Skip to content

KidneyVault PRGNX-8008-001Recalls & reported problems

Paragonix Technologies Inc · KidneyVault Shipper · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850005470335), received between Aug 14, 2025 and Mar 16, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 20, 2026Class IIOpen, Classified

    Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.

    Z-2967-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 16, 2026

    AT THE TIME OF THE PATIENT'S RENAL TRANSPLANT, UPON REMOVING THE CANNISTER ASSEMBLY FROM THE SHIPPING HOUSING, IT WAS DISCOVERED THAT A PORTION OF THE PRESERVATION SOLUTION HAD LEAKED FROM THE CANISTER ASSEMBLY TO THE SHIPPER ASSEMBLY. UPON FURTHER INSPECTION, A TEAR IN THE CLOSED-CIRCUIT TUBING WAS IDENTIFIED APPROXIMATELY AT THE POINT IN WHICH THE TUBING CAME INTO CONTACT WITH THE PERISTALTIC PUMP / CLASPING. THE PATIENT WAS INFORMED OF THE FINDINGS; RISKS/BENEFITS WERE DISCUSSED AND THEY ELECTED TO PROCEED WITH TRANSPLANTATION. THEY HAVE SINCE RECOVERED WITHOUT ANY IDENTIFIED ADVERSE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24619168 ↗
  • MalfunctionAug 14, 2025Fluid/Blood Leak; Fluid/Blood Leak

    AT APPROXIMATELY 05:58 ON (B)(6) 2025 UPON INITIAL INSPECTION AT THE RECIPIENT TRANSPLANT CENTER, IT WAS NOTICED THAT THE FRONT OF THE DEVICE WAS WET AND APPEARED TO BE ACTIVELY LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22804829 ↗

Other System, Perfusion, Kidney devices

All System, Perfusion, Kidney devices →

Sources and updates