intermittent Loss of Power; Residue After Decontamination; Noise, Audible; Intermittent Loss of Power

    THE FOLLOWING WAS REPORTED: MAKO MICS HANDPIECE BECAME LOUD DURING SURGERY. TOWARDS THE END OF THE LAST CUT, THE SAWING FUNCTION OF THE HANDPIECE FAILED. BLACK DEBRIS FORMED INSIDE THE HANDPIECE AND SAW BLADE ATTACHMENT. IN ADDITION, THE STRAIGHT SAW BLADE ATTACHMENT BECAME VERY HOT. LAST CUT HAD TO BE FINISHED WITH A MANUAL SAW. CASE TYPE: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14302815 ↗
  • MalfunctionApr 29, 2022Application Program Problem: Parameter Calculation Error; Incorrect, Inadequate or Imprecise Result or Readings; Application Program Problem: Parameter Calculation Error; Incorrect, Inadequate or Imprecise Result or Readings

    OUR DISTAL CUT WAS SHOWING 1.5 DEEP WITH PLANAR PROBE EVEN THOUGH NO RED ON SCREEN WAS SHOWING. WE HAVE HAD ISSUES IN LAST FEW WEEKS WITH CUTS BEING TOO DEEP. WE CHECKED ANOTHER CUT MADE WITH STRAIGHT ATTACHMENT AND IT WASN¿T DEEP. SUSPECT THAT ANGLED ATTACHMENT MAY BE GOING BAD. NO TOURNIQUET WAS USED. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14253970 ↗
  • MalfunctionApr 1, 2022Display or Visual Feedback Problem; Device Slipped; Display or Visual Feedback Problem; Device Slipped

    MICS SCREW AT BOTTOM OF GREY HANDLE FELL OUT DURING CUTTING. DISTAL CUT WAS CHECKED WITH PLANAR PROBE AND WAS 1 MM PROUD AFTER 3 RECUTS AND RED SHOWING THROUGHOUT CUT. SURGICAL DELAY = 15 MINUTES. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13975128 ↗
  • MalfunctionMar 9, 2022Residue After Decontamination; Residue After Decontamination

    OIL LEAKING FROM STRAIGHT AND ANGLED SAW ATTACHMENTS. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13709944 ↗
  • MalfunctionMar 4, 2022Residue After Decontamination; Residue After Decontamination

    ANGLED AND STRAIGHT SAW ATTACHMENTS LEAKING OIL. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13665963 ↗
  • MalfunctionFeb 25, 2022Residue After Decontamination; Residue After Decontamination

    SAW ATTACHMENTS WERE LEAKING OIL. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13604671 ↗
  • MalfunctionJan 4, 2022Residue After Decontamination; Residue After Decontamination

    OIL OR GREASE PRESENT AFTER WASHING AND STERILIZATION AT CONNECTION END OF SAW ATTACHMENTS. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13150552 ↗
  • MalfunctionDec 21, 2021Residue After Decontamination; Residue After Decontamination

    MICS AND ANGLED SAW ATTACHMENT HAD OIL ALL OVER IT. SPD THINKS A BALL BEARING MIGHT HAVE LEAKED. THEY WOULD LIKE IT REPLACED. LONG STABILIZER FUSED WITH THE PIN. SCRUB TECH CANNOT TAKE IT OFF. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13051873 ↗
  • MalfunctionOct 26, 2021Residue After Decontamination; Residue After Decontamination

    BOTH PARTS LEAKING OIL. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12698469 ↗

Other Orthopedic stereotaxic instrument devices