Skip to content

Bigger Better-Bladder BBB38Recalls & reported problems

Circulatory Technology Inc · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00851997007011), received between Nov 21, 2019 and Mar 25, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Collapse” (1 report).

7 other recalls cover devices cleared under the same FDA submission (K981284); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Collapse1
  • Defective Component1
  • Gas/Air Leak1
  • Pressure Problem1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2020Collapse

    ON 2/24/2020, HOSPITAL NOTIFIED MANUFACTURER THAT TWO BIGGER BETTER-BLADDERS, MODEL BBB38, BEGAN TO COLLAPSE WITHIN 24 HOURS OF CLINICAL USE. THE FIRST BBB38 STARTED TO COLLAPSE UPON ECLS INITIATION ON (B)(6) 2020. THE 2ND COLLAPSED 12 HRS AFTER A CIRCUIT CHANGE ON (B)(6) 2020. THEY KEPT COLLAPSING AND WERE CHANGED WITH A BBB38 FROM ANOTHER LOT. THE PRODUCTS WERE USED ON THE SAME PATIENT. THE UNITS WERE REPLACED AND THERE WAS NO HARM OR ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9879086 ↗
  • MalfunctionNov 21, 2019Defective Component; Gas/Air Leak; Pressure Problem

    FROM MEDWATCH REPORT (B)(4): THE BIGGER BETTER BLADDER IN THE ECMO CIRCUIT HAS AN AIR CHAMBER ON THE OUTSIDE FOR EQUILIBRATING THE PRESSURE DURING THE RUN THAT ENSURES APPROPRIATE NEGATIVE PRESSURE ON THE SYSTEM. DURING THE PREVIOUS NIGHT, SPECIALIST HAS TROUBLE MAINTAINING IN THE AIR CHAMBER TO THE RIGHT PRESSURE. THEY FOUND THE AIR CHAMBER PIGTAIL WAS LEAKING REQUIRING THE BLABBER TO BE CHANGED ON THE CIRCUIT. THE PATIENT WAS OFF BYPASS FOR 1:20 (ONE MINUTE 20 SECONDS) DURING THE CHANGE, BUT TOLERATED THE BRIEF TIME WITHOUT INCIDENT. POSSIBLY A MANUFACTURER DEFECT. EVENT WAS RELATED TO A DEFECTIVE CONSUMABLE WHICH WAS REPLACED WITH NO EVENT. MACHINE DID NOT MALFUNCTION-NO FURTHER ACTION BY BIOMED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9354816 ↗

FDA clearance and classification

Cleared through 510(k) K981284 on Aug 12, 1998 (decision: Substantially Equivalent). Submitted by Circulatory Technology, Inc. as “Better-Bladder (Bb)”. The FDA review took 126 days. FDA record ↗

The FDA regulates product code DTN (reservoir, blood, cardiopulmonary bypass) under 21 CFR 870.4400 as a Class II device. 21 CFR 870.4400 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K981284. One submission often covers many models, so check the product and lot details in each record.

  • Z-1862-2021 · Apr 27, 2021 · Open, Classified
    The Better-Bladder (BB), part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits. Model: BB14
  • Z-1171-2018 · Sep 22, 2017 · Terminated
    **Recall Expanded to add these models: Bigger Better-Bladder cardiopulmonary bypass blood reservoir; Models: BBB38 (individual, sterile, pouched) and BBB38NS (sold bulk to kit manu
  • Z-0817-2018 · Sep 22, 2017 · Terminated
    Better-Bladder cardiopulmonary bypass blood reservoir; Models: BB14 (individual, sterile, pouched) and BB14NS (sold bulk to kit manufacturers)
  • Z-0057-2017 · Jul 1, 2016 · Terminated
    The Better-Bladder With 14" ID tubing, ITEM BB14, STERILE, Measure pressure noninvasively in the extracorporeal circuit and provide compliance in the venous line between the patie
  • Z-2832-2016 · Jul 1, 2016 · Terminated
    The Bigger-Better-Bladder, TM 2 With 3/8"ID tubing, 72" long tubing, ITEM BBB38-72 A device that isolates pressure tranducers from blood contact when measurements of blood pressur
  • Z-2050-2016 · Mar 3, 2016 · Terminated
    Bigger-Better-Bladder With 3/8" ID tubing, ITEM BBB38
  • Z-1323-2010 · Jun 3, 2009 · Terminated
    The Bigger Better-Bladder, Catalog Number BBB38 (with 3/8 inch ID tubing), Sterile, Single use only --- Manufactured for: Circulatory Technology Inc., 21 Singworth Street, Oyster B

Common questions about the Bigger Better-Bladder BBB38

Has the Bigger Better-Bladder BBB38 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00851997007011), based on the FDA recall file dated Nov 27, 2026. However, 7 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Bigger Better-Bladder BBB38?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 21, 2019 and Mar 25, 2020. In those reports, the most common specific device problems are “Collapse” (1), “Defective Component” (1) and “Gas/Air Leak” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Bigger Better-Bladder BBB38?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Bigger Better-Bladder BBB38 FDA approved?

It was cleared through 510(k) K981284 on Aug 12, 1998. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Bigger Better-Bladder BBB38, and is it still sold?

The FDA's device database lists Circulatory Technology Inc as the company responsible for it (brand Bigger Better-Bladder, model BBB38). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS devices

All RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page