Skip to content

Percussionaire P5-TEERecalls & reported problems

Percussionaire Corporation · PHASITRON 5 INLINE VALVE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00849436000723), received between Jul 9, 2025 and Jul 9, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 18, 2026Class IIOpen, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3320-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Sep 17, 2025Class IOpen, Classified

    IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.

    Z-0031-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Gas/Air Leak1

Patient problems

  • Respiratory Insufficiency1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2025Gas/Air Leak; Gas/Air Leak

    ON JUNE 10, 2025, SENTEC WAS NOTIFIED OF AN INCIDENT INVOLVING FOUR NICU PATIENTS WHO EXHIBITED SIGNS OF RESPIRATORY DESCREPENCIES WHILE RECEIVING IPV THERAPY USING THE IN-LINE VALVE (P5-TEE). IT WAS OBSERVED THAT THE PATIENTS WERE RECEIVING ONLY 40-50% OF THE EXHALED VOLUME SET BY THE VENTILATOR, RATHER THAN THE EXPECTED FULL RETURNED VOLUME. THIS DISCREPANCY LED TO THE RETENTION OF CARBON DIOXIDE (CO2) IN THE PATIENTS. THE UNDERLYING CAUSE OF THE CO2 RETENTION WAS NOT INITIALLY KNOWN. AS PART OF TYPICAL IN-LINE VALVE USE, THE PHYSICIAN CONFIRMED THE PRESSURE RELIEF VALVE WAS FULLY CLOSED FOR ALL PATIENT CASES. IN AN EFFORT TO IDENTIFY THE SOURCE OF THE ISSUE, ONE PATIENT WAS REINTUBATED TO RULE OUT COMPLICATIONS RELATED TO THE ENDOTRACHEAL TUBE. THE CO2 RETENTION PERSISTED UNTIL THE IN-LINE VALVE WAS REMOVED, AT WHICH POINT THE PATIENT'S EXHALED TIDAL VOLUMES IMMEDIATELY IMPROVED. DUE TO THE NEED FOR MEDICAL INTERVENTION IN RESPONSE TO THIS EVENT, SENTEC IS SUBMITTING THIS REPORT.

    Patient problems: Respiratory Insufficiency

    MAUDE report 22456912 ↗

FDA clearance and classification

The FDA regulates product code BZO (set, tubing and support, ventilator (w harness)) under 21 CFR 868.5975 as a Class I device. 21 CFR 868.5975 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Percussionaire P5-TEE

Has the Percussionaire P5-TEE been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-3320-2026 (Class II), began on Aug 18, 2026: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Percussionaire P5-TEE?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jul 9, 2025. That report lists device problem “Gas/Air Leak” (1); patient problem “Respiratory Insufficiency” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Percussionaire P5-TEE?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Percussionaire P5-TEE FDA approved?

Devices of this type (product code BZO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Percussionaire Corporation.

Who makes the Percussionaire P5-TEE, and is it still sold?

The FDA's device database lists Percussionaire Corporation as the company responsible for it (brand Percussionaire, model P5-TEE). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Set, Tubing And Support, Ventilator (W Harness) devices

All Set, Tubing And Support, Ventilator (W Harness) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page