Skip to content

NeuroStar Advanced Therapy System Version 3.6Recalls & reported problems

Neuronetics, Inc. · Neurostar Mobile Console 3.6, 120V, US · Class 2

All models of NeuroStar Advanced Therapy System (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00850005944058), received between Feb 8, 2023 and Feb 3, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 24, 2026Class IIOpen, Classified

    Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.

    Z-0142-2027 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Patient Device Interaction Problem2
  • Patient-Device Incompatibility2
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Convulsion/Seizure4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Dizziness1
  • Headache1
  • Tics/Tremor1
  • Tinnitus1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 3, 2026Adverse Event Without Identified Device or Use Problem

    THE TREATING PRACTICE CONTACTED NEURONETICS TO INFORM THEM THAT A PATIENT EXPERIENCED A SEIZURE DURING TMS.

    Patient problems: Convulsion/Seizure

    MAUDE report 24251666 ↗
  • InjurySep 3, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    NEURONETICS RECEIVED A REPORT FROM A PRACTICE ALLEGING WORSENING TINNITUS DURING TMS TREATMENT DESPITE WEARING HEARING PROTECTION.

    Patient problems: Tinnitus

    MAUDE report 22956987 ↗
  • InjuryMay 8, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    A PRACTICE NOTIFIED NEURONETICS THAT A PATIENT STARTED HAVING TREMORS AFTER RECEIVING A FULL COURSE OF TMS TREATMENT.

    Patient problems: Tics/Tremor

    MAUDE report 21984314 ↗
  • InjuryMar 27, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    NEURONETICS WAS MADE AWARE BY A TMS PRACTICE THAT A PATIENT EXPERIENCED RIGHT SIDE ARM AND LEG SHAKING AN HOUR AND A HALF FOLLOWING HER 25TH TMS TREATMENT. PATIENT WAS TAKEN TO THE ER AND CONFIRMED TO BE A SEIZURE.

    Patient problems: Convulsion/Seizure

    MAUDE report 21711669 ↗
  • InjuryNov 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NEURONETICS WAS CONTACTED BY A PROVIDER REPORTING THAT ONE OF THEIR PATIENTS HAD A SEIZURE AFTER COMPLETING HIS FIRST THETA BURST TREATMENT SESSION. EMS WAS CALLED AND THE PATIENT WAS SENT TO THE EMERGENCY ROOM.

    Patient problems: Convulsion/Seizure

    MAUDE report 20599257 ↗
  • InjuryMar 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PROVIDER'S OFFICE CALLED NEURONETICS TO REPORT THAT THEIR PATIENT'S MOTHER IS CONCERNED THAT THE PATIENT (HER SON) HAS EXPERIENCED A SIGNIFICANT WORSENING IN HIS OCD SYMPTOMS AFTER RECEIVING 16 TREATMENTS. THE PATIENT HAS REPORTEDLY BEEN STAYING UP "ALL NIGHT MAKING LISTS AND CHECKING THEM OVER AND OVER, NOT SLEEPING".

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18836090 ↗
  • InjuryAug 28, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    NEURONETICS RECEIVED A CALL FROM A PROVIDER REPORTING THAT ONE OF THEIR PATIENTS EXPERIENCED A SEIZURE DURING THEIR 9TH TREATMENT SESSION AND WAS TAKEN TO THE EMERGENCY ROOM.

    Patient problems: Convulsion/Seizure

    MAUDE report 17631922 ↗
  • InjuryFeb 8, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    A TMS PROVIDER CONTACTED NEURONETICS REGARDING A PATIENT WHO REPORTED THE FOLLOWING: "MY BRAIN HAS GONE A BIT WONKY WITH THE TREATMENTS. ABOUT THREE HOURS AFTER TREATMENT, IT'S LIKE A SEISMIC WAVE GOES OFF IN MY BRAIN, AND I'M HIT WITH A WAVE OF DIZZINESS. THE FIRST TWO TIMES IT RESOLVED COMPLETELY. I DID TAKE RELPAX. THE THIRD TIME IT DIDN'T RESOLVE. MY BRAIN HASN'T GONE BACK TO REGULAR VESTIBULAR WISE SINCE MY TREATMENT ON FRIDAY. I HAD A SMALLER WAVE OF OFFNESS/DIZZINESS HIT ME DURING THE TREATMENT AND AFTER THE SEISMIC WAVE LIKE FEELING IN MY BRAIN, IT ESCALATED THAT NIGHT AND STAYED THROUGH THE WEEKEND."

    Patient problems: Dizziness; Headache

    MAUDE report 16326454 ↗

FDA clearance and classification

Cleared through 510(k) K212289 on May 6, 2022 (decision: Substantially Equivalent). Submitted by Neuronetics, Inc. as “NeuroStar Advanced Therapy for adjunctive treatment of OCD, NeuroStar Advanced Therapy System, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy for Mental Health, NeuroStar”. The FDA review took 289 days. FDA record ↗

Cleared through 510(k) K201158 on Nov 20, 2020 (decision: Substantially Equivalent). Submitted by Neuronetics, Inc. as “NeuroStar Advanced Therapy”. The FDA review took 204 days. FDA record ↗

The FDA regulates product code OBP (transcranial magnetic stimulator) under 21 CFR 882.5805 as a Class II device. 21 CFR 882.5805 ↗

FDA definition of this device type: A transcranial magnetic stimulator is a device that delivers brief duration, rapidly alternating, or pulsed, magnetic fields to induce electrical currents that are directed at spatially discrete regions of the cerebral cortex for the treatment of major depressive disorder.

Common questions about the NeuroStar Advanced Therapy System Version 3.6

Has the NeuroStar Advanced Therapy System Version 3.6 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0142-2027 (Class II), began on Aug 24, 2026: Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NeuroStar Advanced Therapy System Version 3.6?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Feb 8, 2023 and Feb 3, 2026. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (2), “Patient-Device Incompatibility” (2) and “Improper or Incorrect Procedure or Method” (1); the most common patient problems are “Convulsion/Seizure” (4), “Dizziness” (1) and “Headache” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NeuroStar Advanced Therapy System Version 3.6?

None of the reports are classified as deaths, 8 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NeuroStar Advanced Therapy System Version 3.6 FDA approved?

It was cleared through 510(k) K201158 on Nov 20, 2020; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the NeuroStar Advanced Therapy System Version 3.6 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the NeuroStar Advanced Therapy System Version 3.6, and is it still sold?

The FDA's device database lists Neuronetics, Inc. as the company responsible for it (brand NeuroStar Advanced Therapy System, model Version 3.6). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Transcranial Magnetic Stimulator devices

All Transcranial Magnetic Stimulator devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page