Skip to content

SonarMed AirWave AW-M0001Recalls & reported problems

Sonarmed, Inc · AirWave Monitor - Portable/handheld monitor packaged with power transformer and electrical cord, data communication cable, reusable sensor cable, hang strap and software CD. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00851334007001), received between Sep 17, 2024 and Sep 17, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Calibrate” (1 report).

3 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2024Class IOpen, Classified

    Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.

    Z-1535-2024 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 22, 2021Class IIOpen, Classified

    Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.

    Z-0514-2022 · Root cause: Process control · Recall details · FDA record ↗

  3. Apr 17, 2017Class IITerminated

    Potential for the presence of two error codes which would make the monitor inoperable.

    Z-2450-2019 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Calibrate1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THIS SONARMED DEVICE HAD MUFFLED SPEAKER. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238637 ↗
  • MalfunctionSep 17, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED THAT, THIS SONARMED DEVICE WAS UNABLE TO CALIBRATE. IT WAS ALSO NOTED THAT THE DEVICE DISPLAYED "CHECK/CHANGE SENSOR" ERROR MESSAGE AFTER TRYING TO DISCONNECT AND RECONNECT THE SENSOR. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238411 ↗

Other Airway Monitoring System devices

All Airway Monitoring System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page