Dario LC Starter Kit 1147-05Recalls & reported problems
Labstyle Innovation Ltd · The Dario LC Blood Glucose Monitoring System is a compact, all-in-one blood glucose meter that is used in conjunction with a smart mobile device (SMD). The Dario LC meter is connected to a SMD through an Apple Lightning® connector in order to provide the glucose meter with power. The SMD also provides an external display for the user to review blood glucose testing results via a mobile application downloaded to the SMD. The full process of blood glucose measurement is done on the glucose meter; the SMD does not provide any diagnostic or clinical analysis function. When connected to a SMD, the Dario LC glucose meter is agnostic both to the platform of the smart mobile device and to the model of the individual SMD within a platform. (DI# 00850065006154) The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.(DI#00850065006178) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00850065006253), received between Jun 27, 2024 and May 13, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (11 reports).
1 other recall cover devices cleared under the same FDA submission (K150817, K172723); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings11
- High Readings7
- Low Readings5
Patient problems
- Hyperglycemia7
- Hypoglycemia5
- Insufficient Information2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
ON APRIL 13TH, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER A BG READING OF 159 MG/DL FROM ONE TEST STRIP CARTRIDGE COMPARED TO A BG READING OF 128 MG/DL FROM A DIFFERENT TEST STRIP CARTRIDGE. THE USER REPORTED THAT BOTH CARTRIDGES HAD THE SAME LOT NUMBER AND EXPIRATION DATE, BEING THE 15TH OF APRIL. THE USER ALSO REPORTED THAT HE TESTED HIS BG SEVERAL TIMES USING BOTH TEST STRIP CARTRIDGES, AND EACH TIME HE RECEIVED A 30 MG/DL DIFFERENCE IN BG READINGS BETWEEN THE TWO CARTRIDGES.
Patient problems: Insufficient Information
MAUDE report 25158286 ↗MalfunctionMay 5, 2026Incorrect, Inadequate or Imprecise Result or Readings; High Readings
ON APRIL 7TH, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING HIGH BG READINGS FROM HIS DARIO METER WHEN COMPARED TO THE READINGS THAT HE RECEIVED FROM TWO OTHER NON-DAIRO METERS. THE USER ALSO STATED THAT THE BG READINGS FROM THE NON-DARIO METERS WERE TAKEN BY REGISTERED NURSES.
Patient problems: Hyperglycemia
MAUDE report 25076947 ↗MalfunctionDec 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; Low Readings
ON NOVEMBER 4TH, THE USER CONTACTED DARIO TO REPORT LOW BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING AN A1C OF 12.8% FROM HER LABWORK, COMPARED TO AN ESTIMATED A1C OF 7.6% ALONG WITH AN AVERAGE BG READING OF 168 MG/DL FROM HER DARIO METER. THE USER ALSO STATED THAT SHE HAD COMPARED THE READINGS FROM HER DARIO METER AGAINST MULTIPLE OTHER NON-DARIO METERS AND FOUND THAT HER DARIO METER WAS CONSISTENTLY READING LOWER.
Patient problems: Hypoglycemia
MAUDE report 23719599 ↗MalfunctionOct 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6), DARIO'S REPRESENTATIVE CONTACTED THE USER IN ORDER TO FOLLOW UP REGARDING THE REPLACEMENT METER THAT THE USER RECENTLY RECEIVED. THE USER REPORTED A DIFFERENCE IN THE RANGE OF 15 MG/DL TO 20 MG/DL IN HER BG READINGS WHEN PERFORMING CONSECUTIVE TESTS WITH HER DARIO METER.
Patient problems: Insufficient Information
MAUDE report 23301955 ↗MalfunctionSep 9, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings
ON (B)(6), THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER, WHO HAS TWO DARIO METERS, REPORTED RECEIVING A BG READING OF 35 MG/DL FOLLOWED BY A BG READING OF 520 MG/DL FROM ONE OF THE DARIO METERS. THE USER STATED THAT SHE STORES ONE DARIO METER IN HER PURSE, AND THE OTHER AT HOME.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 23002816 ↗MalfunctionAug 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings
ON JULY 13TH, THE USER CONTACTED DARIO TO REPORT INCONSISTENT AND HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER A BG READING OF 418 MG/DL COMPARED TO A BG READING OF 129 MG/DL FROM HIS NON-DARIO METER, WHICH HE RECEIVED APPROXIMATELY 40 MINUTES LATER. THE USER STATED THAT HIS NORMAL BG RANGE IS 129 MG/DL.
Patient problems: Hyperglycemia
MAUDE report 22784362 ↗MalfunctionMay 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings
ON APRIL 28TH, THE USER CONTACTED DARIO TO REPORT HIGH AND INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER BG READINGS THAT WERE 25 MG/DL HIGHER AND VARYING WHEN HE PERFORMED CONSECUTIVE TESTS.
Patient problems: Hyperglycemia
MAUDE report 22078457 ↗MalfunctionSep 6, 2024Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; Low Readings
ON AUGUST 7TH, THE USER CONTACTED DARIO TO REPORT LOW BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED THAT A COUPLE OF MONTHS AGO, HE BEGAN RECEIVING LOW BG READINGS FROM HIS DARIO METER COMPARED TO THE BG READINGS FROM HIS DEXCOM METER AND HIS OTHER NON-DARIO METER. DUE TO THE ABOVE, THE USER REPORTED THAT HE BROUGHT HIS DARIO METER WITH HIM TO THE DOCTOR IN ORDER TO COMPARE HIS BG READINGS WITH THE DOCTOR'S METER.
Patient problems: Hypoglycemia
MAUDE report 20155897 ↗MalfunctionJul 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings
ON (B)(6) 2024, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING A BG READING OF 77 MG/DL FOLLOWED BY A BG READING OF 121 MG/DL FROM HIS DARIO METER.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 19749386 ↗MalfunctionJul 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings
ON JUNE 10TH, 2024, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING A BG READING OF 158 MG/DL FROM HIS DARIO METER. DUE TO THIS BG READING, THE USER OPENED A NEW CARTRIDGE OF STRIPS, WHERE HE TESTED AND RECEIVED A BG READING OF 89 MG/DL, WHICH THE USER STATED IS WITHIN HIS NORMAL FASTING RANGE OF 80 MG/DL TO 100 MG/DL. THE USER ALSO REPORTED THAT HE RECEIVED BG READINGS OF 79 MG/DL ON APRIL 28TH, 64 MG/DL ON MAY 12TH, AND 52 MG/DL ON MAY 26TH, WITHOUT EXPERIENCING ANY HYPOGLYCEMIC SYMPTOMS. DUE TO THESE LOW BG READINGS, THE USER DISCARDED THREE STRIP CARTRIDGES.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 19702414 ↗MalfunctionJun 27, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings
ON (B)(6) 2024, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED THAT HE STARTED RECEIVING HIGH BG READINGS ALL OF A SUDDEN.
Patient problems: Hyperglycemia
MAUDE report 19627659 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150817, K172723. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2020 · Aug 29, 2019 · TerminatedDario Blood Glucose Monitoring System, Model No. 1015-05, running Android software versions 4.3.0-4.3.2
Other System, Test, Blood Glucose, Over The Counter devices
- OneTouch Verio Flex 023-271LifeScan Europe GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER3,703 reports
- FreeStyle Libre 2 71953Abbott Diabetes Care Inc · System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- OneTouch Ultra2 021-105LifeScan U.S. LLC · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- Omnipod DASH PDM PDM-USA1-D001-MG-USA1Insulet Corporation · System, Test, Blood Glucose, Over The Counter1 recall10 deaths1,129 reports
- Omnipod UST400Insulet Corporation · System, Test, Blood Glucose, Over The Counter12 deaths816 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE METER - NOT FOR RETAIL SALETrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter418 reports
- FreeStyle Freedom Lite 70923-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Lite 70829-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page