LimFlow V-Ceiver VC-US-23Recalls & reported problems

Inari Medical, Inc. · The LimFlow V-Ceiver is intended for use in the cardiovascular system to manipulate and retrieve guidewires specified in the IFU. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850041730141), received between Nov 19, 2024 and Nov 19, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Entrapment of Device1

Patient problems

  • Foreign Body Embolism1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2024Break; Entrapment of Device; Break; Entrapment of Device

    A PATIENT WITH CLTI (CHRONIC LIMB-THREATENING ISCHEMIA), CLASSIFIED AS RUTHERFORD CATEGORY 6, UNDERWENT A TADV PROCEDURE (TRANSCATHETER ARTERIALIZATION OF DEEP VEINS) USING THE LIMFLOW DEVICES ON (B)(6) 2024. DURING THE PROCEDURE, THE V-CEIVER WAS INSERTED INTO THE PATIENT'S LATERAL PLANTAR VEIN (LPV) OVER A COMMAND 0.014" GUIDEWIRE. THE GUIDEWIRE HAD TRAVERSED FROM THE POSTERIOR TIBIAL VEIN THROUGH A PERFORATOR AND TO THE PERONEAL VEIN. DUE TO THE TORTUOSITY OF THIS PATIENT'S VENOUS ANATOMY, THE V-CEIVER COULD NOT TRAVERSE THROUGH THE PERONEAL VEIN; THE PHYSICIAN ATTEMPTED TO DELIVER THE V-CEIVER BUT WAS UNSUCCESSFUL. THEY DECIDED TO RETRACT THE V-CEIVER AND SWITCH OUT THE WIRE FOR A STIFFER (V14) GUIDEWIRE HOWEVER THEY WERE STILL NOT ABLE TO DELIVER THE V-CEIVER IN THE PERONEAL VEIN. THE PHYSICIAN STARTED TO RETRACT THE DEVICE WHEN THE TIP BECAME STUCK AND DISCONNECTED FROM THE CATHETER. THE PHYSICIAN MADE AN EFFORT TO RETRIEVE THE DETACHED TIP BY BALLOONING PROXIMAL TO THE TOP IN ATTEMPT TO DRAG THE TIP DOWN INTO THE PEDAL SHEATH WITH NO SUCCESS. GIVEN THE INABILITY TO RETRIEVE THE TIP, THE DECISION WAS MADE TO LEAVE IT IN PLACE. TO PREVENT MIGRATION, A STENT WAS DEPLOYED OVER T

    Patient problems: Foreign Body Embolism

    MAUDE report 20720077 ↗

Other Device, Percutaneous Retrieval devices

All Device, Percutaneous Retrieval devices →