Bigger Better-Bladder BBB38Recalls & reported problems
Circulatory Technology Inc · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00851997007011), received between Nov 21, 2019 and Mar 25, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Collapse” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K981284); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Collapse1
- Defective Component1
- Gas/Air Leak1
- Pressure Problem1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2020Collapse
ON 2/24/2020, HOSPITAL NOTIFIED MANUFACTURER THAT TWO BIGGER BETTER-BLADDERS, MODEL BBB38, BEGAN TO COLLAPSE WITHIN 24 HOURS OF CLINICAL USE. THE FIRST BBB38 STARTED TO COLLAPSE UPON ECLS INITIATION ON (B)(6) 2020. THE 2ND COLLAPSED 12 HRS AFTER A CIRCUIT CHANGE ON (B)(6) 2020. THEY KEPT COLLAPSING AND WERE CHANGED WITH A BBB38 FROM ANOTHER LOT. THE PRODUCTS WERE USED ON THE SAME PATIENT. THE UNITS WERE REPLACED AND THERE WAS NO HARM OR ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9879086 ↗MalfunctionNov 21, 2019Defective Component; Gas/Air Leak; Pressure Problem
FROM MEDWATCH REPORT (B)(4): THE BIGGER BETTER BLADDER IN THE ECMO CIRCUIT HAS AN AIR CHAMBER ON THE OUTSIDE FOR EQUILIBRATING THE PRESSURE DURING THE RUN THAT ENSURES APPROPRIATE NEGATIVE PRESSURE ON THE SYSTEM. DURING THE PREVIOUS NIGHT, SPECIALIST HAS TROUBLE MAINTAINING IN THE AIR CHAMBER TO THE RIGHT PRESSURE. THEY FOUND THE AIR CHAMBER PIGTAIL WAS LEAKING REQUIRING THE BLABBER TO BE CHANGED ON THE CIRCUIT. THE PATIENT WAS OFF BYPASS FOR 1:20 (ONE MINUTE 20 SECONDS) DURING THE CHANGE, BUT TOLERATED THE BRIEF TIME WITHOUT INCIDENT. POSSIBLY A MANUFACTURER DEFECT. EVENT WAS RELATED TO A DEFECTIVE CONSUMABLE WHICH WAS REPLACED WITH NO EVENT. MACHINE DID NOT MALFUNCTION-NO FURTHER ACTION BY BIOMED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9354816 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K981284. One submission often covers many models, so check the product and lot details in each record.
- Z-1862-2021 · Apr 27, 2021 · Open, ClassifiedThe Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14
- Z-0817-2018 · Sep 22, 2017 · TerminatedBetter-Bladder cardiopulmonary bypass blood reservoir; Models: BB14 (individual, sterile, pouched) and BB14NS (sold bulk to kit manufacturers)
- Z-1171-2018 · Sep 22, 2017 · Terminated**Recall Expanded to add these models: Bigger Better-Bladder cardiopulmonary bypass blood reservoir; Models: BBB38 (individual, sterile, pouched) and BBB38NS (sold bulk to kit manu
- Z-0057-2017 · Jul 1, 2016 · TerminatedThe Better-Bladder With 14" ID tubing, ITEM BB14, STERILE, Measure pressure noninvasively in the extracorporeal circuit and provide compliance in the venous line between the patie
- Z-2832-2016 · Jul 1, 2016 · TerminatedThe Bigger-Better-Bladder, TM 2 With 3/8"ID tubing, 72" long tubing, ITEM BBB38-72 A device that isolates pressure tranducers from blood contact when measurements of blood pressur
- Z-2050-2016 · Mar 3, 2016 · TerminatedBigger-Better-Bladder With 3/8" ID tubing, ITEM BBB38
- Z-1323-2010 · Jun 3, 2009 · TerminatedThe Bigger Better-Bladder, Catalog Number BBB38 (with 3/8 inch ID tubing), Sterile, Single use only --- Manufactured for: Circulatory Technology Inc., 21 Singworth Street, Oyster B
Other RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS devices
- Capiox® FX Advance Hollow Fiber Oxygenator w Integrated AF w Hardshell Res. 3CX*FX25RWCTerumo Cardiovascular Systems Corporation · Reservoir, Blood, Cardiopulmonary Bypass1 recall11 deaths286 reports
- Affinity Fusion® BB841Medtronic, Inc. · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS2 recalls36 reports
- NX19 Oxygenator w/4-liter reservoir; right-side outlet port 3CX*NX19RETerumo Cardiovascular Systems Corporation · Reservoir, Blood, Cardiopulmonary Bypass2 recalls27 reports
- Capiox® FX Advance Hollow Fiber Oxygenator w Integrated AF w Hardshell Res. 3CX*FX15RW40CTerumo Cardiovascular Systems Corporation · Reservoir, Blood, Cardiopulmonary Bypass3 deaths79 reports
- Capiox® FX Advance Hollow Fiber Oxygenator w Integrated AF w Hardshell Res. 3CX*FX15RW30CTerumo Cardiovascular Systems Corporation · Reservoir, Blood, Cardiopulmonary Bypass3 deaths57 reports
- Better-Bladder BB14Circulatory Technology Inc · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS1 recall5 reports
- Affinity Pixie™ BBP241Medtronic, Inc. · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS1 recall2 reports
- Affinity Fusion™ BB841Medtronic, Inc. · Reservoir, blood, cardiopulmonary bypass44 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.