Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850014110154), received between Aug 19, 2025 and Jan 13, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2026Fluid/Blood Leak
INTERA ONCOLOGY WAS NOTIFIED BY A PHYSICIAN THAT ON (B)(6) 2025, DURING OR PUMP PREP, THE OR PREP KIT HAD A LEAK BETWEEN THE NEEDLE AND THE TUBING. THE CONNECTION WAS VERY TIGHT BUT THERE WAS STILL LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24056760 ↗MalfunctionSep 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
INTERA ONCOLOGY RECEIVED A REPORT FROM A PHYSICIAN THAT DURING PUMP PREPARATION, THE PUMP WAS FILLED WITH 5ML'S, THE SCRUB TECH NOTICED A LEAK BETWEEN THE NEEDLE AND TUBING. AN INTERA ONCOLOGY REPRESENTATIVE INSTRUCTED THE PHYSICIAN TO CLAMP THE CENTER OF THE TUBING, BUT THE LEAK CONTINUED, LIKELY FROM THE 5ML'S IN THE PUMP PUSHING BACK. OUR REPRESENTATIVE INSTRUCTED THE PHYSICIAN TO PULL THE NEEDLE OUT OF THE PUMP AND TIGHTEN THE CONNECTION AS HARD AS SHE COULD WITH HER HANDS. SHE DID THEN INSERT THE NEEDLE BACK IN THE PUMP, THE LEAK CONTINUED. OUR REPRESENTATIVE THEN INSTRUCTED THE PHYSICIAN TO PULL THE NEEDLE FROM THE PUMP AND USE SOME HEMOSTATS TO FULLY SECURE THE CONNECTION. AFTER THIS THERE WAS NO MORE LEAKING AT THE CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23043343 ↗MalfunctionAug 19, 2025Fluid/Blood Leak; Fluid/Blood Leak
INTERA ONCOLOGY RECEIVED A REPORT THAT DURING PUMP PREPARATION THE TUBING FROM THE OR PREP KIT WAS LEAKING NEAR WHERE IT CONNECTED WITH THE REFILL NEEDLE. ONCE THE CLINIC NOTICED THE LEAK, THEY REPLACED THE TUBING BY OPENING A NEW OR PREP KIT. THE PUMP PREPPED FINE AFTERWARDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22841456 ↗
Other Pump, Infusion, Implanted, Programmable devices
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page