Skip to content

COMPRESSION SLEEVE 5032-000Recalls & reported problems

Advanced Orthopaedic Solutions, Inc. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00848665036480), received between Jun 19, 2024 and May 14, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Defective Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device2
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2025Defective Device; Defective Device

    ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017965 ↗
  • MalfunctionMay 14, 2025Defective Device; Defective Device

    ADDITIONAL INFORMATION WAS RECEIVED ON 05/07/2025: DURING A TROCHANTERIC NAIL PROCEDURE ON AN UNKNOWN DATE, THE COMPRESSION WHEEL AND INSERTER HANDLE FAILED TO PROVIDE PROPER COMPRESSION, EXHIBITING SLIPPING OR CLICKING AT A CERTAIN POINT WHEN ATTEMPTING TO COMPRESS. THIS ISSUE HAD PREVIOUSLY BEEN OBSERVED IN THREE CASES INVOLVING TROCH NAILS, THOUGH SPECIFIC DATES OR CASE DETAILS ARE UNAVAILABLE. WHILE THE ISSUE DID NOT IMPACT THE CASES OR PATIENTS, A RECURRING PATTERN WAS LATER IDENTIFIED. THE CASE WAS COMPLETED IN STANDARD FASHION, ACHIEVING AS MUCH COMPRESSION AS THE INSTRUMENT ALLOWED WITHOUT REQUIRING ADDITIONAL PRODUCTS. THERE WAS NO REPORTED DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE AFFECTED PRODUCTS, IN THIS CASE, INCLUDED THE LAG SCREW INSERTER AND COMPRESSION WHEEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017963 ↗
  • MalfunctionJun 19, 2024Material Fragmentation; Material Fragmentation

    ON 10/25/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 5032-000 COMPRESSION SLEEVE'S RUBBER "O" RING MADE ITS WAY OUT WHEN THE "LEFTY" LAG SCREW WAS INSERTED. THIS OCCURRED DURING A PROCEDURE ON (B)(6) 2021. SURGEON WAS ABLE TO COMPLETE PROCEDURE WITHOUT THE "O" RING AND ASSURED THE LAG SCREW WAS INSERTED WITH GREAT COMPRESSION FUNCTION. DURING INSPECTION OF THE COMPRESSION SLEEVE AFTER THE CASE, IT WAS DETERMINE THAT THE "O" RING APPEARED TO BE CUT IN TWO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19567592 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page