Skip to content

SonarMed AirWave Monitor, Model Number M0001 recallZ-2450-2019 · SonarMed Inc · initiated Apr 17, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SonarMed Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the presence of two error codes which would make the monitor inoperable.

FDA root cause category: Software design

What the company is doing

April 2017, firm notifications were made in person at the time of the visit to upgrade the firmware. All monitors that had not been distributed were upgraded to the new firmware. All subsequent shipments will have the updated firmware.

Recall details

Recall number
Z-2450-2019
Status
Terminated
Initiated by the company
Apr 17, 2017
Recalling firm
SonarMed Inc
Quantity
16 units
Product code
OQU
FDA submissions
K143042
Distribution
CA & TX

Product description

SonarMed AirWave Monitor, Model Number M0001

Code information (models, lots, serial numbers)

UDI Numbers: (01)00851334007001(11)130426(21)339 (01)00851334007001(11)130502(21)342 (01)00851334007001(11)131105(21)365 (01)00851334007001(11)140522(21)369 (01)00851334007001(11)140522(21)373 (01)00851334007001(11)150609(21)379 (01)00851334007001(11)150609(21)381 (01)00851334007001(11)150609(21)382 (01)00851334007001(11)150630(21)383 (01)00851334007001(11)150630(21)385 (01)00851334007001(11)150630(21)386 (01)00851334007001(11)150630(21)388 (01)00851334007001(11)150630(21)389 (01)00851334007001(11)150630(21)390 (01)00851334007001(11)150701(21)391 (01)00851334007001(11)150701(21)393

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.