COMPRESSION SLEEVE 5032-000Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00848665036480), received between Jun 19, 2024 and May 14, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device2
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2025Defective Device; Defective Device
ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22017965 ↗MalfunctionMay 14, 2025Defective Device; Defective Device
ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22017963 ↗MalfunctionJun 19, 2024Material Fragmentation; Material Fragmentation
ON 10/25/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 5032-000 COMPRESSION SLEEVE'S RUBBER "O" RING MADE ITS WAY OUT WHEN THE "LEFTY" LAG SCREW WAS INSERTED. THIS OCCURRED DURING A PROCEDURE ON (B)(6) 2021. SURGEON WAS ABLE TO COMPLETE PROCEDURE WITHOUT THE "O" RING AND ASSURED THE LAG SCREW WAS INSERTED WITH GREAT COMPRESSION FUNCTION. DURING INSPECTION OF THE COMPRESSION SLEEVE AFTER THE CASE, IT WAS DETERMINE THAT THE "O" RING APPEARED TO BE CUT IN TWO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567592 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page