INTERA AP-04013-5Recalls & reported problems
Intera Oncology Inc. · Special Bolus Needle for use with Intera 3000 Hepatic Artery Infusion Pump Straight Length 1 3/8" (3.5cm) · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850014110048), received between Jun 18, 2025 and Jan 30, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component1
- Difficult to Flush1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2026Difficult to Flush
INTERA ONCOLOGY WAS NOTIFIED THAT DURING PUMP PREPARATION FOR AN IMPLANT ON (B)(6) 2026, THE SCRUB TECH WASN'T ABLE TO FLUSH THE PUMP BOLUS PATHWAY WITH LOW DOSE HEPARINIZED SALINE. THEY OPENED THE BACKUP SPECIAL BOLUS NEEDLE AND WAS ABLE TO FLUSH THE PUMP BOLUS PATHWAY. THE PUMP WAS SUCCESSFULLY IMPLANTED AND THE CLINIC SET THE MALFUNCTIONED NEEDLE ASIDE TO BE EXAMINED BY INTERA ONCOLOGY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24227113 ↗MalfunctionJun 18, 2025Defective Component; Defective Component
INTERA ONCOLOGY RECEIVED A REPORT FROM OUR REPRESENTATIVE THAT HE AND THE PHYSICIAN NOTICED DURING AN IMPLANTATION ON (B)(6) 2025, THAT THE SPECIAL BOLUS NEEDLE WAS EXPERIENCING BACKFLOW OF BLOOD, EVEN THOUGH THE NEEDLE WAS CLAMPED WHEN NOT IN USE. OUR REPRESENTATIVE INSTRUCTED THE SCRUB TECH TO UNCLAMP THE NEEDLE AND INJECT 5 ML OF LOW-DOSE HEPARINIZED SALINE TO CLEAR THE BLOOD, THEN RECLAMP IT IMMEDIATELY AFTERWARD. HOWEVER, ONE TO TWO MINUTES LATER, BLOOD WAS AGAIN FLOWING BACK INTO THE NEEDLE AND SYRINGE, DESPITE THE CLAMP BEING IN PLACE. THE NEEDLE WAS REMOVED, AND THE SPARE SPECIAL BOLUS NEEDLE WAS OPENED. WITH THE SPARE NEEDLE, THERE WERE NO ISSUES OF BACKFLOW INTO THE NEEDLE OR SYRINGE WHEN CLAMPED. DURING COMMUNICATION WITH THE CLINIC, THE SURGEON CONFIRMED THE PATIENT DID NOT EXPERIENCE ANY ADVERSE REACTIONS. IN ADDITION, THE CLINIC PROVIDED INTERA ONCOLOGY WITH THREE PICTURES THAT SHOW THE SPECIAL BOLUS NEEDLE EXPERIENCING BACKFLOW OF BLOOD, EVEN THOUGH THE NEEDLE WAS CLAMPED WHEN NOT IN USE. DUE TO SAFETY CONCERNS FOR CONTAMINATED NEEDLE, INTERA ONCOLOGY ASKED THE CLINIC TO DISPOSE THE SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22286574 ↗
Other Non-Coring (Huber) Needle devices
- SafeStep CPA00049Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00040Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00042ABard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- INTERA AP-07014USIntera Oncology Inc. · Non-Coring (Huber) Needle37 reports
- SURECAN SAFETY II 4447006-02B. Braun Medical Inc. · Non-Coring (Huber) Needle13 reports
- SURECAN SAFETY II 4447005-02B. Braun Medical Inc. · Non-Coring (Huber) Needle6 reports
- AllPoints Port Access System SafeStep AN152075MBard Access Systems, Inc. · Non-coring (Huber) needle5 reports
- SURECAN SAFETY II 4447012-02B. Braun Medical Inc. · Non-Coring (Huber) Needle4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.