Skip to content

Sterile Polyester Spun Swab 25 Pack 60566Recalls & reported problems

Steripack (usa) Limited LLC · single use, sterile pouched polyester spun swabs for specimen ciollection sampling from nasal area, i.e. Anterior nares · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Apr 25, 2023Class IIOpen, Classified

    Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab

    Z-1970-2023 · Root cause: Process control · Recall details · FDA record ↗

Other Applicator, Absorbent Tipped, Sterile devices

All Applicator, Absorbent Tipped, Sterile devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page