GALILEO™ LAG SCREW, RIGHT, 10.5mm x 95mm 1099-095Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00848665015843), received between Aug 18, 2023 and May 15, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Physical Resistance/Sticking2
- Fracture1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Protrusion/Extrusion1
- Positioning Failure1
Patient problems
- Insufficient Information4
- Failure of Implant2
- Foreign Body In Patient2
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryMay 15, 2026Break
IT WAS REPORTED THAT A FEW DAYS AFTER THE IMPLANTATION OF A TROCHANTERIC NAIL, IT WAS DISCOVERED THAT THE LAG SCREW HAD BROKEN INTO THE JOINT. THERE WAS NO SPECIFIC INCIDENT THAT HAD LED TO THE FRACTURE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION WAS RECEIVED. UPDATE 08-MAY-2026 - FURTHER INFORMATION WAS RECEIVED: IT WAS REPORTED THAT THE INITIAL SURGERY WAS PERFORMED ON (B)(6) 2026, AND THE REVISION SURGERY ON (B)(6) 2026. BOTH SURGERIES WERE PERFORMED BY THE SAME SURGEON. THE PATIENT HAS NO FURTHER SYMPTOMS.
Patient problems: Foreign Body In Patient
MAUDE report 25191213 ↗MalfunctionMay 15, 2026Fracture
IT WAS REPORTED THAT A PATIENT EXPERIENCED PAIN A FEW DAYS AFTER THE IMPLANTATION OF A TROCHANTERIC NAIL ON (B)(6) 2026. AN X-RAY WAS SUBSEQUENTLY TAKEN, WHICH SHOWED THAT THE LAG SCREW AND THE ANTI-ROTATION SCREW HAD FRACTURED INTO THE JOINT. THERE WAS NO SPECIFIC INCIDENT THAT HAD LED TO THE FRACTURE. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: Pain
MAUDE report 25187053 ↗InjuryOct 24, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON 09/03/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT A 5033-100 CAPTURING ROD ASSEMBLY WAS COLD-WELDED WITH THE SCREW. THIS OCCURRED DURING A REVISION TROCH NAIL PROCEDURE WHEN THE SURGEON REMOVED OUR TROCH NAIL AND IMPLANTED A SYNTHES NAIL WHEN THE GALILEO CAPTURING ROD COLD WELDED TO THE SCREW UPON REMOVAL.
Patient problems: Insufficient Information
MAUDE report 20522252 ↗MalfunctionJul 23, 2024Break; Break
ON 7/1/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 1099-095 TELESCOPING LAG SCREW LOCKING MECHANISM BROKE WHEN TRYING TO LOCK IT. ALL BROKEN FRAGMENTS WERE REMOVED. THE CASE WAS COMPLETED USING A 1099-090 TELESCOPING LAG SCREW. THIS WAS DISCOVERED DURING A TROCHANTERIC NAIL PROCEDURE ON (B)(6) 2024. THE CASE WAS NOT DELAYED, AND ADDITIONAL ANESTHESIA WAS NOT NEEDED.
Patient problems: Insufficient Information
MAUDE report 19805941 ↗MalfunctionDec 22, 2023Material Protrusion/Extrusion; Material Protrusion/Extrusion
ON 11/27/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1099-095 TELESCOPING LAG SCREW RING WAS PROMINENT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18385469 ↗MalfunctionNov 22, 2023Positioning Failure; Loosening of Implant Not Related to Bone-Ingrowth; Positioning Failure; Loosening of Implant Not Related to Bone-Ingrowth
ON 10/23/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT A TROCHANTERIC NAIL PROCEDURE ON 6/4/2023. AFTER THE SURGERY, AN X-RAY REVEALED THAT THE 1099-095 LAG SCREW SHIFTED LATERALLY. THE PATIENT IS NOT IN ANY DISCOMFORT OR PAIN, AND THE IMPLANTS REMAIN IN PLACE. CAUSE FOR THE SHIFT IN THE LAG SCREW IS UNKNOWN.
Patient problems: Failure of Implant; Foreign Body In Patient
MAUDE report 18190910 ↗InjuryNov 1, 2023Break; Break
ON 10/5/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT SURGERY DUE TO AN INTERTROCHANTERIC FRACTURE ON (B)(6) 2023. ON (B)(6) 2023, THE SALES REPRESENTATIVE WAS MADE AWARE VIA PHONE THAT THE LAG SCREW IMPLANTED IN THE PATIENT BROKE. THE PATIENT WILL UNDERGO REVISION SURGERY. NO FURTHER INFORMATION HAS BEEN PROVIDED. ADDITIONAL INFORMATION RECEIVED ON 10/13/2023: ON (B)(6) 2023, THE PATIENT UNDERWENT SURGERY IN WHICH A 1023-395 TROCHANTERIC NAIL AND A 1099-095 LAG SCREW WERE IMPLANTED. ON (B)(6) 2023, THE PATIENT FELT PAIN WHILE WALKING. A REVISION SURGERY TOOK PLACE ON (B)(6) 2023 TO REMOVE THE TROCH NAIL AND LAG SCREW. DURING THE REMOVAL OF THE LAG SCREW, IT WAS NOTICED THAT IT HAD SCRATCHES DISTAL TO THE THREADS. THE CASE WAS COMPLETED AS A TOTAL HIP ARTHROPLASTY PROCEDURE.
Patient problems: Failure of Implant
MAUDE report 18046633 ↗MalfunctionAug 18, 2023Physical Resistance/Sticking; Physical Resistance/Sticking
ON 7/20/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 1099-095 TELESCOPING LAG SCREW WAS BEING REMOVED, AND AS THE SURGEON WAS BACKING THE SCREW OUT, THE LOCKING RING GOT CAUGHT ON THE LATERAL CORTEX. THE SURGEON HAD TO RIP THE LAG SCREW THROUGH THE LOCKING RING, WHICH WAS EMBEDDED IN THE CORTEX. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Insufficient Information
MAUDE report 17577011 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page