NeuroStar Advanced Therapy System. Contains the NeuroStar… recallZ-0142-2027 · Neuronetics, Inc. · initiated Aug 24, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Neuronetics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.
FDA root cause category: Nonconforming Material/Component
What the company is doing
NeuroStar began notification to consignees on about 08/24/2026 via email. Consignees were instructed to check their NeuroStar System for Console and Coil for affected serial numbers and complete and return the response form. A NeuroStar Field Service Engineer will contact you to coordinate product inspection at no expense to you. If the HP Coild Assembly is not affected, no action is required. If it is determined the product is nonconforming, the affected product will be replaced with a HP Coil Assembly with the proper specifications to address the issue at no expense to you. The Field Service Engineer will handle the delivery and installation of the HP Coil Assembly and return of the affected coil. The actions are expected to be completed in the next 12 months.
Recall details
- Recall number
- Z-0142-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Aug 24, 2026
- Posted by the FDA
- Oct 9, 2026
- Recalling firm
- Neuronetics, Inc.
- Quantity
- 179 units
- Product code
- OBP
- FDA submissions
- K231926, K233621
- Distribution
- Worldwide distribution - US Nationwide and the countries of Japan and South Korea.
Product description
NeuroStar Advanced Therapy System. Contains the NeuroStar HP Coil Assembly Part Number 81-01000-200. Affects the below NeuroStar Advanced Therapy Systems: 1. NeuroStar 2.0 US; Part Number: 81-01315-000. 2. NeuroStar 3.0 US; Part Number: 81-02315-000. 3. NeuroStar 2.X to 3.2 Upgrade Kit; Part Number: 81-25000-100. 4. NeuroStar Upgrade Kit 3.0 US; Part Number: 81-35000-000. 5. NeuroStar Mobile Console 3.6, 120V, US; Part Number: 81-03315-000. 6. NeuroStar Mobile Console 3.7, 120V, US; Part Number: 81-04315-000. 7. NeuroStar Mobile Console 3.8, 120V, US; Part Number: 81-05315-000.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
NeuroStar Software Versions 3.6, 3.7, 3.8 1. NeuroStar 2.0 US; Part Number: 81-01315-000; UDI-DI: 00869378000100. 2. NeuroStar 3.0 US; Part Number: 81-02315-000; UDI-DI: 00869378000117. 3. NeuroStar 2.X to 3.2 Upgrade Kit; Part Number: 81-25000-100; UDI-DI: 00869378000193. 4. NeuroStar Upgrade Kit 3.0 US; Part Number: 81-35000-000; UDI-DI: 00869378000155. 5. NeuroStar Mobile Console 3.6, 120V, US; Part Number: 81-03315-000; UDI-DI: 00850005944058. 6. NeuroStar Mobile Console 3.7, 120V, US; Part Number: 81-04315-000; UDI-DI: 00850005944522. 7. NeuroStar Mobile Console 3.8, 120V, US; Part Number: 81-05315-000; UDI-DI: 00850005944621. NeuroStar HP Coil Assembly Part Number: 81-01000-200. Coil Serial Numbers: 40203, 40208, 40202, 40204, 40206, 40205, 40207, 40209, 40210, 40211, 40212, 40213, 40214, 40215, 40216, 40217, 40218, 40219, 40220, 40221, 40222, 40223, 40224, 40225, 40227, 40228, 40226, 40229, 40230, 40231, 40232, 40233, 40234, 40236, 40235, 40238, 40239, 40240, 40241, 40242, 40243, 40244, 40245, 40246, 40247, 40248, 40249, 40250, 40251, 40252, 40253, 40254, 40255, 40256, 40257, 40258, 40259, 40260, 40261, 40262, 40263, 40265, 40266, 40267, 40264, 40283, 40284, 40285, 40286, 40287, 40288, 40289, 40290, 40291, 40293, 40292, 40294, 40295, 40296, 40297, 40298, 40299, 40300, 40301, 40302, 40303, 40304, 40305, 40306, 40307, 40308, 40309, 40310, 40311, 40312, 40313, 40314, 40315, 40316, 40317, 40318, 40320, 40319, 40321, 40322, 40323, 40324, 40325, 40326, 40327, 40328, 40329, 40331, 40332, 40333, 40334, 40335, 40336, 40337, 40338, 40339, 40340, 40341, 40342, 40343, 40344, 40347, 40346, 40345, 40348, 40349, 40350, 40353, 40352, 40351, 40354, 40355, 40356, 40357, 40358, 40359, 40360, 40371, 40372, 40373, 40374, 40375, 40376, 30643, 40274, 30638, 40281, 40199, 30671, 40201, 40272, 30641, 40278, 30646, 30653, 30640, 30647, 30652, 40275, 30648, 40197, 40279, 30650, 30639, 30623, 40198, 40280, 40270, 30645, 30654, 30636, 40200, 30635, 40282, 30606, 30607, 30608, 30609, 30610, 30611, 30624, 30625, 30630, 30631, 30632, 30633, 30634, 30655, 30656, 30657, 30658, 30659, 30660, 30622, 30621, 30626, 30614, 30620, 30616, 30619, 30615, 30617, 30661, 30662, 30663, 30664, 30665, 30627, 30628, 30629, 30644, 40268, 40269.
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- NeuroStar TMS Therapy System NeuroStar 3.0 USNeuronetics, Inc. · Transcranial Magnetic Stimulator1 recall58 reports
- NeuroStar Advanced Therapy System Version 3.7Neuronetics, Inc. · Transcranial Magnetic Stimulator1 recall13 reports
- NeuroStar Advanced Therapy System Version 3.8Neuronetics, Inc. · Transcranial Magnetic Stimulator1 recall12 reports
- NeuroStar Advanced Therapy System Version 3.6Neuronetics, Inc. · Transcranial Magnetic Stimulator1 recall8 reports
- NeuroStar TMS Therapy System NeuroStar 2.0 USNeuronetics, Inc. · Transcranial Magnetic Stimulator1 recall2 reports
- NeuroStar TMS Therapy System Upgrade Kit 3.0 USNeuronetics, Inc. · Transcranial Magnetic Stimulator1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page