Skip to content

INTERA AP-07004 (REF AP07004)Recalls & reported problems

Intera Oncology Inc. · Intera 3000 Hepatic Artery Infusion Pump OR Prep Kit, Contents: (1) Huber Needle 22 Gauge, (1) Special Bolus Needle, (1) Empty Syringe Barrel with Lid · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850014110154), received between Aug 19, 2025 and Jan 13, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026Fluid/Blood Leak

    INTERA ONCOLOGY WAS NOTIFIED BY A PHYSICIAN THAT ON (B)(6) 2025, DURING OR PUMP PREP, THE OR PREP KIT HAD A LEAK BETWEEN THE NEEDLE AND THE TUBING. THE CONNECTION WAS VERY TIGHT BUT THERE WAS STILL LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056760 ↗
  • MalfunctionSep 12, 2025Fluid/Blood Leak; Fluid/Blood Leak

    INTERA ONCOLOGY RECEIVED A REPORT FROM A PHYSICIAN THAT DURING PUMP PREPARATION, THE PUMP WAS FILLED WITH 5ML'S, THE SCRUB TECH NOTICED A LEAK BETWEEN THE NEEDLE AND TUBING. AN INTERA ONCOLOGY REPRESENTATIVE INSTRUCTED THE PHYSICIAN TO CLAMP THE CENTER OF THE TUBING, BUT THE LEAK CONTINUED, LIKELY FROM THE 5ML'S IN THE PUMP PUSHING BACK. OUR REPRESENTATIVE INSTRUCTED THE PHYSICIAN TO PULL THE NEEDLE OUT OF THE PUMP AND TIGHTEN THE CONNECTION AS HARD AS SHE COULD WITH HER HANDS. SHE DID THEN INSERT THE NEEDLE BACK IN THE PUMP, THE LEAK CONTINUED. OUR REPRESENTATIVE THEN INSTRUCTED THE PHYSICIAN TO PULL THE NEEDLE FROM THE PUMP AND USE SOME HEMOSTATS TO FULLY SECURE THE CONNECTION. AFTER THIS THERE WAS NO MORE LEAKING AT THE CONNECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23043343 ↗
  • MalfunctionAug 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    INTERA ONCOLOGY RECEIVED A REPORT THAT DURING PUMP PREPARATION THE TUBING FROM THE OR PREP KIT WAS LEAKING NEAR WHERE IT CONNECTED WITH THE REFILL NEEDLE. ONCE THE CLINIC NOTICED THE LEAK, THEY REPLACED THE TUBING BY OPENING A NEW OR PREP KIT. THE PUMP PREPPED FINE AFTERWARDS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22841456 ↗

Other Pump, Infusion, Implanted, Programmable devices