Tack Endovascular System (6F, 4.0-8.0mm) 206135061Recalls & reported problems
Intact Vascular, Inc. · 135cm Catheter Length · Class 3 · Implant
All models of Tack Endovascular System (6F, 4.0-8.0mm) (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00850003494036), received between Jun 23, 2021 and Nov 4, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Device Damaged by Another Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Damaged by Another Device2
- Ejection Problem2
- Unintended Collision2
- Material Separation1
- Positioning Failure1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Foreign Body In Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2024Ejection Problem; Ejection Problem
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED TO TREAT THE DISTAL SFA. THE FIRST FIVE TACKS WERE DEPLOYED SUCCESSFULLY. THE SIXTH TACK WAS NOT INTENDED TO BE DEPLOYED; HOWEVER, THE PATIENT MOVED HER LEG, AND WAS INADVERTENTLY DEPLOYED IN THE COMMON FEMORAL ARTERY. AN ATTEMPT WAS MADE TO RETRIEVE THE TACK BY PUTTING THE SHEATH OVER IT BUT COULD NOT BE RETRIEVED. A BALLOON WAS USED TO POST DILATE AND TO ENSURE PROPER WALL APPOSITION OF THE SIXTH TACK. THIS ADVERSE EVENT AND PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE THE SIXTH TACK WAS INADVERTENTLY DEPLOYED, REQUIRING ADDITIONAL INTERVENTION, BUT RETAINED IN PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20599624 ↗InjuryJul 8, 2024Positioning Failure; Positioning Failure
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED TO TREAT THE SFA LESION. THREE TACKS WERE DEPLOYED INSIDE THE PATIENT; HOWEVER, ONE TACK WAS SLIGHTLY TILTED. IN ORDER TO ACHIEVE THE DESIRED OUTCOME, THE PHYSICIAN ELECTED TO STENT THE MOST DISTAL TACK, EVEN THOUGH LEAVING IT SLIGHTLY TILTED IN PLACE WAS NOT EXPECTED TO CAUSE ANY RISK OR HARM TO THE PATIENT. NO PATIENT INJURY REPORTED. THIS ADVERSE EVENT AND PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY DUE TO THE TACK POSITION; RESULTING IN ELECTED STENT PLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19696745 ↗MalfunctionJul 3, 2024Positioning Problem; Positioning Problem
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE IN THE DISTAL COMMON FEMORAL ARTERY. DURING USE, A TACK WAS DEPLOYED IN THE ARTERY, REMOVED THE DEVICE, AND PERFORMED ANGIO. THE PHYSICIAN WANTED TO PLACE A SECOND TACK PROXIMAL TO THE ORIGINAL TACK LOCATION; HOWEVER, DID NOT POST DILATE THE FIRST TACK AND PUSHED THE TACK FURTHER DOWN THE ARTERY. THE SYSTEM WAS REMOVED WITHOUT DEPLOYING THE SECOND TACK, AND A BALLOON WAS USED TO POST DILATE THE TACK. THERE WAS NO ADDITIONAL INTERVENTION REQUIRED AND NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE THE TACK MIGRATED. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19667677 ↗MalfunctionJul 3, 2024Ejection Problem; Ejection Problem
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, ONE TACK WAS DEPLOYED SUCCESSFULLY IN THE POPLITEAL ARTERY. THE PHYSICIAN PLANNED TO DEPLOY ADDITIONAL TACKS IN THE OTHER AREAS OF THE SFA, WHICH REQUIRED THE DEVICE TO BE REMOVED FROM THE PATIENT IN ORDER TO REPOSITION. AS THE DEVICE WAS BEING PULLED OVER THE WIRE OUTSIDE OF THE PATIENT'S BODY, THE OUTER SHEATH RETRACTED, AND UNINTENTIONALLY DEPLOYED THE REMAINING 5 TACKS. THE PROCEDURE WAS COMPLETED USING A REGULAR STENT TO TREAT THE SFA. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE THE TACK INADVERTENTLY DEPLOYED. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19665820 ↗MalfunctionNov 17, 2022Material Separation; Material Separation
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED TO TREAT THE SLIGHTLY TORTUOUS AND MODERATELY CALCIFIED PERIPHERAL LESIONS. FOUR TACKS WERE DEPLOYED SUCCESSFULLY. DURING REMOVAL, IT WAS NOTED THAT THE DISTAL PORTION OF THE INNER CORE SEPARATED INSIDE THE INTRODUCER SHEATH. THE INTRODUCER SHEATH WAS REMOVED WITH THE SEPARATED TES CONTAINED WITHIN THE SHEATH. IMAGING CONFIRMED NO PIECE WAS RETAINED INSIDE THE BODY. THE SAME INTRODUCER SHEATH WAS PLACED BACK INSIDE THE PATIENT TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO THE SHAFT SEPARATION. THERE IS A POTENTIAL FOR HARM IF THE MALFUNCTION WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15813490 ↗MalfunctionNov 23, 2021Unintended Collision; Device Damaged by Another Device; Unintended Collision; Device Damaged by Another Device
FIVE 4-8MM TACKS WERE PLACED IN PROXIMAL/MID SFA AROUND (B)(6) 2021. PATIENT HAD BEEN ON DUAL ANTIPLATET THERAPY FROM PRIOR PROCEDURE. INITIAL ANGIOGRAM SHOWED PATENT TACKS AND SFA, WITH A LONG DISSECTION PRESENT IN DISTAL SFA/PROXIMAL POPLITEAL. WE THEN USED 0.014 IVUS TO EVALUATE DISSECTIONS, VESSEL SIZE, DISEASE BURDEN/MORPHOLOGY, IN POPLITEAL, AS PHYSICIAN PRE-PROCEDURE HAD PLANNED ON STENTING WITH SUPERA STENT IN POP AND ELUVIA IN SFA IF NEEDED. IVUS SHOWED SOFT ECCENTRIC PLAQUE IN POPLITEAL AROUND 80% STENOSIS. THERE WAS ALSO A LONG DISSECTION PLANE FROM PROXIMAL POPLITEAL INTO DISTAL SFA, AS WELL AS SOME OTHER AREAS OF DISSECTION THROUGHOUT THE SFA. PHYSICIAN WANTED TO USE HAWK ATHERECTOMY FOR POPLITEAL. ADVISED PHYSICIAN TO PROCEED WITH CAUTION THROUGH TACKS SINCE SOMEWHAT RECENT PLACEMENT, AND THAT THE HAWK MAY BE TO "BULKY" TO PASS THROUGH TWO OF THE PROXIMAL TACKS THAT WEREN'T FULLY EXPANDED DUE TO SOME RESIDUAL DISEASE. SPIDER FILTER/WIRE WAS PLACED INTO THE DISTAL POPLITEAL AND THEN HAWK INSERTED. AT ONE POINT PHYSICIAN HAD SOME DIFFICULTY PASSING HAWK IN SFA. AFTER THREE PASSES WITH THE HAWK, IT WAS REMOVED, AND PHYSICIAN REQUESTED A STELLAREX FOR THE POPLITEAL. NO IS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12862680 ↗MalfunctionJun 23, 2021Unintended Collision; Device Damaged by Another Device; Unintended Collision; Device Damaged by Another Device
UPON POST DILATION OF THE TACKS, THE BALLOON PASSED THE FIRST TACK WITHOUT ISSUE BUT CAME INTO CONTACT WITH THE SECOND TACK. THE BALLOON BECAME CAUGHT ON THE SECOND TACK. THE BALLOON PUSHED THE TACK SIDEWAYS IN THE VESSEL AND THEN DOWN THE SUPERFICIAL FEMORAL ARTERY (SFA). EVENTUALLY, THE OPERATOR WAS ABLE TO USE A STENT TO OVERLAY THE DISPLACED TACK, SECURING IT AGAINST THE VESSEL WALL. A SELF-EXPANDING STENT WAS PLACED IN THE PROXIMAL SFA TO COVER THE DISSECTIONS. AT THE CONCLUSION OF THE PROCEDURE, THERE WAS COMPLETE FLOW THROUGH THE SFA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12051440 ↗
Other Scaffold, Dissection Repair devices
- Tack Endovascular System (6F, 3.5-6.0mm) 156120061Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall2 reports
- Tack Endovascular System (4F, 1.5-4.9mm) 154090041Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall0 reports
- Tack Endovascular System (6F, 4.0-8.0mm) 206120061Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall0 reports
- Tack Endovascular System (4F, 1.5-4.5mm) 154150041Intact Vascular, Inc. · Scaffold, Dissection Repair7 reports
- Tack Endovascular System (6F, 3.5-6.0mm) 156135061Intact Vascular, Inc. · Scaffold, Dissection Repair3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page