PNS Freedom Permanent Tined Neurostimulator Receiver Kit Channel A STQ4-RCV-A0Recalls & reported problems
Curonix LLC · PNS Freedom Permanent Tined Neurostimulator Receiver Kit Channel A · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Curonix PNS Freedom neurostimulator is an implanted pain relief device. Of its 350 reports, 343 describe injuries and 7 mention deaths. Common problems include pain, slow wound healing, skin erosion and infection, and some describe the device moving or being expelled. No recalls name this device, and reports fell to 92 in the last 12 months from 153 before that.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 364
- Deaths reported
- 8
- Injuries reported
- 356
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 364 adverse event reports that list this exact device (DI 00850051034406), received between Jul 17, 2023 and Sep 29, 2026 — classified as 8 deaths, 356 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (174 reports).
6 other recalls cover devices cleared under the same FDA submission (K152178); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem174
- Adverse Event Without Identified Device or Use Problem150
- Expulsion19
- Migration18
- Off-Label Use10
- Migration or Expulsion of Device7
- Fracture5
- Detachment of Device or Device Component3
Patient problems
- Pain78
- Impaired Healing71
- Erosion59
- Skin Infection49
- Post Operative Wound Infection46
- Swelling/ Edema36
- Fluid Discharge21
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 25 most recent of 364 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
The patient reported a stitch was eroding from the implant site. The patient went to the hospital due to pain at the suture site and was given guidance on how to move forward with imaging. On (b)(6) 2026, X-Rays were performed, migration was not suspected, and it was noted the patient did not have stiches poking through their skin. However, the patient had bulging at the pocket site where the coils and stitches were. A revision procedure was performed where the physician buried the stitches and coil deeper under the skin to reduce patient discomfort. No further issues have been reported.
Patient problems: Pain
MAUDE report 50937584 ↗InjurySep 29, 2026Use of Device Problem
The patient reported their incision re-opened and was now infected. Antibiotics were prescribed and an explant procedure was performed on (b)(6) 2026.
Patient problems: Post Operative Wound Infection; Wound Dehiscence
MAUDE report 50936648 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
The patient was taken to the hospital on (b)(6) 2026 due to swelling in the lower extremity and concerns of infection. Antibiotics were prescribed the week prior after a follow up appointment with the implanting clinician. As of (b)(6) 2026, the patient completed their antibiotics, they are doing well, and resumed using the neurostimulator without issue.
Patient problems: Swelling/ Edema
MAUDE report 50886588 ↗InjurySep 24, 2026Use of Device Problem
The patient reported an infection at the receiver site with clear liquid discharge off and on. Antibiotics were prescribed and an explant procedure was approved. Several attempts have been made to schedule the explant procedure. However, the patient refuses to move forward and does not want to take the device out. The physician is planning to speak to the patient again. No further information is available at this time.
Patient problems: Fluid Discharge; Post Operative Wound Infection
MAUDE report 50885184 ↗InjurySep 17, 2026Use of Device Problem
Patient problems: Impaired Healing
MAUDE report 50809527 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
The patient reported their distal incision at the receiver site was opening and started to show signs of infection with redness, pus, and odor. Antibiotics were prescribed and an incision washout was performed on (b)(6) 2026. As of (b)(6) 2026, the patient is fine. No further information is available at this time.
Patient problems: Post Operative Wound Infection; Purulent Discharge; Wound Dehiscence; Wound Odor
MAUDE report 50804957 ↗InjurySep 10, 2026Use of Device Problem
Patient problems: Erosion
MAUDE report 50721125 ↗DeathSep 10, 2026Adverse Event Without Identified Device or Use Problem
Curonix Support was notified the patient had passed on (b)(6) 2026 from non-device related factors. There were no alleged deficiencies, therapy was working properly, and the family would be returning the Transmitter Assemblies.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50718039 ↗InjurySep 10, 2026Use of Device Problem
Patient problems: Pain; Swelling/ Edema
MAUDE report 50717506 ↗InjurySep 8, 2026Use of Device Problem
Patient problems: Cellulitis
MAUDE report 50676335 ↗InjurySep 8, 2026Use of Device Problem
The patient reported blisters between each incision on both of their legs. The patient stopped using the device and the implanting clinician is considering sending the patient to an allergist to test for allergies to implant materials. As of (b)(6) 2026, there are no new updates or information at this time.
Patient problems: Blister
MAUDE report 50673949 ↗InjurySep 8, 2026Use of Device Problem
The patient reported erosion and an explant procedure was performed on (b)(6) 2026. No further issues have been reported.
Patient problems: Erosion
MAUDE report 50671501 ↗InjuryAug 20, 2026Use of Device Problem
The patient reported periodic uncomfortable stimulation. The frequency was lowered on the Transmitter Assemblies and as of (b)(6) 2026, the shocking sensations have gone away and the patient is getting marginal relief.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50356236 ↗InjuryAug 3, 2026Use of Device Problem
Patient problems: Impaired Healing
MAUDE report 50116577 ↗InjuryJul 31, 2026Migration
The patient reported "something" was sticking out of their leg, bruising, as well as the incision not fully closing. The patient was seen on (b)(6) 2026 and it was noted a suture had eroded on both sides of the incision site. However, there were no signs of infection. Additionally, X-Rays were performed which confirmed device migration of the right neurostimulator. The protruding sutures were clipped and a follow-up was scheduled for additional monitoring. Additional information was provided. The bruising had resolved and an explant procedure was performed on (b)(6) 2026. No further issues have been reported.
Patient problems: Bruise/Contusion; Impaired Healing
MAUDE report 50107358 ↗InjuryJul 31, 2026Adverse Event Without Identified Device or Use Problem
The patient experienced severe itching and drainage from the incision sites. Antibiotics were prescribed, the patient has seen wound care, the itchiness was clearing, the wound was wrapped, and the patient was instructed not to resume therapy until the wounds were completely healed. Additional information was received on July 29, 2026. Although the wounds healed under wound care, the itchiness and drainage continued and the wound opened back up. An explant procedure will be performed. However, a date has not yet been scheduled. No further information is available at this time.
Patient problems: Fluid Discharge; Itching Sensation
MAUDE report 50105628 ↗InjuryJul 30, 2026Insufficient Device Problem Information
During an implant procedure, the implanting clinician was unable to get consistent stim artifact and CMAP post suturing. As a result, the patient was under anesthesia for longer than 30 minutes. The neurostimulator was explanted and the implant procedure was aborted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50093836 ↗InjuryJul 29, 2026Use of Device Problem
The reported an infection in their left shoulder, that moved into their prosthetic shoulder joint. Antibiotics were prescribed and an explant procedure was performed on (b)(6)2026. As of (b)(6) 2026, the patient is recovering from the explant procedure. No further issues have been reported.
Patient problems: Medical device site infection
MAUDE report 50077587 ↗InjuryJul 29, 2026Adverse Event Without Identified Device or Use Problem
The patient reported their receiver site had not healed. Cultures were obtained. However, an infection was not confirmed. The patient has been going to wound care for three months where they have been monitoring, debriding, and dressing the wound. Additionally, the patient completed a short term steroid to improve healing. As of (b)(6) 2026, the incision has improved significantly since their previous visit and the wound has reduced in size greater than 50%. The patient will follow up with wound clinic, they are using their device and getting great relief. No further issues have been reported.
Patient problems: Impaired Healing
MAUDE report 50076670 ↗InjuryJul 29, 2026Use of Device Problem
The patient reported increased pain in their lower leg and no longer receiving pain relief. On (b)(6) 2026, the Commercial Team Member met with the patient and re-educated them on where to place the Transmitter Assembly Antenna to receive adequate therapy. Additionally, the power index was decreased on the Transmitter Assembly. Therapy has been restored and the pain resolved. The patient will follow-up with the Commercial Team Member to see how therapy is going.
Patient problems: Pain
MAUDE report 50074049 ↗DeathJul 2, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, THE COMMERCIAL TEAM MEMBER WAS NOTIFIED THE PATIENT PASSED AWAY. HOWEVER, THE SPECIFIC DATE OF THE PASSING IS UNKNOWN. THE HOSPITAL HAS EXPLICITLY CONFIRMED THAT THIS DEATH WAS NOT RELATED TO THE CURONIX SYSTEM AND THERE WERE NO ALLEGED DEFICIENCIES ASSOCIATED WITH THE DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25697814 ↗InjuryJun 30, 2026Use of Device Problem
THE PATIENT REPORTED EROSION OF THE NEUROSTIMULATOR. ON (B)(6) 2026, A REVISION PROCEDURE WAS PERFORMED WHERE THE PHYSICIAN RE-SUTURED THE NEUROSTIMULATOR TO THE FASCIA. ANTIBIOTICS WERE GIVEN DURING THE REVISION PROCEDURE. THE PATIENT ATTENDED A FOLLOW-UP APPOINTMENT ON (B)(6) 2026, WHERE THE STAPLES WERE REMOVED. ADDITIONALLY, THE PATIENT REPORTED THEY NO LONGER HAVE THE BAD ITCHING FROM THE EROSION, AND THEY ARE DOING WELL. NO FURTHER ISSUES HAVE BEEN REPORTED.
Patient problems: Erosion
MAUDE report 25668291 ↗InjuryJun 29, 2026Use of Device Problem
THE PATIENT REPORTED AN INFECTION IN THEIR LEFT LOWER LEG. AN EXPLANT PROCEDURE WAS PERFORMED ON (B)(6) 2026, I.V. ANTIBIOTICS WERE PRESCRIBED, AND THE PATIENT STAYED IN THE HOSPITAL FOR SEVERAL DAYS FOR OBSERVATION. HOWEVER, THE DATES OF THE HOSPITAL OBSERVATION AND SERIAL NUMBER OF THE DEVICE THAT WAS EXPLANTED ARE UNKNOWN. THE PATIENT IS DOING BETTER, CURRENTLY TAKING ORAL ANTIBIOTICS, WEARING A WOUND VAC, AND THE WOUND VAC DRESSING IS BEING CHANGED EVERY FEW DAYS. THE COMMERCIAL TEAM MEMBER HAS MADE SEVERAL ATTEMPTS TO CONTACT THE PATIENT FOR FURTHER INFORMATION WITHOUT SUCCESS.
Patient problems: Skin Infection
MAUDE report 25659393 ↗InjuryJun 24, 2026Migration
THE PATIENT REPORTED LOSS OF THERAPY AND PAIN FROM THE NEUROSTIMULATORS POKING THEM UNDER THE SKIN. SEVERAL ATTEMPTS HAVE BEEN MADE TO CHANGE THE PROGRAMS ON THE TRANSMITTER ASSEMBLY WITHOUT SUCCESS. HOWEVER, X-RAYS CONFIRMED DEVICE MIGRATION. AN EXPLANT PROCEDURE WAS PERFORMED ON (B)(6) 2026. NO FURTHER ISSUES HAVE BEEN REPORTED.
Patient problems: Pain
MAUDE report 25616395 ↗InjuryJun 24, 2026Off-Label Use
THE PATIENT REPORTED MANY INSTANCES OF "SHOCKS" IN THEIR CHEEK WHILE WEARING THE DEVICE. THE PATIENT WAS ADVISED TO STOP USING THE DEVICE UNTIL THE PROGRAMS ON THE TRANSMITTER ASSEMBLY COULD BE CHANGED. ON (B)(6) 2026, THE PROGRAMS ON THE TRANSMITTER ASSEMBLY WERE LOWERED, THE PATIENT WAS GIVEN A PRESCRIPTION OF LIDOCAINE SPRAY AND PATCHES, AND THE PATIENT WAS ASKED TO NOT USE THEIR DEVICES. THE PATIENT WILL TRY THE NEW PROGRAMS ON THE TRANSMITTER ASSEMBLY WHEN THEY RETURN FROM A THREE-WEEK HOLIDAY AND THEIR POST-OPERATIVE PAIN HAS SUBSIDED. THE PATIENT WILL FOLLOW-UP WITH THE HOSPITAL OUTPATIENT CLINIC. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Shock from Patient Lead(s)
MAUDE report 25613357 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K152178. One submission often covers many models, so check the product and lot details in each record.
- Z-2933-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models: 1. FR4A-RCV-AO US and 2. FR4A-RCV-A0 EU - Product Usage: is used for peripheral neural stimulation to provide therapeutic re
- Z-2928-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model 1. STQ4-RCV-A0 US and 2.STQ4-RCV-A0 EU - Product Usage: is used for peripheral neural stimulation to provide therapeutic relie
- Z-2931-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models FR4A-SPR-BO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractab
- Z-2930-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models: 1. FR4A-TRL-BO US and 2. FR4A-TR-BO - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief
- Z-2929-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model FR4A-TRL-AO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractabl
- Z-2932-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model STQ4-SPR-BO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractabl
Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices
- Quattrode™ 3166St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)429 reports
- Quattrode™ 3169St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)401 reports
- Nalu Medical, Inc. 71019-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall225 reports
- StimQ Receiver Stimulator Kit Channel A US w/Receiver STQ4-RCV-A0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)3 deaths270 reports
- Nalu Medical, Inc. 71020-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)213 reports
- StimQ Spare Lead Kit Channel B US w/Receiver STQ4-SPR-B0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)126 reports
- Nalu Medical, Inc. 71005-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall73 reports
- Nalu Medical, Inc. 71020Nalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)101 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page