Adverse event reports
7
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00850008222061), received between Dec 30, 2025 and Sep 30, 2026 — classified as 1 death, 6 malfunctions.

The most frequently reported specific device problem is “Human-Device Interface Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Human-Device Interface Problem3
  • No Apparent Adverse Event2
  • Adverse Event Without Identified Device or Use Problem1
  • Migration1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Human-Device Interface Problem

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50955051 ↗
  • MalfunctionSep 30, 2026Human-Device Interface Problem

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50955049 ↗
  • MalfunctionSep 11, 2026No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50742870 ↗
  • MalfunctionSep 11, 2026No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50742867 ↗
  • MalfunctionApr 3, 2026Appropriate Device Problem Term/Code Not Available; Patient Device Interaction Problem

    A REPRESENTATIVE REPORTED THAT DURING A PAVM PROCEDURE, A LARGE PULMONARY ARTERY BRANCH DID NOT ACHIEVE OCCLUSION APPROXIMATELY 20 MINUTES AFTER A LOBO-9 VASCULAR OCCLUSION SYSTEM DEVICE WAS DETACHED IN THE INTENDED VESSEL. THE VESSEL WAS REPORTED TO MEASURE APPROXIMATELY 8 MM IN DIAMETER. THE PHYSICIAN WAITED AN ADDITIONAL FIVE (5) MINUTES AND SUBSEQUENTLY DEPLOYED A SECOND LOBO-9 DEVICE PROXIMAL TO THE INITIAL DEVICE. AFTER AN ADDITIONAL 15 MINUTES, FLOW WAS OBSERVED TO DECREASE; HOWEVER, COMPLETE OCCLUSION WAS NOT ACHIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24779630 ↗
  • MalfunctionMar 25, 2026Migration

    IT WAS REPORTED THAT A PATIENT WITH AN ENLARGED SPLEEN AND NO EVIDENCE OF ACTIVE SPLENIC BLEEDING WAS SCHEDULED TO UNDERGO A SPLENECTOMY. THE PHYSICIAN ELECTED TO PERFORM A SPLENIC ARTERY EMBOLIZATION PRIOR TO SURGERY TO REDUCE BLEEDING RISK. THE TARGET VESSEL WAS REPORTED TO MEASURE APPROXIMATELY 6.8 MM IN DIAMETER. A LOBO-9 VASCULAR OCCLUSION SYSTEM DEVICE WAS SELECTED AND DEPLOYED FOR EMBOLIZATION. FOLLOWING DEVICE DEPLOYMENT AND DETACHMENT, IT WAS REPORTED THAT THE LOBO-9 DEVICE MIGRATED FORWARD BY APPROXIMATELY 1 CM FROM THE INTENDED PLACEMENT LOCATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24687166 ↗
  • DeathDec 30, 2025Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem; Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2025, IT WAS REPORTED THAT THE PATIENT UNDERWENT PORTAL VEIN THROMBECTOMY AND THROMBOLYSIS, ALONG WITH PLUG-ASSISTED ANTEROGRADE TRANSVENOUS OBLITERATION (PATO) VIA A PERCUTANEOUS TRANSHEPATIC APPROACH, TO REDUCE THE RISK ASSOCIATED WITH RECENTLY BLEEDING HIGH-RISK GASTRIC VARICES AND TO RESTORE PATENCY OF THE MAIN PORTAL VEIN THROMBUS. SUCCESSFUL VARIX EMBOLIZATION WAS ACHIEVED USING THE LOBO VASCULAR OCCLUSION SYSTEM DEVICES AND GELFOAM BY THE CONCLUSION OF THE PROCEDURE. THE PATIENT WAS DISCHARGED TO FLOOR ROOM IN STABLE CONDITION WITHOUT ANY CONCERNS. OVERNIGHT ON (B)(6) 2025, THE PATIENT BEGAN BLEEDING AND WAS BROUGHT EMERGENTLY FOR ACTIVE HEMORRHAGE FROM THE GASTRIC VARIX (AS NOTED BY EMERGENT BEDSIDE UPPER ENDOSCOPY). THE PATIENT BECAME ANEMIC/UNSTABLE AND THUS WAS AGGRESSIVELY RESUSCITATED IN A 'MASSIVE TRANSFUSION PROTOCOL' (MTP) FASHION. EMERGENT TIPS WAS CREATED FOR PORTAL ACCESS AND PORTAL HYPERTENSION DECOMPRESSION. PORTAL VENOGRAPHY DEMONSTRATED PATENT RECENTLY EMBOLIZED VARIX VESSELS - WITH ACTIVE EXTRAVASATION FROM THE GASTRIC VARIX ON THE OTHER SIDE OF THE LOBO VASCULAR OCCLUDERS. TO ADDRESS THE VARIX BLEED, ONE OF THE LOBO VASCULAR OCCLUDERS WAS REPORTED

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23931098 ↗

Other Device, Vascular, For Promoting Embolization devices