Skip to content

Maveric Medical M6234Recalls & reported problems

Maveric Medical LLC · Blood Collection Set, Safety, 23G x 3/4", 12" Tubing, Luer Adapter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850028131046), received between Apr 14, 2026 and Apr 14, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fail-Safe Did Not Operate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fail-Safe Did Not Operate1
  • Lack of Effect1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Fail-Safe Did Not Operate; Lack of Effect

    COMPLAINT 1: PHLEBOTOMIST EXPERIENCED ISSUE THIS MORNING WHEN USING THE HOSPITAL'S NEW MAVERIC BUTTERFLY. THE TUBING PROTRUDED FROM THE SIDE OF THE SAFETY SLEEVE AND PREVENTED PROPER ACTIVATION OF THE SAFETY DEVICE. THE PHLEBOTOMIST STATES THIS IS THE SECOND TIME THIS HAS HAPPENED. SEE SAMPLE PHOTO BELOW OF REPORTED ISSUE: COMPLAINT 2: THE ASSOCIATES INCLUDING OUR LEAD PHLEBOTOMIST SHARED THAT THE NEW BUTTERFLY DEVICES ARE DIFFICULT TO USE. THEY ARE HARD TO MANEUVER. THEY DON'T SEE THE FLASH. EVEN THOUGH YOU'RE IN THE VEIN, THE BLOOD FLOW IS SLUGGISH. AS STATED, ENGAGING THE SAFETY IS DIFFICULT. THEY HAVE CONCERNS THAT A POTENTIAL SHARPS INJURY WILL LIKELY OCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870249 ↗

Other Tubes, Vials, Systems, Serum Separators, Blood Collection devices

All Tubes, Vials, Systems, Serum Separators, Blood Collection devices →