Skip to content

Paragonix Technologies Inc. KidneyVault Portable Renal… recallZ-2967-2026 · Paragonix Technologies Inc. · initiated Jul 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Paragonix Technologies Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.

FDA root cause category: Software design

What the company is doing

On July 20, 2026, Paragonix Technologies Inc./Getinge issued an Urgent Medical Device Correction to affected consignees via FedEX mail. Paragonix requested consignees to take the following actions: 1. An update to the Paragonix App has been released through the application store. 2. We recommend updating all mobile devices to latest version prior to KidneyVault usage. 3. Please continue to inspect all KidneyVault devices per the IFU before use to ensure proper datalogger function. 4. Continue to follow the IFU for using a multi-faceted clinical approach to assessing the organ, including any/all information available on the datalogger screen. 5. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 6. Complete and return the enclosed response form by e-mailing a copy to [email protected].

Recall details

Recall number
Z-2967-2026
Status
Open, Classified
Initiated by the company
Jul 20, 2026
Posted by the FDA
Aug 14, 2026
Recalling firm
Paragonix Technologies Inc.
Quantity
1445 systems
Product code
KDN
FDA submissions
K234060
Distribution
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.

Product description

Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Code information (models, lots, serial numbers)

All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates