Tack Endovascular System (4F, 1.5-4.5mm) 154150041Recalls & reported problems
Intact Vascular, Inc. · 150cm Catheter Length · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00863328000196), received between Jun 21, 2021 and Jul 3, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device5
- Ejection Problem1
- Inadequacy of Device Shape and/or Size1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2024Entrapment of Device; Entrapment of Device
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, A TACK WAS DEPLOYED SUCCESSFULLY. HOWEVER, DURING REMOVAL, THE TES GOT STUCK ON A 0.014" NON-PHILIPS GUIDEWIRE AND HAD TO BE REMOVED AS A SYSTEM. THE PROCEDURE WAS COMPLETED WITH A NEW GUIDEWIRE IN ORDER TO DELIVER A BALLOON FOR POST DILATION. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO REMOVAL AS A SYSTEM. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19672022 ↗MalfunctionJul 3, 2024Entrapment of Device; Entrapment of Device
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, A TACK WAS DEPLOYED SUCCESSFULLY. HOWEVER, DURING REMOVAL, THE TES GOT STUCK ON A 0.014" NON-PHILIPS GUIDEWIRE AND HAD TO BE REMOVED AS A SYSTEM. THE PROCEDURE WAS COMPLETED WITH A NEW GUIDEWIRE IN ORDER TO DELIVER A BALLOON FOR POST DILATION. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO REMOVAL AS A SYSTEM. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19671930 ↗MalfunctionJul 3, 2024Entrapment of Device; Entrapment of Device
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, THE TES GOT STUCK ON A 0.014" NON-PHILIPS GUIDEWIRE AND HAD TO BE REMOVED AS A SYSTEM. AN EXTENDED BALLOON WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO REMOVAL AS A SYSTEM. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19671664 ↗MalfunctionJul 3, 2024Entrapment of Device; Entrapment of Device
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, THE TES WAS ABLE TO ADVANCE WELL AND 3 OF THE 4 TACKS WERE DEPLOYED SUCCESSFULLY. HOWEVER, DURING REMOVAL, THE TES GOT STUCK HALFWAY OUT, AND HAD TO PULL THE TES AND GUIDEWIRE AS A UNIT. A NEW GUIDE WIRE WAS INSERTED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO REMOVAL AS A SYSTEM. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19671540 ↗MalfunctionJul 3, 2024Entrapment of Device; Entrapment of Device
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, THE TES WAS ABLE TO ADVANCE WELL AND TACKS WERE DEPLOYED SUCCESSFULLY. HOWEVER, DURING REMOVAL, THE TES GOT STUCK HALFWAY OUT, AND HAD TO PULL THE TES AND GUIDEWIRE AS A UNIT. A NEW GUIDE WIRE WAS INSERTED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED DUE TO REMOVAL AS A SYSTEM. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19671521 ↗MalfunctionJul 3, 2024Ejection Problem; Ejection Problem
THE TACK ENDOVASCULAR SYSTEM (TES) WAS USED IN A PERIPHERAL PROCEDURE. UPON ENTERING THE SHEATH, A TACK WAS INADVERTENTLY DEPLOYED BUT WAS ABLE TO BE REMOVED FROM THE ENTRANCE OF THE SHEATH. THE PROCEDURE WAS COMPLETED WITH A STENT. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE THE TACK INADVERTENTLY DEPLOYED. THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19665700 ↗MalfunctionJun 21, 2021Inadequacy of Device Shape and/or Size; Unintended Movement; Inadequacy of Device Shape and/or Size; Unintended Movement
THE REPORTER STATED "THE DEVICE WAS ORIGINALLY IMPLANTED (B)(6) 2021 AND TODAY THEY NOTICED DURING THE ANGIOGRAM THAT THE DEVICES HAVE MOVED. PRIOR TO AN INTERVENTION ON ANOTHER VESSEL ALTOGETHER." "IT WAS NOTICED THAT THE DEVICES HAD MOVED BETWEEN THIS PROCEDURE AND THE INITIAL PROCEDURE PERFORMED ON THE (B)(6) 2021." "IT IS IMPORTANT TO NOTE THAT THE MOVEMENT OF THESE DEVICES WERE NOTICED PRIOR TO ANY ADDITIONAL INTERNATIONAL MEASURES BEING TAKEN. BEFORE ANY WIRES REACHED THE STENT THEMSELVES. SOMETHING HAPPENED IN BETWEEN THE PROCEDURES AND CAUSED THE DEVICES TO MOVE. IT WAS NOTED VIA ANGIOGRAM DURING THE INITIAL RUN OFF BEFORE ANY WIRES WENT DOWN TO THE AREA." FOUR TACK ENDOVASCULAR DEVICES (TACKS) WERE IMPLANTED ON (B)(6) 2021 FOR TREATMENT OF A SIGNIFICANT SPIRAL DISSECTION IN THE TIBIAL PERONEAL TRUNK AS PART OF PURCUTANEOUS ATHERECTOMY PROCEDURE FOR A NON-HEALING ULCER OF LOWER EXTREMITY. POST-PROCEDURE ANGIOGRAMS ON (B)(6) 2021 SHOWED THE FOUR TACKS IMPLANTED PROPERLY AT THE DESIRED LOCATION IN THE VESSEL. ON A SUBSEQUENT ANGIOGRAM FROM (B)(6) 2021, THE THREE MOST DISTAL TACKS WERE SEEN IN ALTERED POSITIONS FROM THEIR ORIGINAL PLACEMENT. THREE TACKS APPEAR TO HAVE MOVED PR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12031330 ↗
Other Scaffold, Dissection Repair devices
- Tack Endovascular System (6F, 3.5-6.0mm) 156120061Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall2 reports
- Tack Endovascular System (4F, 1.5-4.9mm) 154090041Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall0 reports
- Tack Endovascular System (6F, 4.0-8.0mm) 206120061Intact Vascular, Inc. · Scaffold, Dissection Repair1 recall0 reports
- Tack Endovascular System (6F, 4.0-8.0mm) 206135061Intact Vascular, Inc. · Scaffold, Dissection Repair7 reports
- Tack Endovascular System (6F, 3.5-6.0mm) 156135061Intact Vascular, Inc. · Scaffold, Dissection Repair3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page