Skip to content

GALILEO® LOCKING TOOL 5031-000Recalls & reported problems

Advanced Orthopaedic Solutions, Inc. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00848665031522), received between May 22, 2024 and May 11, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure3
  • Defective Device2
  • Unintended Movement2
  • Activation Failure1
  • Break1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Break

    ON 22-APR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN 5031-000 GALILEO TOOL IS BROKEN. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135366 ↗
  • MalfunctionMay 14, 2025Defective Device; Defective Device

    ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017957 ↗
  • MalfunctionMay 14, 2025Defective Device; Defective Device

    ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017947 ↗
  • MalfunctionJan 6, 2025Positioning Failure; Activation Failure; Positioning Failure; Activation Failure

    ADDITIONAL INFORMATION WAS RECEIVED ON 01/16/2025: THIS WAS DISCOVERED DURING A HIP FRACTURE PROCEDURE ON (B)(6) 2024. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21085199 ↗
  • MalfunctionJan 3, 2025Positioning Failure; Use of Device Problem; Positioning Failure; Use of Device Problem

    ON 12/06/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN 5031-000 LOCKING TOOL AND 5030-000 LAG INSERTER WERE NOT ABLE TO WORK TOGETHER IN UNISON. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21066923 ↗
  • MalfunctionOct 16, 2024Positioning Failure; Positioning Failure

    ON 10/07/2024 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT A 5030-000 GALILEO LAG SCREW AND A 5031-000 GALILEO LOCKING TOOL DISENGAGED WHEN COMPRESSION WAS APPLIED RESULTING IN THE INSTRUMENTS NOT LOCKING. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20461895 ↗
  • MalfunctionMay 24, 2024Unintended Movement; Unintended Movement

    ON 04/08/2022, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A 5031-000 LOCKING TOOL IS NOT LOCKING. THIS OCCURRED DURING THE LAST 2 TROCH NAIL CASES WHEN THE BLACK T-HANDLE SEEMED TO BE ¿JUMPING¿ WHEN WE WOULD GO TO LOCK THE LAG SCREW INTO THE NAIL. IT WAS A CLEAR DIFFERENCE FROM THE ACTUAL TORQUE LIMITER ENGAGING. THERE WAS NO PATIENT EFFECT REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19392146 ↗
  • MalfunctionMay 22, 2024Unintended Movement; Unintended Movement

    ON 12/30/2021, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT A 5 031-100 LAG SCREW LOCKING TOOL MAY HAVE BEEN IN QUESTION FOR RELIABILITY AND PERFORMANCE OF THE TORQUE LIMITER WHEN IT WAS USED TO LOCK A 1099-095 TELESCOPING LAG SCREW. THIS WAS DISCOVERED POST-OPERATIVE OF A TROCHANTERIC FEMORAL NAIL PROCEDURE THAT WAS PERFORMED ON (B)(6)2021. WHEN THE SURGEON X-RAYED THE PATIENT, IT WAS NOTICES THAT THE TELESCOPING LAG SCREW DID NOT STAY LOCKED UP. PRIOR TO THIS, THE TROCHANTERIC FEMORAL NAIL PROCEDURE HAD BEEN SUCCESSFULLY COMPLETED WITH NO ISSUES, NO BROKEN IMPLANTS AND OR INSTRUMENTS. AT THIS TIME THERE HAS BEEN NO REVISION SURGERY HAS BEEN SCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19375823 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates