N/A 10016NBRecalls & reported problems

Microport Navibot International LLC · Navigation Trolley Assembly · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850030914002), received between May 22, 2025 and Jun 13, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Off-Label Use” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use2
  • Difficult or Delayed Positioning1

Patient problems

  • Bone Notching2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 13, 2025Difficult or Delayed Positioning; Difficult or Delayed Positioning

    ALLEGEDLY, SKYWALKER NUMBER REF REF10016-NB, SN (B)(6) WOULD NOT LOCK INTO THE TIBIA DURING THE CUTTING SEQUENCE WITH MOTION FOLLOW, SURGEON CONVERTED TO MANUAL CASE AND CANCELLED THE FOLLOWING TWO CASES TO USE THE ROBOT.

    Patient problems: Insufficient Information

    MAUDE report 22207406 ↗
  • InjuryMay 22, 2025Off-Label Use; Off-Label Use

    ALLEGEDLY, REP WAS NOTIFIED BY THE SURGEON ON (B)(6) 2025 OF SECOND ANTERIOR FEMORAL NOTCHING SEEN ON POST-OP X-RAYS. THE PROCEDURE WAS DONE USING THE SKYWALKER ROBOT WITH EVO MEDIAL-PIVOT KNEE IMPLANTS. THE ROBOTIC SURGICAL PLAN SHOWED AVOIDANCE OF ANTERIOR FEMORAL NOTCHING.

    Patient problems: Bone Notching

    MAUDE report 22074631 ↗
  • InjuryMay 22, 2025Off-Label Use; Off-Label Use

    ALLEGEDLY, ON (B)(6) 2025 OF ANTERIOR FEMORAL NOTCHING SEEN ON POST-OP XRAYS. THE PROCEDURE WAS DONE USING THE SKYWALKER ROBOT WITH EVO MEDIAL-PIVOT KNEE IMPLANTS. THE ROBOTIC SURGICAL PLAN SHOWED AVOIDANCE OF ANTERIOR FEMORAL NOTCHING.

    Patient problems: Bone Notching

    MAUDE report 22073960 ↗