Breathing circuit set, coaxial 260168Recalls & reported problems

Hamilton Medical AG · Coaxial breathing circuit set with flow sensor, expiratory valve, elbow adapter and expandable limb connection for use with HAMILTON-C1/T1/MR1, 480 cm, adult/ pediatric, single-use · Class 1

All models of Breathing circuit set, coaxial (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630002802987), received between Mar 30, 2026 and Jun 2, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Obstruction of Flow” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2026Class IOpen, Classified

    Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

    Z-2463-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Obstruction of Flow4
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026Obstruction of Flow

    IT WAS REPORTED TO HAMILTON MEDICAL AG, THAT: ON 13TH MAY 2026 THE HAMILTON-T1 VENTILATOR (PN 161006 / SN (B)(6) AND BREATHING CIRCUIT SET (PN UNKNOWN / LOT UNKNOWN) PASSED PRE-USE CHECK, BUT WHEN CONNECTED TO THE PATIENT DISPLAYS "EXHALATION OBSTRUCTION" ALARM AND CANNOT VENTILATE. PATIENT INVOLVEMENT WITH MEDICAL INTERVENTION (NATURAL OF MEDICAL INTERVENTION UNKNOWN) WAS REPORTED. HOWEVER, NO PATIENT, USER OR THIRD PARTY HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25362053 ↗
  • MalfunctionJun 2, 2026Obstruction of Flow

    IT WAS REPORTED TO HAMILTON MEDICAL AG, THAT: ON 13TH MAY 2026 THE HAMILTON-T1 VENTILATOR (PN 1610060 / SN (B)(6) AND BREATHING CIRCUIT SET (PN UNKNOWN / LOT UNKNOWN) PASSED PRE-USE CHECK, BUT WHEN CONNECTED TO THE PATIENT DISPLAYS "EXHALATION OBSTRUCTION" ALARM AND CANNOT VENTILATE. PATIENT INVOLVEMENT WITH MEDICAL INTERVENTION (NATURAL OF MEDICAL INTERVENTION UNKNOWN) WAS REPORTED. HOWEVER, NO PATIENT, USER OR THIRD PARTY HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25360342 ↗
  • MalfunctionJun 2, 2026Obstruction of Flow

    IT WAS REPORTED TO HAMILTON MEDICAL AG, THAT: ON (B)(6) 2026 THE HAMILTON-T1 VENTILATOR (PN 161006 / SN (B)(6) AND BREATHING CIRCUIT SET (PN UNKNOWN / LOT UNKNOWN) PASSED PRE-USE CHECK, BUT WHEN CONNECTED TO THE PATIENT DISPLAYS "EXHALATION OBSTRUCTION" ALARM AND CANNOT VENTILATE. PATIENT INVOLVEMENT WITH MEDICAL INTERVENTION (NATURAL OF MEDICAL INTERVENTION UNKNOWN) WAS REPORTED. HOWEVER, NO PATIENT, USER OR THIRD PARTY HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25359469 ↗
  • MalfunctionMar 30, 2026Obstruction of Flow; Physical Resistance/Sticking

    HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "VENTILATOR PASSED PRE-OP CHECKS, PUT ON PATIENT...ALARM EVIDENT "EXHALATION OBSTRUCTION - LOW MINUTE VOLUME". STAFF CHANGED PATIENT TO DIFFERENT VENT (DRAGER V800)...PATIENT VENTILATED OK. RETRIEVED SPARE T1 VENTILATOR, PASSED PRE-OP, TRANSFERRED PATIENT TO THIS T1 OK. STAFF CHECKED FIRST T1 (27551) AGAIN, PUT ON A TEST LUNG...INITIALLY DIDN'T VENTILATE PROPERLY BUT RESOLVED AFTER 5 MINUTES AND VENTILATED OK." NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728855 ↗

Other Set, Tubing And Support, Ventilator (W Harness) devices

All Set, Tubing And Support, Ventilator (W Harness) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page