Dario LC Starter Kit 1147-05Recalls & reported problems

Labstyle Innovation Ltd · The Dario LC Blood Glucose Monitoring System is a compact, all-in-one blood glucose meter that is used in conjunction with a smart mobile device (SMD). The Dario LC meter is connected to a SMD through an Apple Lightning® connector in order to provide the glucose meter with power. The SMD also provides an external display for the user to review blood glucose testing results via a mobile application downloaded to the SMD. The full process of blood glucose measurement is done on the glucose meter; the SMD does not provide any diagnostic or clinical analysis function. When connected to a SMD, the Dario LC glucose meter is agnostic both to the platform of the smart mobile device and to the model of the individual SMD within a platform. (DI# 00850065006154) The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.(DI#00850065006178) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00850065006253), received between Jun 27, 2024 and May 13, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (11 reports).

1 other recall cover devices cleared under the same FDA submission (K150817, K172723); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings11
  • High Readings7
  • Low Readings5

Patient problems

  • Hyperglycemia7
  • Hypoglycemia5
  • Insufficient Information2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Incorrect, Inadequate or Imprecise Result or Readings

    ON APRIL 13TH, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER A BG READING OF 159 MG/DL FROM ONE TEST STRIP CARTRIDGE COMPARED TO A BG READING OF 128 MG/DL FROM A DIFFERENT TEST STRIP CARTRIDGE. THE USER REPORTED THAT BOTH CARTRIDGES HAD THE SAME LOT NUMBER AND EXPIRATION DATE, BEING THE 15TH OF APRIL. THE USER ALSO REPORTED THAT HE TESTED HIS BG SEVERAL TIMES USING BOTH TEST STRIP CARTRIDGES, AND EACH TIME HE RECEIVED A 30 MG/DL DIFFERENCE IN BG READINGS BETWEEN THE TWO CARTRIDGES.

    Patient problems: Insufficient Information

    MAUDE report 25158286 ↗
  • MalfunctionMay 5, 2026Incorrect, Inadequate or Imprecise Result or Readings; High Readings

    ON APRIL 7TH, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING HIGH BG READINGS FROM HIS DARIO METER WHEN COMPARED TO THE READINGS THAT HE RECEIVED FROM TWO OTHER NON-DAIRO METERS. THE USER ALSO STATED THAT THE BG READINGS FROM THE NON-DARIO METERS WERE TAKEN BY REGISTERED NURSES.

    Patient problems: Hyperglycemia

    MAUDE report 25076947 ↗
  • MalfunctionDec 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; Low Readings

    ON NOVEMBER 4TH, THE USER CONTACTED DARIO TO REPORT LOW BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING AN A1C OF 12.8% FROM HER LABWORK, COMPARED TO AN ESTIMATED A1C OF 7.6% ALONG WITH AN AVERAGE BG READING OF 168 MG/DL FROM HER DARIO METER. THE USER ALSO STATED THAT SHE HAD COMPARED THE READINGS FROM HER DARIO METER AGAINST MULTIPLE OTHER NON-DARIO METERS AND FOUND THAT HER DARIO METER WAS CONSISTENTLY READING LOWER.

    Patient problems: Hypoglycemia

    MAUDE report 23719599 ↗
  • MalfunctionOct 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ON (B)(6), DARIO'S REPRESENTATIVE CONTACTED THE USER IN ORDER TO FOLLOW UP REGARDING THE REPLACEMENT METER THAT THE USER RECENTLY RECEIVED. THE USER REPORTED A DIFFERENCE IN THE RANGE OF 15 MG/DL TO 20 MG/DL IN HER BG READINGS WHEN PERFORMING CONSECUTIVE TESTS WITH HER DARIO METER.

    Patient problems: Insufficient Information

    MAUDE report 23301955 ↗
  • MalfunctionSep 9, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings

    ON (B)(6), THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER, WHO HAS TWO DARIO METERS, REPORTED RECEIVING A BG READING OF 35 MG/DL FOLLOWED BY A BG READING OF 520 MG/DL FROM ONE OF THE DARIO METERS. THE USER STATED THAT SHE STORES ONE DARIO METER IN HER PURSE, AND THE OTHER AT HOME.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 23002816 ↗
  • MalfunctionAug 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings

    ON JULY 13TH, THE USER CONTACTED DARIO TO REPORT INCONSISTENT AND HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER A BG READING OF 418 MG/DL COMPARED TO A BG READING OF 129 MG/DL FROM HIS NON-DARIO METER, WHICH HE RECEIVED APPROXIMATELY 40 MINUTES LATER. THE USER STATED THAT HIS NORMAL BG RANGE IS 129 MG/DL.

    Patient problems: Hyperglycemia

    MAUDE report 22784362 ↗
  • MalfunctionMay 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings

    ON APRIL 28TH, THE USER CONTACTED DARIO TO REPORT HIGH AND INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING FROM HIS DARIO METER BG READINGS THAT WERE 25 MG/DL HIGHER AND VARYING WHEN HE PERFORMED CONSECUTIVE TESTS.

    Patient problems: Hyperglycemia

    MAUDE report 22078457 ↗
  • MalfunctionSep 6, 2024Incorrect, Inadequate or Imprecise Result or Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; Low Readings

    ON AUGUST 7TH, THE USER CONTACTED DARIO TO REPORT LOW BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED THAT A COUPLE OF MONTHS AGO, HE BEGAN RECEIVING LOW BG READINGS FROM HIS DARIO METER COMPARED TO THE BG READINGS FROM HIS DEXCOM METER AND HIS OTHER NON-DARIO METER. DUE TO THE ABOVE, THE USER REPORTED THAT HE BROUGHT HIS DARIO METER WITH HIM TO THE DOCTOR IN ORDER TO COMPARE HIS BG READINGS WITH THE DOCTOR'S METER.

    Patient problems: Hypoglycemia

    MAUDE report 20155897 ↗
  • MalfunctionJul 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings

    ON (B)(6) 2024, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING A BG READING OF 77 MG/DL FOLLOWED BY A BG READING OF 121 MG/DL FROM HIS DARIO METER.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 19749386 ↗
  • MalfunctionJul 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Low Readings

    ON JUNE 10TH, 2024, THE USER CONTACTED DARIO TO REPORT INCONSISTENT BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED RECEIVING A BG READING OF 158 MG/DL FROM HIS DARIO METER. DUE TO THIS BG READING, THE USER OPENED A NEW CARTRIDGE OF STRIPS, WHERE HE TESTED AND RECEIVED A BG READING OF 89 MG/DL, WHICH THE USER STATED IS WITHIN HIS NORMAL FASTING RANGE OF 80 MG/DL TO 100 MG/DL. THE USER ALSO REPORTED THAT HE RECEIVED BG READINGS OF 79 MG/DL ON APRIL 28TH, 64 MG/DL ON MAY 12TH, AND 52 MG/DL ON MAY 26TH, WITHOUT EXPERIENCING ANY HYPOGLYCEMIC SYMPTOMS. DUE TO THESE LOW BG READINGS, THE USER DISCARDED THREE STRIP CARTRIDGES.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 19702414 ↗
  • MalfunctionJun 27, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Incorrect, Inadequate or Imprecise Result or Readings; High Readings

    ON (B)(6) 2024, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED THAT HE STARTED RECEIVING HIGH BG READINGS ALL OF A SUDDEN.

    Patient problems: Hyperglycemia

    MAUDE report 19627659 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150817, K172723. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2020 · Aug 29, 2019 · Terminated
    Dario Blood Glucose Monitoring System, Model No. 1015-05, running Android software versions 4.3.0-4.3.2