AOS GALILEO™ LAG SCREW, RIGHT, 10.5mm x 90mm 1099-090Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00848665015836), received between Nov 27, 2019 and Apr 27, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Physical Resistance/Sticking1
Patient problems
- Insufficient Information2
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 27, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT AFTER THE INITIAL SURGERY ON (B)(6) 2026, A REVISION SURGERY WAS PERFORMED ON THE SAME DAY BECAUSE A LEFT TROCHANTERIC NAIL HAD BEEN MISTAKENLY IMPLANTED INSTEAD OF THE RIGHT TROCHANTERIC NAIL. THE REVISION SURGERY WAS FINISHED SUCCESSFULLY USING THE RIGHT TROCHANTERIC NAIL. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE. UPDATE 15-APR-2026 - FURTHER INFORMATION WAS RECEIVED: THE DISTRIBUTOR REPORTED THAT THE LABEL OF THE DEVICE WAS CORRECT. THE INCIDENT WAS CAUSED BY AN ERROR AT THE HOSPITAL. NO X-RAYS ARE AVAILABLE. THE INITIAL SURGERY AND THE REVISION SURGERY WERE PERFORMED IN THE SAME HOSPITAL THROUGH THE SAME SURGEON.
Patient problems: Insufficient Information
MAUDE report 25001904 ↗MalfunctionJul 25, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON 04/01/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS(B)(4) THAT AN 0857-100 GALILEO EXTRACTION TOOL WAS DAMAGED DURING THE CASE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM. ON 07/10/2024, DURING DEVICE EVALUATION OF THE RETURNED DEVICE, IT WAS FOUND THAT THE 0857-100 WAS STUCK INSIDE A 1099-090 GALILEO LAG SCREW, RIGHT, Ø10.5MMX90MM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19831052 ↗MalfunctionAug 17, 2022Break
IT WAS REPORTED ON (B)(6)2022 BY A FACILITY REPRESENTATIVE VIA PHONE THAT AN 1099-090 GALILEO SCREW LOCKING O-RING BROKE-OFF. THIS WAS DISCOVERED DURING A CASE AND BROKE INSIDE THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 15245108 ↗InjuryNov 27, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2019, AOS RECEIVED AN IMAGE OF A BROKEN GALILEO LAG SCREW, P/N 1099-100. THE LAG SCREW HAD BROKEN INSIDE OF THE PATIENT APPROXIMATELY 5 MONTHS SINCE THE INITIAL SURGERY THAT TOOK PLACE ON (B)(6) 2019. THE REASON FOR INITIAL SURGERY IS UNKNOWN AS NO X-RAYS FROM INITIAL SURGERY HAVE BEEN PROVIDED. THERE HAS BEEN NO FURTHER COMMUNICATION REGARDING THIS COMPLAINT DESPITE MULTIPLE REQUESTS TO THE SALES REPRESENTATIVE. BECAUSE OF THIS, A ROOT CAUSE FOR THIS BREAK IS UNKNOWN. IT CAN BE ASSUMED THAT THE IMPLANT BROKE FROM ONE OF TWO POSSIBLE REASONS. ONE OPTION IS A SECONDARY TRAUMA THAT CAUSED BOTH THE BREAKAGE OF THE FEMUR AND THE FAILURE OF THE IMPLANT. THE SECOND OPTION IS THAT THE FRACTURE DID NOT HEAL DUE TO THE LACK OF REDUCTION IN THE INITIAL FRACTURE. THE IMPLANT IS NOT DESIGNED AS A LOAD-BEARING DEVICE, BUT LOAD-SHARING; ONLY TO REGAIN STRUCTURE AND AID IN THE HEALING PROCESS. WITH IMPROPER REDUCTION, THERE WOULD BE NO BONE GROWTH TO SUPPORT THE LAG SCREW, THEREFORE THE LAG SCREW COULD FRACTURE. WITH NO ADDITIONAL INFORMATION OTHER THAN AN X-RAY, NO DEFINITE CONCLUSION CAN BE MADE. HOWEVER, A FAILURE AT 5 MONTHS POST-OPERATION IS NOT INDICATIVE OF AN ISSUE WITH THE IMPL
Patient problems: Failure of Implant
MAUDE report 9387388 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page