Quattrode™ 3166 (UDI 05414734401661)Recalls & reported problems

St. Jude Medical, Inc. · Lead Wide Spaced, 60 cm · Class 2 · Implant

Quick summary

This St. Jude Medical Quattrode implanted nerve stimulator lead has 429 reports, all listed as injuries, and no deaths. Common problems described include pain relief failing, often because a lead moved out of position, and some patients needed surgery to reposition or remove the lead. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
429
Deaths reported
0
Injuries reported
429
Recalls naming it
0

Summary

The FDA's MAUDE database contains 429 adverse event reports that list this exact device (DI 05414734401661), received between Jul 12, 2018 and Sep 28, 2026 — classified as 429 injuries.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (117 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 4 reportsMar 2024: 1 reportsApr 2024: 4 reportsMay 2024: 2 reportsJun 2024: 6 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 8 reportsOct 2024: 1 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 4 reportsMar 2025: 12 reportsApr 2025: 6 reportsMay 2025: 1 reportsJun 2025: 7 reportsJul 2025: 10 reportsAug 2025: 3 reportsSep 2025: 6 reportsOct 2025: 1 reportsNov 2025: 10 reportsDec 2025: 2 reportsJan 2026: 7 reportsFeb 2026: 6 reportsMar 2026: 10 reportsApr 2026: 3 reportsMay 2026: 2 reportsJun 2026: 6 reportsJul 2026: 9 reportsAug 2026: 3 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem137
  • Therapeutic or Diagnostic Output Failure117
  • Migration65
  • High impedance56
  • Fracture23
  • Inappropriate/Inadequate Shock/Stimulation23
  • Insufficient Device Problem Information16
  • Failure to Deliver Energy4

Patient problems

  • Inadequate Pain Relief255
  • Implant Pain54
  • Undesired Nerve Stimulation26
  • Erosion18
  • No Consequences Or Impact To Patient17
  • Unspecified Infection16
  • No Clinical Signs, Symptoms or Conditions15
  • Post Operative Wound Infection14

Recent adverse event reports

The 25 most recent of 429 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Migration

    It was reported that the patient experienced ineffective therapy and headaches. X-ray imaging revealed migration of the Left Occipital lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue.

    Patient problems: Headache; Inadequate Pain Relief

    MAUDE report 50917646 ↗
  • InjurySep 24, 2026Migration

    It was reported that the patient's lead migrated. X-ray imaging revealed migration of the left supraorbital lead migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue.

    Patient problems: Insufficient Information

    MAUDE report 50888635 ↗
  • InjurySep 22, 2026Use of Device Problem

    It was reported that the patient's right Supra-orbital lead had migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue. X-ray imaging revealed the right occipital lead was accidentally moved during the procedure. As a result, the patient returned to surgery on (b)(6) 2026 wherein the right occipital lead was repositioned to address the issue. Effective therapy was restored post operatively. It is unknown which leads were the right Supra-orbital and right occipital lead.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860382 ↗
  • InjurySep 22, 2026Migration

    It was reported that the patient experienced ineffective therapy. X-ray imaging revealed migration of the lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50849947 ↗
  • InjurySep 21, 2026Material Twisted/Bent

    It was reported that following recent weight loss, the patient experienced pain at the IPG site. Additionally, the right Occipital lead migrated and bent and another lead migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was repositioned, the right Occipital lead was explanted and replaced, and the other lead was repositioned to address the issue. It is unknown which lead migrated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846748 ↗
  • InjurySep 21, 2026Migration

    It was reported that following recent weight loss, the patient experienced pain at the IPG site. Additionally, the right Occipital lead migrated and bent and another lead migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was repositioned, the right Occipital lead was explanted and replaced, and the other lead was repositioned to address the issue. It is unknown which lead migrated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846746 ↗
  • InjurySep 16, 2026Migration

    It was reported that the patient experienced ineffective therapy, the leads were uncomfortable and visible. X-ray imaging revealed migration of two leads. As a result, surgical intervention was undertaken wherein the system was explanted to address the issue.

    Patient problems: Implant Pain; Inadequate Pain Relief

    MAUDE report 50789583 ↗
  • InjurySep 16, 2026Migration

    It was reported that the patient experienced ineffective therapy, the leads were uncomfortable and visible. X-ray imaging revealed migration of two leads. As a result, surgical intervention was undertaken wherein the system was explanted to address the issue.

    Patient problems: Implant Pain; Inadequate Pain Relief

    MAUDE report 50789581 ↗
  • InjuryAug 20, 2026Fracture

    It was reported that the patient's lead exhibited high impedances. X-ray imaging revealed fracture of left supraorbital lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50355734 ↗
  • InjuryAug 6, 2026Inappropriate/Inadequate Shock/Stimulation

    It was reported that following a recent fall the patient experienced ineffective therapy and uncomfortable stimulation of the right supraorbital lead. Reprogramming has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue

    Patient problems: Inadequate Pain Relief

    MAUDE report 50166725 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that following weight loss, the patient experienced skin erosion at the lead and extension sites. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Erosion

    MAUDE report 50119165 ↗
  • InjuryJul 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported patient underwent surgical intervention on (b)(6) 2026 wherein the right occipital lead was repositioned due to being superficial. Therapy was restored post-op.

    Patient problems: Erosion

    MAUDE report 50074399 ↗
  • InjuryJul 28, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a jet-ski accident. Diagnostics revealed high impedances on all 4 leads. As a result, surgical intervention may be undertaken in the future.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50063606 ↗
  • InjuryJul 28, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a jet-ski accident. Diagnostics revealed high impedances on all 4 leads. As a result, surgical intervention may be undertaken in the future.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50063600 ↗
  • InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient experienced skin erosion at the lead sites. As a result, surgical intervention was undertaken wherein the entire system was explanted on (b)(6) 2026 to address the issue.

    Patient problems: Skin Erosion

    MAUDE report 50060748 ↗
  • InjuryJul 17, 2026Migration

    IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25862052 ↗
  • InjuryJul 8, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. VISUAL CONFIRMATION REVEALED THE LEFT SUPRA-ORBITAL LEAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPOSITIONED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25748354 ↗
  • InjuryJul 7, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO MIGRATION OF TWO LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN (B)(6) 2026 WHEREIN THE LEADS WERE REPOSITIONED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25736219 ↗
  • InjuryJul 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT'S LEADS WIRES COMING ACROSS HER CLAVICLE CAUSING PATIENT PAIN. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 25683077 ↗
  • InjuryJul 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT'S LEADS WIRES COMING ACROSS HER CLAVICLE CAUSING PATIENT PAIN. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 25682916 ↗
  • InjuryJun 26, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN TO REPOSITION THE LEFT SUPRA-ORBITAL LEAD.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25637752 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD DISCOMFORT AT THE LEAD SITES AND INEFFECTIVE STIMULATION THAT WAS NOT RESOLVED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHERE THE SYSTEM WAS REMOVED TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 25465617 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AT THE IPG AND LEAD SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE SYSTEM WAS EXPLANTED.

    Patient problems: Implant Pain

    MAUDE report 25461288 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AT THE IPG AND LEAD SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE SYSTEM WAS EXPLANTED.

    Patient problems: Implant Pain

    MAUDE report 25461287 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AT THE IPG AND LEAD SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE SYSTEM WAS EXPLANTED.

    Patient problems: Implant Pain

    MAUDE report 25461285 ↗

Other STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF) devices

All STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.