GALILEO® LEFTY TIGHTY™ CRT²™ LAG SCREW, LEFT, 10.5mm x 90mm 1192-090Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00848665036534), received between Aug 10, 2022 and Feb 3, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Dislodged or Dislocated2
- Damaged Thread1
- Failure to Advance1
Patient problems
- Insufficient Information4
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2025Failure to Advance; Failure to Advance
ON 01/07/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1192-090 LEFT LAG SCREW WOULD NOT ADVANCE THROUGH THE NAIL AND WAS SPINNING INSTEAD. THE SURGEON TRIED MULTIPLE TIMES TO REDRILL AND REINSERT THE SCREW WITHOUT SUCCESS AND DEEMED IT NOT FUNCTIONAL. NO PIECES BROKE OFF INSIDE THE PATIENT. THE SURGEON PROCEEDED TO IMPLANT A STANDARD RIGHT-THREADED LAG SCREW WHICH WAS IMPLANTED SUCCESSFULLY. THIS WAS DISCOVERED DURING A HIP FRACTURE PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON 1/20/2025: THE CASE WAS COMPLETED USING A 1099-090 TELESCOPING LAG SCREW. THE CASE WAS NOT DELAYED, AND ADDITIONAL ANESTHESIA WAS NOT NEEDED. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293645 ↗MalfunctionAug 21, 2024Device Dislodged or Dislocated; Damaged Thread; Device Dislodged or Dislocated; Damaged Thread
ON 07/19/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1192-090 LEFTY TELESCOPING LAG SCREW UNTHREADED INSIDE THE PATIENT. THE THREADS BROKE, AND THE SCREW WAS LEFT IMPLANTED. THIS WAS DISCOVERED DURING A HIP NAIL PROCEDURE ON (B)(6) 2024.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 20017672 ↗MalfunctionJun 4, 2024Break; Break
ON 6/1/2022 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A LOCKING NUT ON THE SCREW CAPTURING ROD SNAPPED OFF. WHILE TRYING TO REMOVE THE SCREW, THE LOCKING MECHANISM WAS DAMAGED. THIS WAS DISCOVERED DURING A HIP FRACTURE REPAIR ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON 6/2/2022 OVER THE PHONE: THIS WAS DISCOVERED DURING A HIP FRACTURE PROCEDURE AND NOTHING BROKE INSIDE THE PATIENT. ANOTHER TRAY WAS OPENED AND CASE WAS COMPLETED WITHOUT FURTHER ISSUES.
Patient problems: Insufficient Information
MAUDE report 19460563 ↗InjuryAug 18, 2022Device Dislodged or Dislocated
IT WAS REPORTED ON (B)(6) 2022 BY A ARTHREX EMPLOYEE VIA SEMS THAT AN 1192-090 LAG SCREW BACKED OUT. THIS WAS DISCOVERED DURING AFTER THE SCREW WAS IMPLANTED FOR AN UNSPECIFIED TIME, A REVISION SURGERY IS PLANNED. REVISION DATE (B)(6) 2022.
Patient problems: Insufficient Information
MAUDE report 15250098 ↗MalfunctionAug 10, 2022Break
ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1192-090 LEFTY TIGHT LAG SCREW BROKE. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2021. WHEN THE SURGEON ATTEMPTED TO ADVANCE THE SCREW, IT BROKE INSIDE THE PATIENT. THE SCREW WAS RETRIEVED AND WAS REPLACED WITH AN 1099-090 GALILEO LAG SCREW.
Patient problems: Insufficient Information
MAUDE report 15200736 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page