Mako Surgical Corp. · Right Angle Saw Attachment · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Mako right angle saw attachment is an orthopedic stereotaxic instrument. The 81 reports mostly describe locking mechanisms failing so the saw blade is loose, along with grease or residue after cleaning and occasional inaccurate cuts. There are 81 reports, one injury and no deaths. One recall started April 2019, now closed, was about ungreased bearings.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 81
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 00848486032128), received between Jun 11, 2018 and Dec 17, 2025 — classified as 1 injury, 80 malfunctions.
The most frequently reported specific device problem is “Residue After Decontamination” (24 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 25, 2019Terminated
The bearings of the saw attachments were ungreased.
Z-1823-2019 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Residue After Decontamination24
- Detachment of Device or Device Component20
- Mechanical Problem20
- Non Reproducible Results10
- Contamination of Device Ingredient or Reagent9
- Premature Separation7
- Device Slipped6
- Noise, Audible6
Patient problems
- No Known Impact Or Consequence To Patient46
- No Clinical Signs, Symptoms or Conditions36
- Insufficient Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 17, 2025Mechanical Problem; Application Program Problem: Parameter Calculation Error; Mechanical Problem; Application Program Problem: Parameter Calculation Error
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23825988 ↗MalfunctionApr 21, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21874829 ↗InjuryMar 26, 2025Difficult to Insert; Difficult to Insert
LOCKING MECHANISM BROKE ON ANGLED SAW ATTACHMENT (CASE WAS CANCELLED). CASE TYPE / APPLICATION: TKA. IT WAS PREOP AND THE CASE WAS CANCELLED BECAUSE THE SAW ATTACHMENT LOCKING MECHANISM BROKE ONE THE SECOND SIDE OF A BILATERAL KNEE CASE AND THE DOCTOR DID NOT WANT TO WAIT AN HOUR AND A HALF FOR THE NEXT SAW ATTACHMENT TO BE PROCESSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21699144 ↗MalfunctionOct 25, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
MAKO ANGLED SAW ATTACHMENT LOCKING MECHANISM IS BROKEN. IT WILL NOT HOLD THE SAW IN PLACE AND AS A RESULT HAS ABOUT 2MM OF TOGGLE IN THE BLADE. THIS INTRODUCES 2MM OF INACCURACY DURING CUTS AND WAS EVIDENT DURING THE CASE. INITIALLY THE SURGEON ALIGNED THE SAW FOR THE DISTAL FEMORAL CUT AND STARTED THE SAW BLADE OFF THE BONE. BEFORE CUTTING THE BONE, THE SURGEON MADE A COMMENT THAT THE SAW SEEMED TO HAVE MORE VIBRATION THAN USUAL SO WE RE CHECKED THE SAW CHECKPOINT WITH THE GREEN PROBE AND IT PASSED. THE SURGEON PERFORMED BOTH ANGLED SAW CUTS AND PROCEEDED TO THE STRAIGHT ATTACHMENT CUTS. WE THEN TRAILED THE FEMUR AND IT WASN'T SITTING FLUSH. I GOT THE SURGEON TO CHECK HIS CUTS WITH THE PLANAR PROBE AND BOTH THE DISTAL AND POSTERIOR CHAMFER WERE OFF BY ABOUT 2MM. WE THEN SWITCHED THE SAW ATTACHMENT AND RECUT BOTH THAT WERE OFF. THE SURGEON NOTICED THE DIFFERENCE IN AMOUNT OF BONE THAT WAS CUT AND WAS SATISFIED WHEN HE RE CHECKED THE CUTS AND TRAILED THE FEMUR AGAIN. UPON EXAMINING THE OLD ATTACHMENT WITH THE SAW ATTACHED, THERE WAS VISIBLE ¿PLAY/TOGGLE¿ WHEN THE SAW WAS ¿LOCKED IN¿. MAKO TKA 2.0.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20540078 ↗MalfunctionApr 12, 2024
STRYKER MAKO ROBOT SAGGITAL SAW RIGHT ANGLE ATTACHMENT AND STRAIGHT SAGITTAL SAW ATTACHMENT 2.7 BREAKING AND UNABLE TO HOLD SAW BLADE SECURELY. REPAIRS AND REPLACEMENT REQUIRED FREQUENTLY. RISK TO PATIENT HARM DURING PROCEDURE. MFG IFU FOLLOWED FOR PROCESSING. REF REPORT MW5153825.
Patient problems: Insufficient Information
MAUDE report 19104558 ↗MalfunctionMay 26, 2023Residue After Decontamination; Residue After Decontamination
SPD SAYS THAT THE ATTACHMENTS ARE LEAKING LUBRICANT. THAT THEY STERILIZED THEM AND THAT WHEN OPENED IN THE ROOM THAT THERE WAS GREASE ALL OVER ALL THE INSTRUMENTS. THEY AREN'T SURE WHICH PART IS LEAKING THE LUBRICANT AND WANT TO GET NEW ONE. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17012442 ↗MalfunctionApr 27, 2023Residue After Decontamination; Noise, Audible; Residue After Decontamination; Noise, Audible
SAW AND ATTACHMENTS VERY LOUD. CHECKPOINTS AT THE EDGE OF RANGE AT 1.2 AND 1.4 FOR STRAIGHT AND SAGITTAL ATTACHMENTS. ALL PIECES SHOWING SIGNS OF WEAR WITH DARK GREASE LIKE MATERIAL AROUND MOUTH OF MICS AND ATTACHMENTS. CUTS WERE 1-2 MM DEEP BUT STILL ACCEPTABLE FOR PRESSFIT FOR SURGEON. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16827230 ↗MalfunctionApr 11, 2023Positioning Failure; Incorrect, Inadequate or Imprecise Result or Readings; Positioning Failure; Incorrect, Inadequate or Imprecise Result or Readings
DURING A RIGHT MAKO TKA, AFTER THE FIRST 2 CUTS WITH THE ANGLED SAW ATTACHMENT (DISTAL AND POSTERIOR CHAMFER), BOTH CUTS WERE CHECKED AND WERE BOTH 2-2.5MM DEEP. WE CONFIRMED THIS WITH THE X-ARRAY AS WELL AS THE GREEN PROBE AND BOTH SHOWED THE SAME DEEP RESULT. WHEN CUTTING THERE WAS NO YELLOW WARNING TO SUGGEST THE CUT WAS DEEP. ALSO, THE ANGLED SAW ATTACHMENT HAD BEEN CHECKED WITH THE SAWBLADE CHECKPOINT PRIOR TO AND AFTER THE CUTS AND WAS 0.4. THE FEMORAL ARRAY WAS ALSO CHECKED AND WAS 0.3. WHEN TRIALLING THE FEMORAL IMPLANT HAD A ~2MM GAP DISTALLY SO WE HAD TO USE CEMENTED IMPLANTS WHICH IS NOT THE SURGEON¿S PREFERENCE. ALL OF THE CUTS MADE WITH THE STRAIGHT SAW ATTACHMENT WERE WITHIN 0.5 OF THE PLANNED CUT. NOT AS PRECISE TKA WHICH IS WHY THE SURGEON USES MAKO. WE HAD TO USE A CEMENTED IMPLANT TO FILL THE 2MM GAP DUE TO THE DEEP DISTAL AND POSTERIOR CHAMFER CUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16712808 ↗MalfunctionMar 20, 2023Residue After Decontamination; Residue After Decontamination
BOTH SAWS LEAKING FLUID. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16574188 ↗MalfunctionFeb 9, 2023Residue After Decontamination; Residue After Decontamination
AFTER MULTIPLE WASHES, STERILIZATION AND DETAILED CLEANING, THERE IS PRESENT OIL OR FLUIDS COMING OUT OF SAW ATTACHMENT. ALSO BLACK RESIDUE LEFT ON LIP OF SAW ATTACHMENT THAT IS COMING OFF ON GLOVES BUT IS NOT COMING OFF WITH STERILIZATION. DISCOVERED DURING PRE-OP, NO PATIENT UNDER ANESTHESIA, NO DELAY. CASE TYPE / APPLICATION: PKA 3.0.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16342210 ↗MalfunctionOct 18, 2022Residue After Decontamination; Residue After Decontamination
OIL LEAKING FROM ANGLED SAW ATTACHMENT. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15624188 ↗MalfunctionOct 5, 2022Residue After Decontamination; Residue After Decontamination
ANGLED ATTACHMENT IS LEAKING OIL. NO DELAY. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15543549 ↗MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination
ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632666 ↗MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination
ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632531 ↗MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination
ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632528 ↗MalfunctionJun 8, 2022Residue After Decontamination; Residue After Decontamination
ATTACHMENTS ALL FOUND TO HAVE GREASY RESIDUE LEFT EVEN AFTER DEEP CLEANING, STERILITY CONCERNS OF HOSPITAL REQUIRE REPLACEMENT OF ALL PIECES IN SETS. SURGICAL DELAY 16-30 MINUTES. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14632456 ↗MalfunctionMay 6, 2022Residue After Decontamination; Noise, Audible; Intermittent Loss of Power; Residue After Decontamination; Noise, Audible; Intermittent Loss of Power
THE FOLLOWING WAS REPORTED: MAKO MICS HANDPIECE BECAME LOUD DURING SURGERY. TOWARDS THE END OF THE LAST CUT, THE SAWING FUNCTION OF THE HANDPIECE FAILED. BLACK DEBRIS FORMED INSIDE THE HANDPIECE AND SAW BLADE ATTACHMENT. IN ADDITION, THE STRAIGHT SAW BLADE ATTACHMENT BECAME VERY HOT. LAST CUT HAD TO BE FINISHED WITH A MANUAL SAW. CASE TYPE: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14302815 ↗MalfunctionApr 29, 2022Application Program Problem: Parameter Calculation Error; Incorrect, Inadequate or Imprecise Result or Readings; Application Program Problem: Parameter Calculation Error; Incorrect, Inadequate or Imprecise Result or Readings
OUR DISTAL CUT WAS SHOWING 1.5 DEEP WITH PLANAR PROBE EVEN THOUGH NO RED ON SCREEN WAS SHOWING. WE HAVE HAD ISSUES IN LAST FEW WEEKS WITH CUTS BEING TOO DEEP. WE CHECKED ANOTHER CUT MADE WITH STRAIGHT ATTACHMENT AND IT WASN¿T DEEP. SUSPECT THAT ANGLED ATTACHMENT MAY BE GOING BAD. NO TOURNIQUET WAS USED. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14253970 ↗MalfunctionApr 1, 2022Display or Visual Feedback Problem; Device Slipped; Display or Visual Feedback Problem; Device Slipped
MICS SCREW AT BOTTOM OF GREY HANDLE FELL OUT DURING CUTTING. DISTAL CUT WAS CHECKED WITH PLANAR PROBE AND WAS 1 MM PROUD AFTER 3 RECUTS AND RED SHOWING THROUGHOUT CUT. SURGICAL DELAY = 15 MINUTES. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13975128 ↗MalfunctionMar 9, 2022Residue After Decontamination; Residue After Decontamination
OIL LEAKING FROM STRAIGHT AND ANGLED SAW ATTACHMENTS. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13709944 ↗MalfunctionMar 4, 2022Residue After Decontamination; Residue After Decontamination
ANGLED AND STRAIGHT SAW ATTACHMENTS LEAKING OIL. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13665963 ↗MalfunctionFeb 25, 2022Residue After Decontamination; Residue After Decontamination
SAW ATTACHMENTS WERE LEAKING OIL. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13604671 ↗MalfunctionJan 4, 2022Residue After Decontamination; Residue After Decontamination
OIL OR GREASE PRESENT AFTER WASHING AND STERILIZATION AT CONNECTION END OF SAW ATTACHMENTS. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13150552 ↗MalfunctionDec 21, 2021Residue After Decontamination; Residue After Decontamination
MICS AND ANGLED SAW ATTACHMENT HAD OIL ALL OVER IT. SPD THINKS A BALL BEARING MIGHT HAVE LEAKED. THEY WOULD LIKE IT REPLACED. LONG STABILIZER FUSED WITH THE PIN. SCRUB TECH CANNOT TAKE IT OFF. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13051873 ↗MalfunctionOct 26, 2021Residue After Decontamination; Residue After Decontamination
BOTH PARTS LEAKING OIL. PATIENT WAS NOT UNDER ANESTHESIA. CASE TYPE / APPLICATION: TKA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12698469 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page